Bioprosthetic Valve Fracture to Facilitate Transcatheter Valve-in-Valve Implantation.

Keith B Allen1, Adnan K Chhatriwalla2, David J Cohen2, John T Saxon3, Sanjeev Aggarwal3, Anthony Hart3, Suzanne Baron2, J Russell Davis3, Alex F Pak3, Danny Dvir4, A Michael Borkon2.   

Abstract

BACKGROUND: Valve-in-valve transcatheter aortic valve replacement is less effective in small surgical bioprostheses. We evaluated the feasibility of bioprosthetic valve fracture with a high-pressure balloon to facilitate valve-in-valve transcatheter aortic valve replacement.
METHODS: In vitro bench testing on aortic tissue valves was performed on 19-mm and 21-mm Mitroflow (Sorin, Milan, Italy), Magna and Magna Ease (Edwards Lifesciences, Irvine, CA), Trifecta and Biocor Epic (St. Jude Medical, Minneapolis, MN), and Hancock II and Mosaic (Medtronic, Minneapolis, MN). High-pressure balloons Tru Dilation, Atlas Gold, and Dorado (C.R. Bard, Murray Hill, NJ) were used to determine which valves could be fractured and at what pressure fracture occurred.
RESULTS: Mitroflow, Magna, Magna Ease, Mosaic, and Biocor Epic surgical valves were successfully fractured using high-pressures balloon 1 mm larger than the labeled valve size whereas Trifecta and Hancock II surgical valves could not be fractured. Only the internal valve frame was fractured, and the sewing cuff was never disrupted. Manufacturer's rated burst pressures for balloons were exceeded, with fracture pressures ranging from 8 to 24 atmospheres depending on the surgical valve. Testing further demonstrated that fracture facilitated the expansion of previously constrained, underexpanded transcatheter valves (both balloon and self-expanding) to the manufacturer's recommended size.
CONCLUSIONS: Bench testing demonstrates that the frame of most, but not all, bioprosthetic surgical aortic valves can be fractured using high-pressure balloons. The safety of bioprosthetic valve fracture to optimize valve-in-valve transcatheter aortic valve replacement in small surgical valves requires further clinical investigation.
Copyright © 2017 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28669505     DOI: 10.1016/j.athoracsur.2017.04.007

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  23 in total

Review 1.  Transcatheter valve-in-valve implantation for degenerated surgical bioprostheses.

Authors:  Dale J Murdoch; John G Webb
Journal:  J Thorac Dis       Date:  2018-11       Impact factor: 2.895

Review 2.  Is Bioprosthetic Valve Fracture Beneficial for Patients with Larger Surgical Valves Undergoing Valve in Valve TAVR?

Authors:  Adnan K Chhatriwalla; John T Saxon; David J Cohen; Keith B Allen
Journal:  Curr Cardiol Rep       Date:  2018-08-22       Impact factor: 2.931

3.  Bioprosthetic Valve Fracture During Valve-in-valve TAVR: Bench to Bedside.

Authors:  John T Saxon; Keith B Allen; David J Cohen; Adnan K Chhatriwalla
Journal:  Interv Cardiol       Date:  2018-01

4.  Hemodynamic outcomes after valve-in-valve transcatheter aortic valve replacement: a single-center experience.

Authors:  R Yazan Kherallah; Srikanth Koneru; Zvonimir Krajcer; Ourania Preventza; Kathryn G Dougherty; Melissa L McCormack; Briana T Costello; Stephanie Coulter; Neil E Strickman; Juan Carlos Plana Gomez; Ali Mortazavi; Jose G Díez; James J Livesay; Joseph S Coselli; Guilherme V Silva
Journal:  Ann Cardiothorac Surg       Date:  2021-09

Review 5.  3D Printing Applications for Transcatheter Aortic Valve Replacement.

Authors:  Dmitry Levin; G Burkhard Mackensen; Mark Reisman; James M McCabe; Danny Dvir; Beth Ripley
Journal:  Curr Cardiol Rep       Date:  2020-02-17       Impact factor: 2.931

Review 6.  Transcatheter Aortic and Mitral Valve-in-Valve Implantation Using the Edwards Sapien 3 Heart Valve.

Authors:  Anupama Shivaraju; Jonathan Michel; Antonio H Frangieh; Ilka Ott; Christian Thilo; Heribert Schunkert; Adnan Kastrati; Martin B Leon; Danny Dvir; Susheel Kodali; Vinayak Bapat; Mayra Guerrero; Albert M Kasel
Journal:  J Am Heart Assoc       Date:  2018-07-07       Impact factor: 5.501

7.  Long-Term Results (up to 20 Years) of 19 mm or Smaller Prostheses in the Aortic Position. Does Size Matter? A Propensity-Matched Survival Analysis.

Authors:  Horea Feier; Andrei Grigorescu; Lucian Falnita; Oana Rachita; Marian Gaspar; Constantin T Luca
Journal:  J Clin Med       Date:  2021-05-11       Impact factor: 4.241

8.  The Future of Transcatheter Interventions.

Authors:  Tiffany Patterson; Christopher J Allen; Nicholas Aroney; Simon Redwood; Bernard Prendergast
Journal:  JACC Case Rep       Date:  2020-11-18

9.  Frequency of Bioprosthetic Valve Fracturing Utilization in an All-Comers Valve-in-Valve TAVR Cohort.

Authors:  Hendrik Ruge; Magdalena Erlebach; Ruediger Lange
Journal:  Front Cardiovasc Med       Date:  2021-07-01

Review 10.  Recent updates in transcatheter aortic valve implantation.

Authors:  Jeonghwan Cho; Ung Kim
Journal:  Yeungnam Univ J Med       Date:  2018-06-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.