Abstract
In 30 patients blood samples were taken from both radial arteries and analysed for PCO2, PO2 and pH. The influence of sampling technique on the accuracy of blood-gas analysis was determined. The 95% confidence limits for PCO2 measurements were +/-0.32 kPa (2.4 mmHg), and for PO2 measurements were +/-1.16 kPa (8.8 mmHg) in the range below 20 kPa, and +/-2.94 kPa (22.0 mmHg) above 20 kPa. For pH, the 95% confidence limits were +/-0.014. Only the PO2 measurements were significantly influenced by the sampling procedure.Entities:
Mesh:
Year: 1978 PMID: 28631
Source DB: PubMed Journal: Acta Anaesthesiol Scand Suppl ISSN: 0515-2720