Abstract

INTRODUCTION: The XN-550 is a new, automated, compact, haematology analyser designed to generate a full blood count with a standard five-part white blood cell differential and an immature granulocyte count, as well as an optional reticulocyte and optical platelet (PLT) counts. The aim of the study was to evaluate the performance of the XN-550 and compare it to the established XN-20 system.
METHODS: We evaluated the basic parameter and special measurement channels of the XN-550, using the XN-20 (which has a similar operating system), as a reference analyser. Precision, carry-over and throughput evaluations were performed. In addition, a total of 202 samples including normal controls and various pathological samples were studied for comparability.
RESULTS: Good correlations with the reference analyser were obtained for all parameters except basophils. The XN-550 offers impedance and optical PLT counts and the latter showed a better correlation and less scatter than the impedance count and was comparable to the XN-20 fluorescent count at PLT counts ≤40×109 /L. Precision was good, and no significant carry-over was detected.
CONCLUSIONS: The XN-550 was simple and easy to use, while maintaining the good diagnostic sensitivity seen with high-range systems such as the XN-20, making this compact device suitable for near-patient services and smaller satellite laboratories.
© 2017 John Wiley & Sons Ltd.

Entities:  

Keywords:  Sysmex; general hematology; laboratory automation; leukocytes; platelets; reticulocytes

Mesh:

Year:  2017        PMID: 28649771     DOI: 10.1111/ijlh.12701

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  3 in total

1.  Biomarkers of Exposure and Potential Harm in Exclusive Users of Nicotine Pouches and Current, Former, and Never Smokers: Protocol for a Cross-sectional Clinical Study.

Authors:  David Azzopardi; Linsey Ellen Haswell; Justin Frosina; Michael McEwan; Nathan Gale; Jesse Thissen; Filimon Meichanetzidis; George Hardie
Journal:  JMIR Res Protoc       Date:  2022-10-06

2.  A comparison of the analysis of 3 types of body fluids using the XN-350 hematology analyzer versus light microscopy assessment.

Authors:  Jiwon Lee; Younggeun Cho; Han-Sung Kim; Hee Jung Kang; Miyoung Kim; Young Kyung Lee
Journal:  Medicine (Baltimore)       Date:  2021-03-19       Impact factor: 1.817

3.  The Sysmex XN-L (XN-350) hematology analyzer offers a compact solution for laboratories in niche diagnostics.

Authors:  Tania A Khartabil; Magda M de Frankrijker; Yolanda B de Rijke; Henk Russcher
Journal:  Int J Lab Hematol       Date:  2020-09-19       Impact factor: 2.877

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.