Equal Primary Fixation of Resurfacing Stem, but Inferior Cup Fixation With Anterolateral vs Posterior Surgical Approach. A 2-Year Blinded Randomized Radiostereometric and Dual-energy X-Ray Absorptiometry Study of Metal-on-Metal Hip Resurfacing Arthroplasty.

Mette H Hjorth1, Nina D Lorenzen1, Kjeld Søballe1, Stig S Jakobsen1, Maiken Stilling1.   

Abstract

BACKGROUND: The anterolateral (AntLat) surgical approach may spare the blood supply to the femoral head and improve the accuracy of cup positioning in metal-on-metal hip resurfacing arthroplasty. Thereby, potentially lessen complications such as avascular head necrosis, femoral neck narrowing and fracture, improve implant fixation, and lessen periprosthetic bone mineral density (BMD) loss.
METHODS: Between November 2008 and January 2012, a randomized clinical trial was performed at Aarhus University Hospital. A total of 49 patients (28 males) were allocated to metal-on-metal hip resurfacing arthroplasty by the AntLat (n = 25) or the posterior (Post; n = 24) surgical approach. Patients were followed with radiostereometric analysis, measurements of periprosthetic BMD, clinical outcome scores of Harris hip score and visual analogue scale, serum metal ions, and conventional radiographs.
RESULTS: At 3 months, cups in the AntLat group had higher total translations of mean 1.00 ± 0.70 mm vs mean 0.64 ± 0.45 mm in the post group (P = .04), and higher total rotations of mean 2.44° ± 1.36° vs mean 1.39° ± 1.17° in the Post group (P = .002). All migrations of cup and stem were similar at 1 and 2 years postoperative (P > .07). At 1 year, periprosthetic BMD since postoperative at the medial side of the stem was reduced to mean 98.45% ± 8.57% in the AntLat group, and increased to mean 105.57% ± 11.07% in the Post group (P = .02), but measurements were comparable at 2 years (P = .05).
CONCLUSION: Cups inserted by the AntLat approach migrated more until 3 months postoperative. This illustrates a less good primary cup fixation with the AntLat approach; however, all cups were well-fixed after 3 months' follow-up indicating a good secondary fixation.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  clinical outcome scores; dual-energy x-ray absorptiometry; metal-on-metal hip resurfacing arthroplasty; radiostereometric analysis; surgical approach

Mesh:

Substances:

Year:  2017        PMID: 28641972     DOI: 10.1016/j.arth.2017.05.034

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  2 in total

1.  Higher early proximal migration of hemispherical cups with electrochemically applied hydroxyapatite (BoneMaster) on a porous surface compared with porous surface alone: a randomized RSA study with 53 patients.

Authors:  Peter Bo Jørgensen; Henrik Daugaard; Stig Storgaard Jakobsen; Martin Lamm; Kjeld Søballe; Maiken Stilling
Journal:  Acta Orthop       Date:  2019-11-08       Impact factor: 3.717

2.  The correlation between activity level, serum-ion concentrations and pseudotumours in patients with metal-on-metal hip articulations and metal-on-polyethylene total hip articulations.

Authors:  Mette Holm Hjorth; Inger Mechlenburg; Kjeld Soballe; Lone Roemer; Maiken Stilling
Journal:  J Orthop Translat       Date:  2018-12-17       Impact factor: 5.191

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.