Comparison of delay times of various respiratory gating systems
| Author | modality | ▵tbeam‐on (ms) | ▵tbeam‐off (ms) |
|---|---|---|---|
| Chang et al.(19) | Novalis Tx + ExacTrac | 200 ± 30 | |
| Novalis Tx + RPM | 90 ± 10 | ||
| Smith et al.(20) | Trilogy + Calypso | 75.0 ± 12.7 | 65.1 ± 12.9 |
| Woods et al.(21) | TrueBeam + RPM | 139 ± 10 | |
| AAPM TG142(22) | 100 | 100 | |
| This study | TrueBeam + SyncTraX | ||
| 6 MV‐FF |
| 71.5 ± 12.1 | |
| 6 MV‐FFF |
| 62.7 ± 14.1 | |
| 10 MV‐FF |
| 68.1 ± 7.2 | |
| 10 MV‐FFF |
| 69.0 ± 10.7 |
RPM, real‐time positioning management; FF, flattened; FFF = flattening‐filter‐free.