| Literature DB >> 28652851 |
Abstract
Oxygen (O2) is an essential element for aerobic respiration. Atmospheric concentration of O2 is approximately 21%. Mammalian cells, however, are generally adapted to O2 levels much lower than atmospheric conditions. The pericellular levels of O2 must also be maintained within a fairly narrow range to meet the demands of cells. This applies equally to cells in vivo and cells in primary cultures. There has been growing interest in the performance of cell culture experiments under various O2 levels to study molecular and cellular responses. To this end, a range of technologies (e.g. gas-permeable technology) and instruments (e.g. gas-tight boxes and gas-controlled incubators) have been developed. It should be noted, however, that some of these have limitations and they are still undergoing refinement. Nevertheless, better results should be possible when technical concerns are taken into account. This paper aims to review various aspects of O2 level adjustment in primary cell cultures, regulation of pericellular O2 gradients and possible effects of the cell culture medium.Entities:
Keywords: O2 level adjustment; culture medium effects; pericellular O2 gradients regulation; primary cell cultures
Year: 2017 PMID: 28652851 PMCID: PMC5464680 DOI: 10.1515/intox-2016-0011
Source DB: PubMed Journal: Interdiscip Toxicol ISSN: 1337-6853
An overview of the commercially available gas-tight/flush boxes with associated product information.
| Commercial Name | Model | Size | Supplier | |
|---|---|---|---|---|
| Hypoxia/modular incubator chamber, Flush box | MIC | 12″D × 4.7″H | STEMCELL Technologies, Inc., | |
| Hypoxia chamber for cell culture | C-Chamber | 14″W × 12″D × 5.25″H | BioSpherix, Ltd. | |
| O2 control cabinet for | Model 1 (1-shelf chamber) | 16″W × 15″D × 9.5″H | Coy Laboratory Products, Inc. |
External dimensions.
Specifications and product information for gas-tight/flush box gas controllers.
| Model | Function | Control range | Accuracy | Resolution (precision) | Supplier |
| ProOx P110 | Control O2 in any semi-sealable chamber | 0.1–99.9% O2 | ±1% at constant temperature/pressure ±2% full scale over operating temperature range | 0.1% | BioSpherix, Ltd. |
| ProOx C21 | Control O2 and CO2 in any chamber | 0.1–99.9% O2, 0.1–20% CO2 | O2: ±1% at constant temperature/pressure ±2% over entire temperature range. CO2: .1% or 5% of measurement, whichever is greater | 0.1% | BioSpherix, Ltd. |
| ProCO2 P120 | Control CO2 in any chamber | 0.1–20% CO2 | ±.3% (@0%) to ±.7% (@20%) at 25°C and 1013hPa | 0.1% | BioSpherix, Ltd. |
| OxyCycler C42 | Control O2 and CO2 in multiple chambers | O2: 0.1–99.9%, CO2: 0.1–20% | O2: ±1% at constant temperature and pressure ±2% at entire temperature range. | 0.1% | BioSpherix, Ltd. |
| O2 controller | Control O2 and N2/air (or a mixture of CO2 and N2/air) | 0–100% or 0–60% O2 atmospheric | - | 0.1% | Coy Laboratory Products, Inc. |
| CO2 controller | Control CO2 and N2/air | 0–19.9% in 0.1% increments | <+/− [0.02% CO2 +2% of reading] | 0.1% | Coy Laboratory Products, Inc. |
| Single flow meter | Control the gas flow of premixed gases | Calibrated for 0.1–10 LPM | - | - | STEMCELL Technologies, Inc. |
| Dual flow meter | Used to mix gases in addition to controlling the gas flow rate | Calibrated for 0.1–1.0 and 0.1–10 LPM | - | - | STEMCELL Technologies, Inc. |
LPM: Litres per minute
The dual flow meter should be used in combination with an O2 detector.
Blank fields represent information lacking from the supplier.