LOW-DOSE CT PROTOCOL OPTIMIZATION FOR THE ASSESSMENT OF ACUTE APPENDICITIS: THE OPTICAP PHANTOM STUDY.

Hannele Niiniviita1,2, Paulina Salminen3, Juha M Grönroos3, Irina Rinta-Kiikka4, Saija Hurme5, Timo Kiljunen6, Jarmo Kulmala1, Mika Teräs1, Suvi Sippola3, Johanna Virtanen2.   

Abstract

The aim was to evaluate effects of voltage, noise input (NI) and iterative reconstruction (IR) on radiation dose and image quality in order to establish a contrast enhanced low-dose protocol for assessment of acute appendicitis. An anthropomorphic abdominal phantom mimicking contrast enhanced abdomen was scanned with 80, 100 and 120 kV, standard and strong IR and 11 NIs (66 protocols). A total of 14 test tubes of increasing iodine dilutions and one tube with an appendicolith were evaluated within the phantom. The dose, HUs, noise, contrast-to-noise ratio (CNR) and figure of merit (FOM) were determined. Visual quality scores were assessed by two readers. A clinically used voltage-IR combination (120 kV, standard IR) was used as a reference. Overall, 100 kV with standard IR (p = 0.002) and 80 kV with both IRs (p < 0.001) showed higher CNR than the reference, but noise was most pronounced at 80 kV (p < 0.001). The highest FOM was found in the 100 kV protocols (p < 0.001). The reference and 100 kV with standard IR had highest image quality scores, where the 100 kV protocol enabled a distinct dose reduction. Lowering the voltage seems to be a more favorable tool than IR changes in optimizing the dose in contrast enhanced abdominal CT. TRIAL REGISTRATION: ClinicalTrials.gov number, NCT01022567.
© The Author 2017. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 28591824     DOI: 10.1093/rpd/ncx070

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  2 in total

1.  A randomised placebo-controlled double-blind multicentre trial comparing antibiotic therapy with placebo in the treatment of uncomplicated acute appendicitis: APPAC III trial study protocol.

Authors:  Suvi Sippola; Juha Grönroos; Ville Sallinen; Tero Rautio; Pia Nordström; Tuomo Rantanen; Saija Hurme; Ari Leppäniemi; Sanna Meriläinen; Johanna Laukkarinen; Heini Savolainen; Johanna Virtanen; Paulina Salminen
Journal:  BMJ Open       Date:  2018-11-03       Impact factor: 2.692

2.  A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre-A phantom study.

Authors:  Patrizio Barca; Fabio Paolicchi; Giacomo Aringhieri; Federica Palmas; Daniela Marfisi; Maria Evelina Fantacci; Davide Caramella; Marco Giannelli
Journal:  PLoS One       Date:  2021-01-14       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.