Scan time adapted contrast agent injection protocols with low volume for low-tube voltage CT angiography: An in vitro study.

Matthias R Benz1, Zsolt Szucs-Farkas2, Johannes M Froehlich3, Geraldine Stadelmann4, Georg Bongartz4, Luc Bouwman5, Sebastian T Schindera6.   

Abstract

PURPOSE: The aims of this study were twofold. First, we investigated the extent of changes in arterial peak enhancement and changes in the duration of a diagnostic arterial enhancement when small amounts of CA volumes (≤30mL) were administered at varying tube voltages. Second, we investigated how to optimize CA injection protocols for CT-angiography with long scan times at various tube voltages to achieve optimal vascular enhancement at the lowest reasonable CA dose.
MATERIALS AND METHODS: Measurements were performed with a custom-made dynamic flow phantom. For CTA protocols with a short scan time, we investigated the effect of various tube voltages (70-120kVp) on the arterial enhancement profile with very small CA volumes (20 and 30mL of Iobitridol 350mg I/mL) at a flow rate of 5mL/s. For CTA protocols with a long scan time, we utilized an optimized multi-bolus technique switching rapidly between 13 "micro-boli" of CA (total, 60mL) and saline (total, 24mL) at a flow rate of 4mL/s. The peak arterial enhancement (PAE) and the time period of diagnostic aortic enhancement ≥200 HU (T200) were analyzed.
RESULTS: For the short scan time protocols, a diagnostic peak enhancement was achieved using 20mL of CA at 70 and 80kVp (PAE: 327±10 and 255±15 HU, respectively) or 30mL of CA at 70, 80 and 100kVp (PAE 451±10, 367±9, and 253±15 HU). For the long scan time, the optimized multi-bolus injection protocol extended T200 at 100kVp by 6s (40%) compared to a linear injection protocol (21±1s and 15±1s, respectively; p<0.001).
CONCLUSION: Optimized CTA protocols comprising alternations of tube voltage and the CA injection protocol can save radiation doses and CA volumes at the same time.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  CT-angiography; Contrast agent; Low tube voltage; Scan time

Mesh:

Substances:

Year:  2017        PMID: 28668433     DOI: 10.1016/j.ejrad.2017.05.017

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  2 in total

1.  Potential value of the PixelShine deep learning algorithm for increasing quality of 70 kVp+ASiR-V reconstruction pelvic arterial phase CT images.

Authors:  Shi-Feng Tian; Ai-Lian Liu; Jing-Hong Liu; Yi-Jun Liu; Ju-Dong Pan
Journal:  Jpn J Radiol       Date:  2018-12-06       Impact factor: 2.374

2.  Optimizing low contrast volume thoracic CT angiography: From the basics to the advanced.

Authors:  Katherine Kaproth-Joslin; Susan Hobbs; Prabhakar Rajiah; Apeksha Chaturvedi; Abhishek Chaturvedi
Journal:  J Clin Imaging Sci       Date:  2022-07-29
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.