Dosimetric properties of a Solid Water High Equivalency (SW557) phantom for megavoltage photon beams.

Fujio Araki1.   

Abstract

The dosimetric properties of the recently developed SW557 phantom have been investigated by comparison with those of the existing SW457 phantom in megavoltage photon beams. The electron fluence ratio φplw, and chamber ionization ratio kpl, of water to SW457 and water to SW557 for 4-15MV photons were calculated as a function of depth using Monte Carlo simulations, and compared with measured values. Values of φplw for SW457 were in the range of 1.004-1.014 for 4MV, and 1.014-1.018 for 15MV photons. The φplw for SW557 ranged from 1.005 to 1.008 for 4MV and from 1.010 to 1.015 for 15MV photons and the variation of φplw with depth for each beam energy was within ±0.5%. Values of kpl were obtained with a PTW 30013 Farmer-type ionization chamber. The kpl for SW457 ranged from 0.997 to 1.011 for 4-15MV photons. Values of kpl for SW557 were almost unity for 4 and 6MV photons, while in the case of 10 and 15MV photons they were less than 1.006, excepting the build-up region. The measured and calculated kpl values of water to SW557 were in the range of 0.997-1.002 and 1.000-1.006, respectively, for 4-15MV photons, at a depth of 10cm with a source-to-axis distance of 100cm. The measured and calculated kpl values were in agreement within their uncertainty ranges. As a water-equivalent phantom, SW557 can be used with a dosimetric difference within±0.6%, for 4-15MV photons, and is more water-equivalent than SW457 in megavoltage photon beams.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Chamber ionization ratio; Electron fluence ratio; Megavoltage photon beam; Monte Carlo simulation; Solid Water 557

Mesh:

Substances:

Year:  2017        PMID: 28662851     DOI: 10.1016/j.ejmp.2017.06.011

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

1.  Water equivalence of a solid phantom material for radiation dosimetry applications.

Authors:  Maegan A Gargett; Adam R Briggs; Jeremy T Booth
Journal:  Phys Imaging Radiat Oncol       Date:  2020-05-28
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.