Safety assessment of the use of ultrasonic energy in the proximity of the recurrent laryngeal nerve in a porcine model.

Karla V Chávez1, Elpidio M Barajas1, Francisco Soroa2, Armando Gamboa-Dominguez3, Samuel Ordóñez1, Juan P Pantoja1, Mauricio Sierra1, David Velázquez-Fernández1, Miguel F Herrera4.   

Abstract

BACKGROUND: Advanced bipolar and ultrasonic energy have demonstrated reduction of operating time and blood loss in thyroidectomy. However, these devices generate heat and thermal dispersion that may damage adjacent structures such as the recurrent laryngeal nerve (RLN). This study was designed to evaluate the safety profile of the Harmonic Focus+® (HF+) device through the evaluation of thermal injury to the RLN using different algorithms of distance and time with state of the art technology.
METHODS: 25 Vietnamese pigs underwent activation of HF+ in the proximity of their RLN. They were divided into 4 groups according to activation distance (3 mm, 2 mm, 1 mm and on the RLN). Time of activation, time between tones of the ultrasonic generator, changes in the electromyographic signal using continuous nerve neuromonitoring, vocal fold mobility assessed by direct laryngoscopy and histological thermal damaged were evaluated.
RESULTS: None of the pigs had loss of signal in the electromyography during the procedure; only one pig had isolated transient decrease in amplitude and one increase in latency. One pig had transient vocal fold paresis in the group with activation on the nerve. Evaluation of the nerves by histology and immunohistochemistry did not show significant changes attributed to thermal injury.
CONCLUSIONS: The use of ultrasonic energy close to the RLN is safe, provided that activation time does not exceed the necessary time to safely transect the tissue.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Energy devices; Harmonic focus; Recurrent laryngeal nerve; Thermal damage; Ultrasonic energy

Mesh:

Year:  2017        PMID: 28622836     DOI: 10.1016/j.amjsurg.2017.04.013

Source DB:  PubMed          Journal:  Am J Surg        ISSN: 0002-9610            Impact factor:   2.565


  7 in total

1.  Lateral thermal spread induced by energy devices: a porcine model to evaluate the influence on the recurrent laryngeal nerve.

Authors:  Masaru Hayami; Masayuki Watanabe; Shinji Mine; Yu Imamura; Akihiko Okamura; Masami Yuda; Kotaro Yamashita; Tasuku Toihata; Yoshiaki Shoji; Naoki Ishizuka
Journal:  Surg Endosc       Date:  2019-03-07       Impact factor: 4.584

2.  Thermal effects of the Thunderbeat device on the recurrent laryngeal nerve during thyroid surgery.

Authors:  Atsumi Tamura; Hidemitsu Tsutsui; Yukiko Yano; Ryoji Ohara; Masae Hoshi; Takeshi Nagai; Toshitaka Nagao; Norihiko Ikeda
Journal:  Surg Today       Date:  2018-07-09       Impact factor: 2.549

3.  Hyperthermal liquid, spray, and smog may be potential risk factors for recurrent laryngeal nerve thermal injury during thyroid surgeries.

Authors:  Xing Yu; Chang Liu; Maoxiao Yan; Weihua Gong; Yong Wang
Journal:  Endocrine       Date:  2020-08-10       Impact factor: 3.633

4.  Clinical observation of end-to-end neuroanastomosis in the treatment of complete injury of the unilateral recurrent laryngeal nerve.

Authors:  Runzhang Wu; Chaojie Zhang; Huiling Wang; Meiliang Li; Shanshan Lei; Jie Zeng; Jie He
Journal:  Gland Surg       Date:  2020-12

5.  Safety Profile of the New Harmonic Focus: Different Emissivity and Temperature Behavior Between the Active and the Inactive Blade.

Authors:  Saverio Latteri; Giuseppe Stella; Anna M Gueli; Stefania Mazzaglia; Valentina Palumbo; Tommaso Guastella
Journal:  Surg Laparosc Endosc Percutan Tech       Date:  2019-10       Impact factor: 1.719

Review 6.  Improving Voice Outcomes After Thyroid Surgery - Review of Safety Parameters for Using Energy-Based Devices Near the Recurrent Laryngeal Nerve.

Authors:  Jia Joanna Wang; Tzu-Yen Huang; Che-Wei Wu; Yi-Chu Lin; Hsin-Yi Tseng; Cheng-Hsin Liu; I-Cheng Lu; Pi-Ying Chang; Hui-Chun Chen; Hsiu-Ya Chen; Gianlorenzo Dionigi; Feng-Yu Chiang; Ling-Feng Wang
Journal:  Front Endocrinol (Lausanne)       Date:  2021-11-24       Impact factor: 5.555

7.  Effects of Intraoperative Neural Tunnel in Protecting Recurrent Laryngeal Nerve: Experiences in Open, Trans Breast, and Transoral Endoscopic Thyroidectomy.

Authors:  Xing Yu; Yujun Li; Chang Liu; Yuancong Jiang; Zhaodi Liu; Qionghua He; Yong Wang; Ping Wang
Journal:  Front Oncol       Date:  2022-02-23       Impact factor: 6.244

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.