Literature DB >> 28681227

High incidence of early subtrochanteric lateral cortical atrophy after hip arthroplasty using bone-conserving short stem.

Yoon Je Cho1, Chan Il Bae2, Wan Keun Yoon2, Young Soo Chun2, Kee Hyung Rhyu3.   

Abstract

PURPOSE: Load transfer to the bone is believed to be more physiological around the short stem in total hip arthroplasty (THA). However, we found unusual bony remodeling around the shortened tapered stem. This study was performed to investigate the extent and frequency of this phenomenon and to find the possible risk factors of it.
METHODS: Among 121 consecutive THA using the same short stem, 80 THAs were enrolled. Radiographic measurements were made using anteroposterior (AP) radiographs taken immediately and at two years after surgery. The thickness of the lateral cortex at the level of the distal end of the coated surface and at 10, 20, 30, and 40 mm proximal to it were measured.
RESULTS: Significant atrophy was noted in all regions (P < 0.001 each). In 46 cases, this presented as an intra-cortical osteolytic line (IOL). Sixty-one cases showed either an IOL or atrophy >10%. The risk of a mean reduction >20% was related to an increased operating time (odds ratio [OR] = 0.981; 0.966 < 95% confidence interval [CI] < 0.996) and lower body mass index (BMI) (OR = 1.216; 1.043 < 95% CI < 1.417). Periprosthetic fracture through the lateral cortex occurred in one case.
CONCLUSION: Even with THA using a shortened stem, high incidence of proximal stress shielding was noted in the form of lateral cortical atrophy, especially for the patient with low BMI.

Entities:  

Keywords:  Arthroplasty; BMI; Hip; Short stem; Stress shielding

Mesh:

Year:  2017        PMID: 28681227     DOI: 10.1007/s00264-017-3544-9

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  24 in total

1.  Bone remodelling after femoral short stem implantation in total hip arthroplasty: 1-year results from a randomized DEXA study.

Authors:  Tobias Freitag; Marie-Anne Hein; Dirk Wernerus; Heiko Reichel; Ralf Bieger
Journal:  Arch Orthop Trauma Surg       Date:  2015-11-27       Impact factor: 3.067

2.  Biomechanics of a short stem: In vitro primary stability and stress shielding of a conservative cementless hip stem.

Authors:  Ralf Bieger; Anita Ignatius; Heiko Reichel; Lutz Dürselen
Journal:  J Orthop Res       Date:  2013-03-28       Impact factor: 3.494

3.  The NANOS short stem in total hip arthroplasty: a mid term follow-up.

Authors:  Max Ettinger; Peter Ettinger; Matthias Lerch; Kerstin Radtke; Stefan Budde; Marco Ezechieli; Christoph Becher; Fritz Thorey
Journal:  Hip Int       Date:  2011 Sep-Oct       Impact factor: 2.135

4.  Difference in Postoperative Periprosthetic Bone Mineral Density Changes Between 3 Major Designs of Uncemented Stems: A 3-Year Follow-Up Study.

Authors:  Yutaka Inaba; Naomi Kobayashi; Masatoshi Oba; Hiroyuki Ike; So Kubota; Tomoyuki Saito
Journal:  J Arthroplasty       Date:  2016-02-08       Impact factor: 4.757

5.  Stable fixation of short-stem femoral implants in patients 70 years and older.

Authors:  Ronak M Patel; Matthew C Smith; Chase C Woodward; S David Stulberg
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

6.  Structural and cellular assessment of bone quality of proximal femur.

Authors:  L D Dorr; M C Faugere; A M Mackel; T A Gruen; B Bognar; H H Malluche
Journal:  Bone       Date:  1993 May-Jun       Impact factor: 4.398

7.  Patterns of osteolysis around total hip components inserted with and without cement.

Authors:  B Zicat; C A Engh; E Gokcen
Journal:  J Bone Joint Surg Am       Date:  1995-03       Impact factor: 5.284

8.  A prospective short-term outcome study of a short metaphyseal fitting total hip arthroplasty.

Authors:  Young-Hoo Kim; Jun-Shik Kim; Jong-Hwan Joo; Jang-Won Park
Journal:  J Arthroplasty       Date:  2011-03-23       Impact factor: 4.757

Review 9.  Short bone-conserving stems in cementless hip arthroplasty.

Authors:  Harpal S Khanuja; Samik Banerjee; Deepak Jain; Robert Pivec; Michael A Mont
Journal:  J Bone Joint Surg Am       Date:  2014-10-15       Impact factor: 5.284

10.  Biomechanics of a new short-stemmed uncemented hip prosthesis: An in-vitro study in human bone.

Authors:  F M Westphal; N Bishop; K Puschel; M M Morlock
Journal:  Hip Int       Date:  2006       Impact factor: 1.756

View more
  4 in total

1.  Outcome of short versus conventional stem for total hip arthroplasty in the femur with a high cortical index: a five year follow-up prospective multicentre comparative study.

Authors:  Francesco Pogliacomi; Paolo Schiavi; Guido Grappiolo; Francesco Ceccarelli; Enrico Vaienti
Journal:  Int Orthop       Date:  2019-04-24       Impact factor: 3.075

2.  Development of periprosthetic bone mass density around the cementless Metha® short hip stem during three year follow up-a prospective radiological and clinical study.

Authors:  Laura Augustin; Sarah Boller; Cathrein Bobach; Alexander Jahnke; Gafar Adam Ahmed; Markus Rickert; Bernd Alexander Ishaque
Journal:  Int Orthop       Date:  2018-09-03       Impact factor: 3.075

3.  Fourteen-year experience with short cemented stems in total hip replacement.

Authors:  Nicola Santori; Francesco Falez; Domenico Potestio; Francesco Saverio Santori
Journal:  Int Orthop       Date:  2018-11-08       Impact factor: 3.075

4.  Incidence and pattern of periprosthetic hip fractures around the stem in different stem geometry.

Authors:  Umberto Cottino; Federico Dettoni; Giorgia Caputo; Davide E Bonasia; Paolo Rossi; Roberto Rossi
Journal:  Int Orthop       Date:  2019-05-16       Impact factor: 3.075

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.