OBJECTIVES: Orthodontic appliances are often removed before magnetic resonance imaging (MRI) scans because they are known to produce artefacts. The purpose of this study was to find the exact indications for removal of various fixed attachments when imaging four specific areas of the head and neck.
MATERIALS AND METHODS: Sixty patients requiring an MRI scan of the head for medical reasons volunteered for this investigation. One of four different types of fixed attachments (stainless steel brackets, titanium brackets, ceramic brackets with metal slots, and stainless steel retainers) were assigned to a patient. Each patient had two scans at 1.5 T: with an "empty wax jig" and with a wax jig including the attachment. Archwires were not investigated as they are easily removed before a scan. Two radiologists evaluated the scans of each patient and each of the four areas under investigation: maxillary sinus, oral cavity, temporomandibular joints and posterior cerebral fossa.
RESULTS: Stainless steel brackets always caused non-interpretability of all anatomic areas (100 per cent). Titanium brackets, ceramic brackets with metal slots, and stainless steel retainers caused artefact in the oral cavity only (for 20, 16.65, and 86.65 per cent of the subjects).
CONCLUSIONS: Our results show that ceramic brackets with metal slots and titanium brackets do not always have to be removed before MRI scans of the head and neck, depending on the area under investigation. Metal fixed retainers should only be removed if the oral cavity itself is under investigation. Stainless steel brackets should always be removed before MRI scans of the head and neck. © EDP Sciences, SFODF, 2017.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28597834     DOI: 10.1051/orthodfr/2017004

Source DB:  PubMed          Journal:  Orthod Fr        ISSN: 0078-6608


  2 in total

1.  MRI compatibility of orthodontic brackets and wires: systematic review article.

Authors:  Adrienn Dobai; Fanni Dembrovszky; Tamás Vízkelety; Péter Barsi; Fanni Juhász; Csaba Dobó-Nagy
Journal:  BMC Oral Health       Date:  2022-07-19       Impact factor: 3.747

Review 2.  Unwanted effects due to interactions between dental materials and magnetic resonance imaging: a review of the literature.

Authors:  Sherin Jose Chockattu; Deepak Byathnal Suryakant; Sophia Thakur
Journal:  Restor Dent Endod       Date:  2018-08-30
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.