Literature DB >> 28643228

A urologic stethoscope? Urologist performed sonography using a pocket-size ultrasound device in the point-of-care setting.

Arnon Lavi1, Sharon Tzemah2, Anan Hussein2, Ibrahim Bishara2, Nikolay Shcherbakov2, Genady Zelichenko2, Alon Mashiah2, Michael Gross2, Leonid Cherbinski3, Ziv Neeman3, Michael Cohen2.   

Abstract

PURPOSE: Ultrasound is commonly used in urology. Technical advances with reduced size and cost led to diffusion of small ultrasound devices to many clinical settings. Even so, most ultrasound studies are performed by non-urologists. We aimed to evaluate the utility of a pocket-size ultrasound device (Vscan™ GE Healthcare) and the quality of urologist performed study.
METHODS: Three consecutive studies were performed: (1) a urologist using the pocket ultrasound, (2) a sonographist using the pocket ultrasound, and (3) a sonographist using a standard ultrasound device. Thirty-six patients were evaluated with a basic urologic ultrasound study. An excepted deviation between studies was preset for numeric parameters and t test performed. Ordinal parameters were analyzed using Cohen's kappa coefficient.
RESULTS: Kidney length, renal pelvis length, renal cyst diameter, post-void residual and prostate volume (transabdominal) differences were found to be insignificant when comparing a urologist pocket ultrasound study to a sonographist standard ultrasound study (P = 0.15; P = 0.21; P = 0.81; P = 0.32; P = 0.07, respectively). Hydronpehrosis evaluation (none, mild, moderate and severe) and the presence of ureteral jet signs conferred a high inter-observer agreement when comparing the above studies using the Cohen's kappa coefficient (K = 0.63; K = 0.62, respectively).
CONCLUSIONS: Urologist performed pocket ultrasound study is valid in evaluating the upper and lower urinary tract and is practical in many clinical scenarios. The urologic stethoscope is now becoming a reality within reach.

Entities:  

Keywords:  Point-of-care ultrasound scanning (POCUS); Urinary tract evaluation; Urologic stethoscope; Urologist performed ultrasound

Mesh:

Year:  2017        PMID: 28643228     DOI: 10.1007/s11255-017-1641-8

Source DB:  PubMed          Journal:  Int Urol Nephrol        ISSN: 0301-1623            Impact factor:   2.370


  18 in total

1.  Our new stethoscope in the emergency department: handheld ultrasound.

Authors:  Figen Coşkun; Emine Akıncı; Mehmet Ali Ceyhan; Havva Sahin Kavaklı
Journal:  Ulus Travma Acil Cerrahi Derg       Date:  2011-11

Review 2.  Is it time to include point-of-care ultrasound in general surgery training? A review to stimulate discussion.

Authors:  Maximilian Mollenkopf; Noel Tait
Journal:  ANZ J Surg       Date:  2013-08-26       Impact factor: 1.872

3.  The diagnostic accuracy of pocket-size cardiac ultrasound performed by unselected residents with minimal training.

Authors:  Vidar Ruddox; Thomas Muri Stokke; Thor Edvardsen; Jøran Hjelmesæth; Erlend Aune; Morten Bækkevar; Ingvild B Norum; Jan Erik Otterstad
Journal:  Int J Cardiovasc Imaging       Date:  2013-08-23       Impact factor: 2.357

4.  Application of office ultrasound in the management of the spinal cord injury patient.

Authors:  D R Bodner; M Witcher; M I Resnick
Journal:  J Urol       Date:  1990-05       Impact factor: 7.450

5.  Modification of the hand-held Vscan ultrasound and verification of its performance for transvaginal applications.

Authors:  J M Troyano Luque; O Ferrer-Roca; M J Barco-Marcellán; R Sabatel López; T Pérez-Medina; F R Pérez-Lopez
Journal:  Ultrasonics       Date:  2012-08-14       Impact factor: 2.890

6.  Diagnostic ultrasound in the urologist's office.

Authors:  S A Leonard; R Thomas
Journal:  Urology       Date:  1987-06       Impact factor: 2.649

7.  Bedside ultrasound: a useful tool for the on-call urologist?

Authors:  R S Surange; N S Jeygopal; S D Chowdhury; N K Sharma
Journal:  Int Urol Nephrol       Date:  2001       Impact factor: 2.370

8.  Comparison of a pocket-size ultrasound device with a premium ultrasound machine: diagnostic value and time required in bedside ultrasound examination.

Authors:  Konrad Friedrich Stock; Bettina Klein; Dominik Steubl; Christian Lersch; Uwe Heemann; Stefan Wagenpfeil; Florian Eyer; Dir-Andre Clevert
Journal:  Abdom Imaging       Date:  2015-10

9.  Impact of point-of-care ultrasound training on surgical residents' confidence.

Authors:  Meera Kotagal; Elina Quiroga; Benjamin J Ruffatto; Adeyinka A Adedipe; Brandon H Backlund; Robert Nathan; Anthony Roche; Dana Sajed; Sachita Shah
Journal:  J Surg Educ       Date:  2015-04-21       Impact factor: 2.891

10.  EchoScopy in scanning abdominal diseases: initial clinical experience.

Authors:  A P Barreiros; X W Cui; A Ignee; C De Molo; C Pirri; C F Dietrich
Journal:  Z Gastroenterol       Date:  2014-03-12       Impact factor: 2.000

View more
  5 in total

1.  Assessment of the renal collecting system using a pocket-sized ultrasound device.

Authors:  Toru Kameda; Kumiko Uebayashi; Kazuko Wagai; Fukiko Kawai; Nobuyuki Taniguchi
Journal:  J Med Ultrason (2001)       Date:  2018-05-02       Impact factor: 1.314

2.  Miniaturised ultrasound evaluation at the bedside.

Authors:  Katharina Hollerieth; Minh-Truc Vo-Cong; Stephanie Preuss; Stephan Kemmner; Konrad Friedrich Stock
Journal:  World J Urol       Date:  2022-05-18       Impact factor: 4.226

3.  Can Diaphragm Dysfunction Be Reliably Evaluated with Pocket-Sized Ultrasound Devices in Intensive Care Unit?

Authors:  Gul Gursel; Kamil Inci; Zenfira Alasgarova
Journal:  Crit Care Res Pract       Date:  2018-04-01

4.  Development and evaluation of automated ultrasonographic detection of bladder diameter for estimation of bladder urine volume.

Authors:  Masaru Matsumoto; Takuya Tsutaoka; Koichi Yabunaka; Mayumi Handa; Mikako Yoshida; Gojiro Nakagami; Hiromi Sanada
Journal:  PLoS One       Date:  2019-09-05       Impact factor: 3.240

5.  Integrating Basic and Clinical Sciences Using Point-of-Care Renal Ultrasound for Preclerkship Education.

Authors:  Stephen Alerhand; April Choi; Ilya Ostrovsky; Sophia Chen; Christine Ramdin; Maria Laboy; Sangeeta Lamba
Journal:  MedEdPORTAL       Date:  2020-12-09
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.