| Literature DB >> 28661047 |
Yuki Yamamoto1, Daisuke Yamamoto1, Makoto Takada1, Hiroyoshi Naito1,2, Takayuki Arie1, Seiji Akita1, Kuniharu Takei1.
Abstract
Wearable, flexible healthcare devices, which can monitor health data to predict and diagnose disease in advance, benefit society. Toward this future, various flexible and stretchable sensors as well as other components are demonstrated by arranging materials, structures, and processes. Although there are many sensor demonstrations, the fundamental characteristics such as the dependence of a temperature sensor on film thickness and the impact of adhesive for an electrocardiogram (ECG) sensor are yet to be explored in detail. In this study, the effect of film thickness for skin temperature measurements, adhesive force, and reliability of gel-less ECG sensors as well as an integrated real-time demonstration is reported. Depending on the ambient conditions, film thickness strongly affects the precision of skin temperature measurements, resulting in a thin flexible film suitable for a temperature sensor in wearable device applications. Furthermore, by arranging the material composition, stable gel-less sticky ECG electrodes are realized. Finally, real-time simultaneous skin temperature and ECG signal recordings are demonstrated by attaching an optimized device onto a volunteer's chest.Keywords: electrocardiogram sensors; film thickness dependencies; healthcare patches; temperature sensors
Mesh:
Substances:
Year: 2017 PMID: 28661047 DOI: 10.1002/adhm.201700495
Source DB: PubMed Journal: Adv Healthc Mater ISSN: 2192-2640 Impact factor: 9.933