BACKGROUND: Antibodies directed against dsDNA are a highly specific diagnostic marker for the presence of systemic lupus erythematosus and of particular importance in its diagnosis. To assess anti-dsDNA antibodies, the Crithidia luciliae-based indirect immunofluorescence test (CLIFT) is one of the assays considered to be the best choice. To overcome the drawback of subjective result interpretation that inheres indirect immunofluorescence assays in general, automated systems have been introduced into the market during the last years. Among these systems is the EUROPattern Suite, an advanced automated fluorescence microscope equipped with different software packages, capable of automated pattern interpretation and result suggestion for ANA, ANCA and CLIFT analysis.
METHODS: We analyzed the performance of the EUROPattern Suite with its automated fluorescence interpretation for CLIFT in a routine setting, reflecting the everyday life of a diagnostic laboratory. Three hundred and twelve consecutive samples were collected, sent to the Central Diagnostic Laboratory of the Maastricht University Medical Centre with a request for anti-dsDNA analysis over a period of 7 months.
RESULTS: Agreement between EUROPattern assay analysis and the visual read was 93.3%. Sensitivity and specificity were 94.1% and 93.2%, respectively. The EUROPattern Suite performed reliably and greatly supported result interpretation.
CONCLUSIONS: Automated image acquisition is readily performed and automated image classification gives a reliable recommendation for assay evaluation to the operator. The EUROPattern Suite optimizes workflow and contributes to standardization between different operators or laboratories.

Entities:  

Keywords:  Crithidia luciliae; autoantibodies; automation; dsDNA; indirect immunofluorescence; systemic lupus erythematosus (SLE)

Mesh:

Substances:

Year:  2017        PMID: 28672732     DOI: 10.1515/cclm-2017-0326

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  1 in total

1.  Performance analysis of automated evaluation of antinuclear antibody indirect immunofluorescent tests in a routine setting.

Authors:  Joyce J B C van Beers; Melanie Hahn; Johanna Fraune; Kathleen Mallet; Christopher Krause; Wymke Hormann; Kai Fechner; Jan G M C Damoiseaux
Journal:  Auto Immun Highlights       Date:  2018-09-21
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.