Literature DB >> 28659193

Feasibility of a prototype newborn resuscitation monitor to study transition at birth, measuring heart rate and ventilator parameters, an animal experimental study.

Jørgen E Linde1, Joar Eilevstjønn2, Knut Øymar3,4, Hege L Ersdal5.   

Abstract

BACKGROUND: Every year, an estimated 10 million babies are born, non-breathing and in need of resuscitation. Advances in management have been made over the past decades, however, approximately 700.000 yearly deaths result from this global problem. A prototype newborn resuscitation monitor (NRM) (Laerdal Global Health, Stavanger, Norway) has been developed with the purpose of studying newborn resuscitation. The monitor has the ability to continuously display HR using dry electrode ECG technology, to measure tidal volume, pressure and end tidal CO2, and to store the results for later analysis. Such monitor could enhance the care providers performance, and hence survival of neonates, by displaying the quality and response of the given care. The aim of this preclinical study was to describe the abilities of the NRM to measure ventilation and heart rate parameters against pathophysiological responses to different induced conditions in a piglet i.e. increased deadspace, pressure and washout of surfactant.
METHODS: Piglets were chosen for the study, as they have tidal volumes of approximately 6 ml/kg, resembling the human neonate. Five piglets were anesthetized and intubated before starting positive pressure ventilation (PPV). The dry electrode ECG sensor of the NRM was placed over the abdomen, and experiments performed: (1) inducing different ventilation scenarios and (2) lavage of surfactant.
RESULTS: The NRM was capable of continuously displaying HR and detecting inflicted changes in ventilation and compliance of piglets. It could measure inflated and exhaled volume, the pressure of the ventilations and also the end tidal CO2.
CONCLUSIONS: The NRM provides objective feedback in anesthetized animals, and may be used in clinical studies and hopefully generate new knowledge on neonatal transition and resuscitation. The monitor may be further developed for use in both low and high-resource settings.

Entities:  

Keywords:  Monitoring; Newborn; Resuscitation

Mesh:

Substances:

Year:  2017        PMID: 28659193      PMCID: PMC5490173          DOI: 10.1186/s13104-017-2530-z

Source DB:  PubMed          Journal:  BMC Res Notes        ISSN: 1756-0500


Background

Every year, an estimated 10 million babies are born non-breathing and in need of resuscitation [1]. Advances in management of newborn infants requiring delivery room resuscitation have been made over the past decades, however, approximately 700.000 yearly deaths result from this global problem [2]. In 2010, the International Liaison Committee on Resuscitation (ILCOR) raised the important question of whether improved monitoring of the newborn would be of benefit in facilitating effective resuscitation including positive pressure ventilation [2]. During positive pressure ventilation, pressure and volume may individually play important roles in the establishment of effective or harmful ventilations. Experimental data indicate that the ventilation volume does not correlate well with the peak ventilation pressure, and that the provider cannot correctly estimate the volume given when only pressure is displayed as feedback [3]. Moreover, a recent study has shown the value of end tidal carbon dioxide (EtCO2) as a useful measure of lung aeration and pulmonary blood flow [4]. Assessment of the heart rate (HR) is essential during neonatal resuscitation, as an indicator of effective ventilation and immediate feedback to the provider [5]. However, previous studies have shown that common methods for assessing HR are inaccurate, such as palpation of the umbilical cord and auscultation of the chest [6]. ECG seems to be the most reliable and accurate technology for immediate monitoring of the newborn HR and continuous feedback during resuscitation [7, 8]. Nevertheless, this technology is not commonly available in the delivery rooms. Monitors to guide ventilations are rarely available. A prototype newborn resuscitation monitor (NRM) (Laerdal Global Health, Stavanger, Norway) has been developed as a research tool with the purpose to study delivery room resuscitation and explore possibilities for monitoring resuscitation. However, there was a need to demonstrate the abilities of the monitor on live subjects in a controlled setting prior to the clinical studies on newborns. The NRM was constructed to record ventilation pressure and volume, hence compliance, as well as EtCO2 and HR during varying ventilation rates, inflating volumes and dead space. The monitor complies with current medical device technical standards, is CE-marked and approved for patient use. Piglets were chosen for this purpose, partly because their required tidal volume approximates 6 ml/kg, resembling the human neonate [9]. The objective of this preclinical study was to demonstrate the functions of the NRM to record physiological changes in live subjects through exploring how (1) EtCO2 measurements were affected by changes in ventilation rate, inflated volume, and different dead space (simulating a face mask ventilation situation) and (2) lavage of surfactant affected the compliance and HR (simulating different clinical conditions in an infant).

Methods

The study was performed at SEARCH (Sandnes Educational and Research Centre Høyland) animal laboratory in Sandnes, Norway. The protocol was approved by The Research Animal Council for the Rogaland Region. We included five piglets with postnatal age 1.5 (n = 1) and 2.5 weeks (n = 4), with median weight 5.1 kg (range 3.7–5.2 kg). As the monitor was CE marked and approved for use in humans, we chose a small sample size to test the functions and the signal data management before the clinical studies. The piglets were handled in accordance with the European Guidelines for Use of Experimental Animals by certified Federation of European Laboratory Animal Science Association category C researchers (FELASA C).

Measurements

The NRM has sensors capable of synchronously recording HR, airway pressure, flow, volume, and EtCO2. Flow and volume are measured using a flow sensor (MIM Gmbh, Krugzell, Germany) placed between the endotracheal tube (ETT) and the resuscitator bag. The flow sensor and its electronics are the same as used in the Florian respiration monitor (Acutronic AG, Zürich, Switzerland) that has previously been used in similar studies [10, 11]. The sensor has negligible flow resistance and dead space (1 ml), and measures airflow using hot wire anemometer technology. In addition, an airway adapter connecting two plastic tubes to the main cabinet is placed between the ETT and the bag. The first tube draws a sample of exhaled air (50 ml/min) for side stream CO2 measurement (ISA CO2 sensor, Masimo/PhaseIn AB,Danderyd, Sweden) (Fig. 1). The second tube measures airway pressure using a piezoresistive pressure sensor (MPXV5010, Freescale Semiconductor Inc, Austin, Tx).
Fig. 1

Recordings of ventilation and ECG signal data: examples of manual baseline ventilations and ECG from piglet B

Recordings of ventilation and ECG signal data: examples of manual baseline ventilations and ECG from piglet B HR is calculated from ECG measured by dry electrodes: Two stainless steel discs are mounted on each side of a flexible arch that is placed gently over thorax or abdomen of the subject (Fig. 2). The pressure exerted by the flexible arch is comparable to that of a stethoscope used for auscultation. The ECG is recorded and HR calculated using an proprietary algorithm based on a zero crossing count algorithm [12 ]. The dry-electrode ECG technology and HR algorithm was validated (by the manufacturer) on adult human subjects (n = 30, 396 data points) against HR from a Philips HeartStart MRx (Philips Healthcare, Andover, MA) and showed a rate equivalence within 1 bpm in 84.3% of the data points, 99.8% within 4 bpm. The HR sensor is connected to the main cabinet through a cable, and the HR is displayed on the monitor.
Fig. 2

The newborn resuscitation monitor

The newborn resuscitation monitor The monitor thus uses reliable and proven technology put together into one unit. Safety for patient and operator is ensured by complying with current medical device technical standards, and is tested and approved for patient use according to the EU Medical Device Directive. During the experiment, the HR sensor was placed over the abdomen of the piglet, and a Laerdal Neonatal Resuscitator size 1 (240 ml) was used for manual ventilations. Data from all sensors were stored in an internal memory card, downloaded to a computer through an USB connection, and analyzed using Matlab (MathWorks, Natick, MA).

Protocol

The piglets were anesthetized in accordance with Norwegian guidelines for animal experiments. An ETT with cuff was inserted into the trachea and secured, and the HR sensor was placed over the abdomen. Anesthetic maintenance was provided. No interventions took place before the piglets were fully anesthetized. An intra-arterial line was established in one femoral artery to enable measurements of arterial pressures. The instrumented resuscitator (Fig. 2) was connected to the ETT, and baseline patterns and measurements of the inspiratory and expiratory pressure, volume, flow and EtCO2 at rest were recorded in the NRM. Each piglet underwent several experiments, and was euthanized with pentobarbital when mean arterial pressure decreased to ≤15 mbar. Three experiments are described in this report; (1) measurements of EtCO2 with various tidal volumes, ventilation rates and dead space, and (2) blood was drawn from the arterial catheter for blood gas analysis and comparison of PaCO2 with EtCO2 every 4th minute. EtCO2 was recorded continuously by the monitor. (3) Lavage of surfactant. See Table 1 for description of the timeline for each piglet.
Table 1

Overview: overview of experiments and ventilations performed on each piglet

Type of experimentDuration (min)Dead space (ml)Volume (ml/kg)Rate (vpm)
EtCO2 measurements401040
4201040
4101040
4301040
101020
101040
10640
101040
101540
1201040
1201020
1201040
120640
1201540
Pause201040
Surfactant deficiencyLavage of the lungs through the ETT with 10 ml NaCl ten times in each series
Repeated lavage as long as mean arterial pressure >15
Ventilations with 100% O2, volume: 10 ml/kg, 40 ventilations per minute

In the first part, EtCO2 was measured with different duration, dead space, volume and rates of inflation. Thereafter surfactant were lavaged out with 10 ml NaCl × 10 in each series and compliance and HR was measured after each series

Overview: overview of experiments and ventilations performed on each piglet In the first part, EtCO2 was measured with different duration, dead space, volume and rates of inflation. Thereafter surfactant were lavaged out with 10 ml NaCl × 10 in each series and compliance and HR was measured after each series Ventilations in room air were performed with different tidal volumes and ventilation rates (Table 2). Different dead space (0–30 ml) were provided and controlled by an expandable syringe connected between the tube and airway sensors. Each combination of volume and rate were kept for 1 min, whereas each dead space series lasted 4 min. EtCO2 was recorded continuously by the monitor.
Table 2

Experiment with different dead space and volumes: levels of EtCO2 for different tidal volumes and dead space, and then levels of EtCO2 with different rates of inflation per minute (ipm) and dead space

Volume (ml/kg)EtCO2 (mmHg) 0 ml dead spacea EtCO2 (mmHg) 20 ml dead spacea p value
67.0 (5.3–7.8)5.2 (4.3–6.0)0.066
105.9 (4.8–7.2)5.0 (4.2–6.2)0.068
154.6 (4.1–5.9)5.1 (3.6–5.5)0.465

aResults are given as median (range)

Experiment with different dead space and volumes: levels of EtCO2 for different tidal volumes and dead space, and then levels of EtCO2 with different rates of inflation per minute (ipm) and dead space aResults are given as median (range) During the surfactant deficiency experiment, the piglets were ventilated using 100% Oxygen, 40 ventilations/min and a volume of 10 ml/kg. We measured dynamic compliance [total expired volume/(peak inspiratory pressure − positive end expiratory pressure)]. Surfactant was lavaged from the piglet’s lungs with 10 ml saline ten times in each series (=100 ml in one series). Between each series of lavage, the piglets were ventilated for 3 min to allow steady state to occur (i.e. HR, ventilation pressure, and volume stabilizing at a certain level). The HR was registered continuously during the whole experiment. The series were repeated until the mean arterial pressure was ≤15 mbar.

Data analysis

Difference between groups were analyzed by Wilkoxon signed rank test and results given as median (range). Changes in compliance during surfactant lavage were analyzed using a linear mixed model. A two sided p value of <0.05 was considered to be statistically significant. The levels of EtCO2 and PaCO2 and the difference in compliance was compared between groups by Wilkoxon signed rank test. A p value of <0.05 was considered to be statistically significant. Data were analyzed using the IBM-SPSS statistical package (IBM SPSS Statistics for Windows, Version 22.0. Armonk, NY: IBM Corp), and figures were made using Matlab R2014a (MathWorks, Natick, MA).

Results

Data can be viewed in detail in availability of data and materials. One piglet was excluded in both experiments due to excessive leakage around the endotracheal tube (ETT). The ventilation and HR signals were stored in the monitor correctly aligned in time, and data from each ventilation could be analyzed individually (Fig. 1).

EtCO2 readings with different dead space, volumes and ventilation rates

Different clinical scenarios were simulated and Table 2 shows the levels of EtCO2 during different tidal volumes (inflated volumes) (6–15 ml/kg), dead space (different size of mask/tube in airway) (0 and 20 ml) and ventilation rates (the rate of inflations given) (20 and 40 ventilations per minute). There was a tendency of lower EtCO2 levels with increasing dead space, i.e. 0 ml versus 20 ml, for the two smallest tidal volumes, i.e. 6 and 10 ml/kg (p = 0.066 and p = 0.068, respectively). Ventilation rates of 20 compared to 40 ventilations per minute did not influence the levels of EtCO2 with 0 and 20 ml dead space. EtCO2 and PaCO2 were comparable and stable with dead spaces of 10 and 20 ml However, at 30 ml dead space there was a tendency for lower EtCO2 compared to PaCO2 (Table 3).
Table 3

Dynamic compliance and HR before and after lavage of surfactant from the lungs

PigletDynamic compliance before lavage (ml/mbar)HR before lavage (bpm)Number of lavage series (10 × 10 ml) performedDynamic compliance after lavage (ml/mbar)HR after lavage (bpm)
B (4.7 kg)6.614993.5138
C (5.2 kg)4.3144132.5143
D (5.2 kg)2.7136121.5193
E (3.7 kg)8.013232.0132

Measurements from before lavage the first series (10 times 10 ml saline) and after the last lavage series performed

Dynamic compliance and HR before and after lavage of surfactant from the lungs Measurements from before lavage the first series (10 times 10 ml saline) and after the last lavage series performed

Surfactant lavage

The dynamic compliance was measured in each piglet before the lavage of surfactant. The dynamic compliance and HR before and after the lavage series for each piglet are shown in Table 4. There was an average decrease in compliance by 3% after each lavage series (p = 0.004) (Fig. 3, panel a). The piglets tolerated 3–13 series of lavage before the mean arterial pressure decreased to ≤15 mbar. Piglet E with the lowest body weight tolerated three lavage series until mean arterial pressure was ≤15 mbar. ECG was continuously recorded and successfully stored during all the experiments. The HR remained stable throughout the lavage (Fig. 3, panel b).
Table 4

Levels of EtCO2 and PaCO2 measured every 4 min with different dead space

Dead space (ml)EtCO2 (mmHg)a PaCO2 (mmHg)a p value
07.1 (5.0–9.7)7.3 (5.0–9.2)0.502
106.7 (5.2–8.0)6.7 (5.0–7.7)0.724
206.7 (5.0–7.7)6.3 (4.9–8.4)0.141
305.0 (3.9–5.6)7.3 (5.4–8.4)0.075

Ventilation volume was 10 ml/kg and ventilation rate was 40 ventilations per minute

aResults are given as median (range)

Fig. 3

Lavage of surfactant: measured HR and dynamic compliance over time for piglets B (green line), C (turquoise line), D (blue line) and E (red line) during lavage of surfactant with 10 × 10 ml of saline. a shows the dynamic compliance measured at the corresponding times. b shows HR measured after each series of 10 × 10 ml saline lavage

Levels of EtCO2 and PaCO2 measured every 4 min with different dead space Ventilation volume was 10 ml/kg and ventilation rate was 40 ventilations per minute aResults are given as median (range) Lavage of surfactant: measured HR and dynamic compliance over time for piglets B (green line), C (turquoise line), D (blue line) and E (red line) during lavage of surfactant with 10 × 10 ml of saline. a shows the dynamic compliance measured at the corresponding times. b shows HR measured after each series of 10 × 10 ml saline lavage

Discussion

The monitor has the ability to continuously display HR using dry electrode ECG technology, to measure tidal volume, pressure and EtCO2 and to store the results for later analysis. The planned clinical studies will hopefully provide new knowledge on neonatal transition and resuscitation; about ventilation pressure and volume and thereby compliance, as well as mask leakage, EtCO2, and HR during face mask ventilation. The NRM will be used to study if measurements and feedback to the provider will facilitate more objective assessments and thus improve decision making and quality of care during newborn resuscitation. Accuracy level of volumes and pressures registered during the controlled ventilations was in accordance with our expectations. We further observed a steady and significant decrease in compliance when surfactant was lavaged from the piglets lungs. The continuous fall in compliance of mean 3% for each lavage, underlines the monitors ability to detect minute changes in volume and pressure over time. This observed ability of the NRM to register physiological signal data is reassuring. Previous findings suggest that EtCO2 levels can indicate lung aeration, but is not an accurate measure of PaCO2 [4]. Our findings were in line with these previous findings although we had a small number of measurements. It seemed EtCO2 was varying according to dilution of expired gas in terms of increased dead space and change in tidal volumes. This, as well as the delays in side stream technology, is important to consider when moving to clinical studies, and suggests EtCO2 may only be useful as an indication of presence or change in aeration in the setting of bag/mask ventilation. The dry electrode ECG sensor detected satisfactory ECG signals throughout the study. In the clinical setting, the skin of newborns is likely to be moister which would further increase the dry electrodes’ capability of detecting ECG signals. Importantly, HR is the recommended indicator of successful ventilations [13], and the dry electrode ECG technique can provide valuable real-time feedback to the provider. Validation to conventional ECG was satisfactory. However, it was performed on adults and it precision and accuracy in newborns is yet to be studied.

Conclusions

The prototype NRM was capable of continuously displaying HR and detecting inflicted changes in ventilation and compliance of piglets. It can record parameters such as given volume and pressure and the responses (changes in HR and EtCO2). This monitor may facilitate research during neonatal transition and resuscitation that can provide valuable new insight.
  11 in total

1.  Neonatal resuscitation: in pursuit of evidence gaps in knowledge.

Authors:  Jeffrey Perlman; John Kattwinkel; Jonathan Wyllie; Ruth Guinsburg; Sithembiso Velaphi
Journal:  Resuscitation       Date:  2012-01-13       Impact factor: 5.262

2.  Part 11: Neonatal resuscitation: 2010 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science With Treatment Recommendations.

Authors:  Jeffrey M Perlman; Jonathan Wyllie; John Kattwinkel; Dianne L Atkins; Leon Chameides; Jay P Goldsmith; Ruth Guinsburg; Mary Fran Hazinski; Colin Morley; Sam Richmond; Wendy M Simon; Nalini Singhal; Edgardo Szyld; Masanori Tamura; Sithembiso Velaphi
Journal:  Circulation       Date:  2010-10-19       Impact factor: 29.690

Review 3.  Part 13: Neonatal Resuscitation: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care.

Authors:  Myra H Wyckoff; Khalid Aziz; Marilyn B Escobedo; Vishal S Kapadia; John Kattwinkel; Jeffrey M Perlman; Wendy M Simon; Gary M Weiner; Jeanette G Zaichkin
Journal:  Circulation       Date:  2015-11-03       Impact factor: 29.690

4.  Responding to compliance changes in a lung model during manual ventilation: perhaps volume, rather than pressure, should be displayed.

Authors:  John Kattwinkel; Corrine Stewart; Brian Walsh; Matthew Gurka; Alix Paget-Brown
Journal:  Pediatrics       Date:  2009-02-02       Impact factor: 7.124

5.  Breathing patterns in preterm and term infants immediately after birth.

Authors:  Arjan B te Pas; Connie Wong; C Omar F Kamlin; Jennifer A Dawson; Colin J Morley; Peter G Davis
Journal:  Pediatr Res       Date:  2009-03       Impact factor: 3.756

6.  Pulse oximetry measures a lower heart rate at birth compared with electrocardiography.

Authors:  Jeroen J van Vonderen; Stuart B Hooper; Jacco K Kroese; Arno A W Roest; Ilona C Narayen; Erik W van Zwet; Arjan B te Pas
Journal:  J Pediatr       Date:  2014-10-14       Impact factor: 4.406

7.  Accuracy of clinical assessment of infant heart rate in the delivery room.

Authors:  C Omar F Kamlin; Colm P F O'Donnell; Neil J Everest; Peter G Davis; Colin J Morley
Journal:  Resuscitation       Date:  2006-09-20       Impact factor: 5.262

Review 8.  Volume-targeted versus pressure-limited ventilation in the neonate.

Authors:  N McCallion; P G Davis; C J Morley
Journal:  Cochrane Database Syst Rev       Date:  2005-07-20

Review 9.  Every Newborn: progress, priorities, and potential beyond survival.

Authors:  Joy E Lawn; Hannah Blencowe; Shefali Oza; Danzhen You; Anne C C Lee; Peter Waiswa; Marek Lalli; Zulfiqar Bhutta; Aluisio J D Barros; Parul Christian; Colin Mathers; Simon N Cousens
Journal:  Lancet       Date:  2014-05-19       Impact factor: 79.321

10.  Expired CO2 levels indicate degree of lung aeration at birth.

Authors:  Stuart B Hooper; Andreas Fouras; Melissa L Siew; Megan J Wallace; Marcus J Kitchen; Arjan B te Pas; Claus Klingenberg; Robert A Lewis; Peter G Davis; Colin J Morley; Georg M Schmölzer
Journal:  PLoS One       Date:  2013-08-12       Impact factor: 3.240

View more
  3 in total

1.  Predictors for expired CO2 in neonatal bag-mask ventilation at birth: observational study.

Authors:  Kari Holte; Hege Langli Ersdal; Joar Eilevstjønn; Monica Thallinger; Jørgen Linde; Claus Klingenberg; Rene Holst; Samwel Bayo; Hussein Kidanto; Ketil Stordal
Journal:  BMJ Paediatr Open       Date:  2019-09-26

2.  Delivery Room ST Segment Analysis to Predict Short Term Outcomes in Near-Term and Term Newborns.

Authors:  Jørgen Linde; Anne Lee Solevåg; Joar Eilevstjønn; Ladislaus Blacy; Hussein Kidanto; Hege Ersdal; Claus Klingenberg
Journal:  Children (Basel)       Date:  2022-01-03

3.  A Randomised Controlled Study of Low-Dose High-Frequency In-Situ Simulation Training to Improve Newborn Resuscitation.

Authors:  Joanna Haynes; Siren Rettedal; Jeffrey Perlman; Hege Ersdal
Journal:  Children (Basel)       Date:  2021-12-02
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.