Literature DB >> 28681982

Application of paper spray ionization for explosives analysis.

Chia-Wei Tsai1,2, Christopher A Tipple3, Richard A Yost1.   

Abstract

RATIONALE: A desired feature in the analysis of explosives is to decrease the time of the entire analysis procedure, including sampling. A recently utilized ambient ionization technique, paper spray ionization (PSI), provides the possibility of combining sampling and ionization. However, an interesting phenomenon that occurs in generating negatively charged ions pose some challenges in applying PSI to explosives analysis. The goal of this work is to investigate the possible solutions for generating explosives ions in negative mode PSI.
METHODS: The analysis of 2,4,6-trinitrotoluene (TNT), pentaerythritol tetranitrate (PETN), octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazocine (HMX), and 1,3,5-trinitroperhydro-1,3,5-triazine (RDX) was performed. Several solvent systems with different surface tensions and additives were compared to determine their effect on the ionization of explosives. The solvents tested include tert-butanol, isopropanol, methanol, and acetonitrile. The additives tested were carbon tetrachloride and ammonium nitrate.
RESULTS: Of the solvents tested, isopropanol yielded the best results. In addition, adding ammonium nitrate to the isopropanol enhanced the analyte signal. Experimentally determined limits of detection (LODs) as low as 0.06 ng for PETN, on paper, were observed with isopropanol and the addition of 0.4 mM ammonium nitrate as the spray solution. In addition, the explosive components of two plastic explosive samples, Composition 4 and Semtex, were successfully analyzed via surface sampling when using the developed method.
CONCLUSIONS: The analysis of explosives using PSI-MS in negative ion mode was achieved. The addition of ammonium nitrate to isopropanol, in general, enhanced the analyte signal and yielded better ionization stability. Real-world explosive samples were analyzed, which demonstrates one of the potential applications of PSI-MS analysis.
Copyright © 2017 John Wiley & Sons, Ltd.

Entities:  

Year:  2017        PMID: 28681982     DOI: 10.1002/rcm.7932

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  9 in total

1.  Paper spray ionization-high-resolution mass spectrometry (PSI-HRMS) of peroxide explosives in biological matrices.

Authors:  Michelle D Gonsalves; Alexander Yevdokimov; Audreyana Brown-Nash; James L Smith; Jimmie C Oxley
Journal:  Anal Bioanal Chem       Date:  2021-03-15       Impact factor: 4.142

2.  Direct Analysis of Doping Agents in Raw Urine Using Hydrophobic Paper Spray Mass Spectrometry.

Authors:  Eduardo Luiz Rossini; Dmytro S Kulyk; Emelia Ansu-Gyeabourh; Taghi Sahraeian; Helena Redigolo Pezza; Abraham K Badu-Tawiah
Journal:  J Am Soc Mass Spectrom       Date:  2020-05-14       Impact factor: 3.109

Review 3.  Recent advances in ambient mass spectrometry of trace explosives.

Authors:  Thomas P Forbes; Edward Sisco
Journal:  Analyst       Date:  2018-04-30       Impact factor: 4.616

Review 4.  Emerging trends in paper spray mass spectrometry: Microsampling, storage, direct analysis, and applications.

Authors:  Benjamin S Frey; Deidre E Damon; Abraham K Badu-Tawiah
Journal:  Mass Spectrom Rev       Date:  2019-09-06       Impact factor: 10.946

5.  Paper spray screening and liquid chromatography/mass spectrometry confirmation for medication adherence testing: A two-step process.

Authors:  Catia Costa; Cecile Frampas; Katherine A Longman; Vladimir Palitsin; Mahado Ismail; Patrick Sears; Ramin Nilforooshan; Melanie J Bailey
Journal:  Rapid Commun Mass Spectrom       Date:  2019-08-14       Impact factor: 2.419

Review 6.  Interpol review of detection and characterization of explosives and explosives residues 2016-2019.

Authors:  Douglas J Klapec; Greg Czarnopys; Julie Pannuto
Journal:  Forensic Sci Int       Date:  2020-06-17       Impact factor: 2.395

7.  Detection of Trace Explosives Using a Novel Sample Introduction and Ionization Method.

Authors:  Lingfeng Li; Tianyi Zhang; Wei Ge; Xingli He; Yunjing Zhang; Xiaozhi Wang; Peng Li
Journal:  Molecules       Date:  2022-07-17       Impact factor: 4.927

8.  Exploring a route to a selective and sensitive portable system for explosive detection- swab spray ionisation coupled to of high-field assisted waveform ion mobility spectrometry (FAIMS).

Authors:  C Costa; E M van Es; P Sears; J Bunch; Vladimir Palitsin; H Cooper; M J Bailey
Journal:  Forensic Sci Int       Date:  2019-08-27       Impact factor: 2.395

9.  Utilizing Surface Acoustic Wave Nebulization (SAWN) for the Rapid and Sensitive Ambient Ionization Mass Spectrometric Analysis of Organic Explosives.

Authors:  Lauren Pintabona; Alina Astefanei; Garry L Corthals; Arian C van Asten
Journal:  J Am Soc Mass Spectrom       Date:  2019-10-28       Impact factor: 3.109

  9 in total