Evaluation of automatic dose rate control for flat panel imaging using a spatial frequency domain figure of merit.

M Dehairs1, H Bosmans, W Desmet, N W Marshall.   

Abstract

Current automatic dose rate controls (ADRCs) of dynamic x-ray imaging systems adjust their acquisition parameters in response to changes in patient thickness in order to achieve a constant signal level in the image receptor. This work compares a 3 parameter (3P) ADRC control to a more flexible 5-parameter (5P) method to meet this goal. A phantom composed of 15 composite poly(methyl) methacrylate (PMMA)/aluminium (Al) plates was imaged on a Siemens Artis Q dynamic system using standard 3P and 5P ADRC techniques. Phantom thickness covered a water equivalent thickness (WET) range of 2.5 cm to 37.5 cm. Acquisition parameter settings (tube potential, tube current, pulse length, copper filtration and focus size) and phantom entrance air kerma rate (EAKR) were recorded as the thickness changed. Signal difference to noise ratio (SDNR) was measured using a 0.3 mm iron insert centred in the PMMA stack, positioned at the system isocentre. SDNR was then multiplied by modulation transfer function (MTF) based correction factors for focal spot penumbral blurring and motion blurring, to give a spatial frequency dependent parameter, SDNR(u). These MTF correction factors were evaluated for an object motion of 25 mm s-1 and at a spatial frequency of 1.4 mm-1 in the object plane, typical for cardiac imaging. The figure of merit (FOM) was calculated as SDNR(u)²/EAKR for the two ADRC regimes. Using 5P versus 3P technique showed clear improvements over all thicknesses. Averaged over clinically relevant adult WET values (20 cm-37.5 cm), EAKR was reduced by 13% and 27% for fluoroscopy and acquisition modes, respectively, while the SDNR(u) based FOM increased by 16% and 34% for fluoroscopy and acquisition. In conclusion, the generalized FOM, taking into account the influence of focus size and object motion, showed benefit in terms of image quality and patient dose for the 5-parameter control over 3-parameter method for the ADRC programming of dynamic x-ray imaging systems.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28632501     DOI: 10.1088/1361-6560/aa7a9d

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Tantalum-specific contrast-to-noise ratio or conventional detector dose-driven exposure control in angiography: radiation dose and image quality evaluation in a porcine model.

Authors:  Thomas Werncke; Timo Christian Meine; Jan B Hinrichs; Sabine K Maschke; Lena Sophie Becker; Inga Brüsch; Regina Rumpel; Frank K Wacker; Bernhard C Meyer
Journal:  Eur Radiol Exp       Date:  2022-05-17

2.  Performance of sparse-view CT reconstruction with multi-directional gradient operators.

Authors:  Chia-Jui Hsieh; Shih-Chun Jin; Jyh-Cheng Chen; Chih-Wei Kuo; Ruei-Teng Wang; Woei-Chyn Chu
Journal:  PLoS One       Date:  2019-01-07       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.