Oxygen-Sensing Paint-On Bandage: Calibration of a Novel Approach in Tissue Perfusion Assessment.

Pieter G L Koolen1,2, Zongxi Li1,2, Emmanuel Roussakis1,2, Marek A Paul1,2, Ahmed M S Ibrahim1,2, Robina Matyal1,2, Thomas Huang1,2, Conor L Evans1,2, Samuel J Lin1,2.   

Abstract

BACKGROUND: Knowledge of tissue oxygenation status is fundamental in the prevention of postoperative flap failure. Recently, the authors introduced a novel oxygen-sensing paint-on bandage that incorporated an oxygen-sensing porphyrin with a commercially available liquid bandage matrix. In this study, the authors extend validation of their oxygen-sensing bandage by comparing it to the use of near-infrared tissue oximetry in addition to Clark electrode measurements.
METHODS: The oxygen-sensing paint-on bandage was applied to the left hind limb in a rodent model. Simultaneously, a near-infrared imaging device and Clark electrode were attached to the right and left hind limbs, respectively. Tissue oxygenation was measured under normal, ischemic (aortic ligation), and reperfused conditions.
RESULTS: On average, the oxygen-sensing paint-on bandage measured a decrease in transdermal oxygenation from 85.2 mmHg to 64.1 mmHg upon aortic ligation. The oxygen-sensing dye restored at 81.2 mmHg after unclamping. Responses in both control groups demonstrated a similar trend. Physiologic changes from normal to ischemic and reperfused conditions were statistically significantly different in all three techniques (p < 0.001).
CONCLUSIONS: The authors' newly developed oxygen-sensing paint-on bandage exhibits a comparable trend in oxygenation recordings in a rat model similar to conventional oxygenation assessment techniques. This technique could potentially prove to be a valuable tool in the routine clinical management of flaps following free tissue transfer. Incorporating oxygen-sensing capabilities into a simple wound dressing material has the added benefit of providing both wound protection and constant wound oxygenation assessment.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28654595     DOI: 10.1097/PRS.0000000000003421

Source DB:  PubMed          Journal:  Plast Reconstr Surg        ISSN: 0032-1052            Impact factor:   4.730


  3 in total

Review 1.  Flap Monitoring Using Transcutaneous Oxygen or Carbon Dioxide Measurements.

Authors:  Sameer H Halani; Austin S Hembd; Xingchen Li; Ben Kirby; Courtney C Beard; Nicholas T Haddock; Thomas M Suszynski
Journal:  J Hand Microsurg       Date:  2020-11-16

Review 2.  Imaging of oxygen and hypoxia in cell and tissue samples.

Authors:  Dmitri B Papkovsky; Ruslan I Dmitriev
Journal:  Cell Mol Life Sci       Date:  2018-05-14       Impact factor: 9.261

3.  Sensing, monitoring, and release of therapeutics: the translational journey of next generation bandages.

Authors:  Zongxi Li; Haley Marks; Conor Evans; Gabriela Apiou-Sbirlea
Journal:  J Biomed Opt       Date:  2018-12       Impact factor: 3.170