| Literature DB >> 28690266 |
Yuya Hasegawa1,2, Yasutada Suzuki3, Susumu Kawakubo2.
Abstract
An on-site determination method for trace arsenic has been developed by collecting it as molybdenum blue (MB) in the presence of tetradecyldimethylbenzylammonium chloride on a mixed cellulose ester membrane filter and by measuring reflection absorbance (RA) of MB on the filter using a laboratory-made palm-top size reflection-absorbance colorimeter with a red light-emitting diode. The value of RA was proportional to the amount of arsenic up to 0.5 μg with a detection limit of 0.01 μg. The proposed method was successfully applied to soil extract and hot-spring water samples.Entities:
Keywords: Molybdenum blue; arsenic; colorimeter; hot-spring; on-site analysis; reflection; soil extract
Year: 2017 PMID: 28690266 DOI: 10.2116/analsci.33.859
Source DB: PubMed Journal: Anal Sci ISSN: 0910-6340 Impact factor: 2.081