Literature DB >> 28604571

HVAD Waveform Analysis as a Noninvasive Marker of Pulmonary Capillary Wedge Pressure: A First Step Toward the Development of a Smart Left Ventricular Assist Device Pump.

Jonathan Grinstein, Daniel Rodgers, Sara Kalantari, Gabriel Sayer, Gene H Kim, Nitasha Sarswat, Sirtaz Adatya, Takeyoshi Ota, Valluvan Jeevanandam, Daniel Burkhoff, Nir Uriel.   

Abstract

Flow waveforms are an important feature of the HVAD left ventricular assist device (LVAD) that provides information about HVAD function and patient hemodynamics. We assessed the properties of one specific aspect of the waveform, the slope of the ventricular filling phase (VFP), and its correlation with pulmonary capillary wedge pressure (PCWP). A total of 101 screenshots from the HVAD monitor and simultaneous hemodynamic measurements were obtained simultaneously during sequential stages of invasive hemodynamic ramp studies. Each screenshot was digitized (IGOR Pro, WaveMetrics Inc., Oswego, OR) and properties of the flow waveforms including instantaneous flow and rate of change of flow were analyzed. Ventricular filling phase slope (VFPS) was calculated for each screenshot and correlated to PCWP. Ventricular filling phase slope was significantly higher in patients with PCWP ≥ 18 mm Hg than in patients with PCWP < 18 mm Hg [6.25 (5.84-7.37) L/min/s vs. 3.27 (2.00-4.69) L/min/s, p ≤ 0.0001]. A VFPS threshold of 5.8 L/min/s predicted a PCWP ≥ 18 mm Hg with a sensitivity of 87% and specificity of 95% (AUC 0.95). Ventricular filling phase slope of the HVAD flow waveform is a novel noninvasive parameter that correlates with PCWP and can discriminate elevated versus normal or low PCWP. Automated reporting of this parameter may help clinical assessment and management of patients supported by an HVAD and may serve as the basis of a smart LVAD pump that can adapt in response to changes in a patient's physiology.

Entities:  

Mesh:

Year:  2018        PMID: 28604571      PMCID: PMC5741440          DOI: 10.1097/MAT.0000000000000604

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  9 in total

1.  Accuracy of the HVAD Pump Flow Estimation Algorithm.

Authors:  Carlos Reyes; Neil Voskoboynikov; Katherine Chorpenning; Jeffrey A LaRose; Michael C Brown; Nathalie J Nunez; Daniel Burkhoff; Daniel Tamez
Journal:  ASAIO J       Date:  2016 Jan-Feb       Impact factor: 2.872

2.  Use of an intrapericardial, continuous-flow, centrifugal pump in patients awaiting heart transplantation.

Authors:  Keith D Aaronson; Mark S Slaughter; Leslie W Miller; Edwin C McGee; William G Cotts; Michael A Acker; Mariell L Jessup; Igor D Gregoric; Pranav Loyalka; O H Frazier; Valluvan Jeevanandam; Allen S Anderson; Robert L Kormos; Jeffrey J Teuteberg; Wayne C Levy; David C Naftel; Richard M Bittman; Francis D Pagani; David R Hathaway; Steven W Boyce
Journal:  Circulation       Date:  2012-05-22       Impact factor: 29.690

3.  Decoupling Between Diastolic Pulmonary Artery Pressure and Pulmonary Capillary Wedge Pressure as a Prognostic Factor After Continuous Flow Ventricular Assist Device Implantation.

Authors:  Teruhiko Imamura; Ben Chung; Ann Nguyen; Daniel Rodgers; Gabriel Sayer; Sirtaz Adatya; Nitasha Sarswat; Gene Kim; Jayant Raikhelkar; Takeyohi Ota; Tae Song; Colleen Juricek; Viktoriya Kagan; Valluvan Jeevanandam; Mandeep Mehra; Daniel Burkhoff; Nir Uriel
Journal:  Circ Heart Fail       Date:  2017-09       Impact factor: 8.790

4.  Medical management of patients with continuous-flow left ventricular assist devices.

Authors:  Adam D Devore; Robert J Mentz; Chetan B Patel
Journal:  Curr Treat Options Cardiovasc Med       Date:  2014-02

5.  Left Ventricular Decompression During Speed Optimization Ramps in Patients Supported by Continuous-Flow Left Ventricular Assist Devices: Device-Specific Performance Characteristics and Impact on Diagnostic Algorithms.

Authors:  Nir Uriel; Allison P Levin; Gabriel T Sayer; Kanika P Mody; Sunu S Thomas; Sirtaz Adatya; Melana Yuzefpolskaya; Arthur R Garan; Alexander Breskin; Hiroo Takayama; Paolo C Colombo; Yoshifumi Naka; Daniel Burkhoff; Ulrich P Jorde
Journal:  J Card Fail       Date:  2015-06-25       Impact factor: 5.712

6.  Hemodynamic Ramp Tests in Patients With Left Ventricular Assist Devices.

Authors:  Nir Uriel; Gabriel Sayer; Karima Addetia; Savitri Fedson; Gene H Kim; Daniel Rodgers; Eric Kruse; Keith Collins; Sirtaz Adatya; Nitasha Sarswat; Ulrich P Jorde; Colleen Juricek; Takeyoshi Ota; Valluvan Jeevanandam; Daniel Burkhoff; Roberto M Lang
Journal:  JACC Heart Fail       Date:  2015-12-30       Impact factor: 12.035

7.  Development of a novel echocardiography ramp test for speed optimization and diagnosis of device thrombosis in continuous-flow left ventricular assist devices: the Columbia ramp study.

Authors:  Nir Uriel; Kerry A Morrison; Arthur R Garan; Tomoko S Kato; Melana Yuzefpolskaya; Farhana Latif; Susan W Restaino; Donna M Mancini; Margaret Flannery; Hiroo Takayama; Ranjit John; Paolo C Colombo; Yoshifumi Naka; Ulrich P Jorde
Journal:  J Am Coll Cardiol       Date:  2012-10-03       Impact factor: 24.094

8.  Seventh INTERMACS annual report: 15,000 patients and counting.

Authors:  James K Kirklin; David C Naftel; Francis D Pagani; Robert L Kormos; Lynne W Stevenson; Elizabeth D Blume; Susan L Myers; Marissa A Miller; J Timothy Baldwin; James B Young
Journal:  J Heart Lung Transplant       Date:  2015-10-08       Impact factor: 10.247

Review 9.  Hemodynamics of Mechanical Circulatory Support.

Authors:  Daniel Burkhoff; Gabriel Sayer; Darshan Doshi; Nir Uriel
Journal:  J Am Coll Cardiol       Date:  2015-12-15       Impact factor: 24.094

  9 in total
  14 in total

1.  HVAD Flow Waveform Estimates Left Ventricular Filling Pressure.

Authors:  Teruhiko Imamura; Nikhil Narang; Daniel Rodgers; Daisuke Nitta; Jonathan Grinstein; Takeo Fujino; Gene Kim; Ann Nguyen; Valluvan Jeevanandam; Gabriel Sayer; Nir Uriel
Journal:  J Card Fail       Date:  2020-01-23       Impact factor: 5.712

Review 2.  Contemporary Perspectives in Durable Mechanical Circulatory Support: What Did We Learn in the Last 3 Years?

Authors:  Jayant Raikhelkar; Nir Uriel
Journal:  Curr Cardiol Rep       Date:  2018-01-29       Impact factor: 2.931

Review 3.  Physiology of the Assisted Circulation in Cardiogenic Shock: A State-of-the-Art Perspective.

Authors:  Julien Guihaire; Francois Haddad; Mita Hoppenfeld; Myriam Amsallem; Jeffrey W Christle; Clark Owyang; Khizer Shaikh; Joe L Hsu
Journal:  Can J Cardiol       Date:  2019-11-09       Impact factor: 5.223

Review 4.  Clinical implications of hemodynamic assessment during left ventricular assist device therapy.

Authors:  Teruhiko Imamura; Ben Chung; Ann Nguyen; Gabriel Sayer; Nir Uriel
Journal:  J Cardiol       Date:  2017-12-26       Impact factor: 3.159

Review 5.  Physiology and Clinical Utility of HeartMate Pump Parameters.

Authors:  Mark N Belkin; Viktoriya Kagan; Colleen Labuhn; Sean P Pinney; Jonathan Grinstein
Journal:  J Card Fail       Date:  2021-12-31       Impact factor: 6.592

Review 6.  Advancements in mechanical circulatory support for patients in acute and chronic heart failure.

Authors:  Thomas A Csepe; Ahmet Kilic
Journal:  J Thorac Dis       Date:  2017-10       Impact factor: 2.895

7.  Estimation of the Severity of Aortic Insufficiency by HVAD Flow Waveform.

Authors:  Teruhiko Imamura; Nikhil Narang; Daniel Rodgers; Daisuke Nitta; Takeo Fujino; Sara Kalantari; Bryan Smith; Gene Kim; Ann Nguyen; Ben Chung; Luise Holzhauser; Tae Song; Takeyoshi Ota; Valluvan Jeevanandam; Gabriel Sayer; Nir Uriel
Journal:  Ann Thorac Surg       Date:  2019-11-19       Impact factor: 4.330

8.  Estimation of Central Venous Pressure by Pacemaker Lead Impedances in Left Ventricular Assist Device Patients.

Authors:  Teruhiko Imamura; Joshua D Moss; Erin Flatley; Daniel Rodgers; Gene Kim; Jayant Raikhelkar; Nitasha Sarswat; Sara Kalantari; Ann Nguyen; Colleen Juricek; Daniel Burkhoff; Tae Song; Takeyoshi Ota; Valluvan Jeevanandam; Gabriel Sayer; Nir Uriel
Journal:  ASAIO J       Date:  2020-01       Impact factor: 3.826

9.  Continuous LVAD monitoring reveals high suction rates in clinically stable outpatients.

Authors:  Christoph Gross; Heinrich Schima; Thomas Schlöglhofer; Kamen Dimitrov; Martin Maw; Julia Riebandt; Dominik Wiedemann; Daniel Zimpfer; Francesco Moscato
Journal:  Artif Organs       Date:  2020-03-01       Impact factor: 3.094

10.  Lund exhaust on hemodynamic parameters and inflammatory mediators in patients undergoing cardiac valve replacement under cardiopulmonary bypass.

Authors:  Jihong Zhu; Weimin Zhang; Guoying Shen; Xin Yu; Jige Guo; Taidi Zhong
Journal:  Exp Ther Med       Date:  2018-06-26       Impact factor: 2.447

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.