Comminuted olecranon fractures: biomechanical testing of locked versus minifragment non-locked plate fixation.

David S Wellman1, Scott M Tucker2, Josh R Baxter2, Nadine C Pardee1, Lionel E Lazaro1, Christopher S Smith1, Dean G Lorich1, David L Helfet3.   

Abstract

INTRODUCTION: Open reduction and internal fixation has long been accepted as optimal treatment for displaced olecranon fractures based on poor results seen with conservative management. With the presence of comminution, tension-band wiring constructs are contraindicated due to tendency to compress through fragments, thereby shortening the articular segment. Therefore, plate fixation is typically employed. Our hypothesis was that in a comminuted fracture model, 2.7 mm reconstruction plating without locking screws will perform equally to 3.5 mm locked plating in terms of fracture displacement and rotation (shear).
MATERIALS AND METHODS: A three-part comminuted olecranon fracture pattern was created in nine matched pairs of cadaveric specimen using an oscillating saw in standardized, reproducible fashion. Each matched pair was then randomized to receive either 2.7 mm reconstruction plating or 3.5 mm proximal ulna locked plating. Random allocation software was used to assign the 2.7 mm plate construct to either the right or left side of each pair with the contralateral receiving the 3.5 mm plate construct. Specimens were cyclically loaded simulating passive range of motion exercises commonly performed during rehabilitation. Displacement and rotation in relation to the long axis of the ulna were measured through motion capture. Fragment gapping and rotation was quantified following 100 cycles at 10 N and again following 100 cycles at 500 N.
RESULTS: No significant differences were detected between the 2.7 and 3.5 mm plates in fracture rotation or gapping following loads at 10 N (0.5° and 0.7°; 0.6 and 1.2 mm; respectively; p > 0.05) or 500 N (2.3° and 1.6°; 3.8 and 3.1 mm; respectively; p > 0.05) loading. Fragment rotation and gapping were positively correlated within each plate construct (R 2 > 0.445; p < 0.05).
CONCLUSIONS: 2.7 mm plating is an alternative to 3.5 mm locked plating with decreased plate prominence without significantly sacrificing displacement and rotational control. This is beneficial in fracture patterns where the traditional dorsal plating does not offer optimal screw trajectory.

Entities:  

Keywords:  Biomechanical study; Cyclic loading; Locking plates; Non-locking plates; Olecranon fractures

Mesh:

Year:  2017        PMID: 28634743     DOI: 10.1007/s00402-017-2735-6

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  3 in total

1.  Mini-fragment plating of olecranon fractures is comparable to precontoured small-fragment plating.

Authors:  Harsh Wadhwa; Yousi A Oquendo; L Henry Goodnough; Malcolm R DeBaun; Julius A Bishop; Michael J Gardner
Journal:  J Orthop       Date:  2022-02-11

2.  Elimination of irreducible intercalary fragment and fixation using locking plate for Mayo type IIB olecranon fracture-outcomes compared with type IIA.

Authors:  Soo Min Cha; Jong Woo Kang; Hyun Dae Shin; Cheol Won Lee
Journal:  Arch Orthop Trauma Surg       Date:  2021-08-19       Impact factor: 2.928

3.  Comparison of clinical outcomes of three internal fixation techniques in the treatment of olecranon fracture: a retrospective clinical study.

Authors:  Hongfei Qi; Zhong Li; Shuai Ji; Ming Li; Yao Lu; Teng Ma; Bing Du; Qiang Huang; Kun Zhang; Yanling Yang
Journal:  BMC Musculoskelet Disord       Date:  2022-06-01       Impact factor: 2.562

  3 in total