Asbjørn M Scholtens1, Laurens E Swart2, Henryk J Te Kolste3, Ricardo P J Budde2, Marnix G E H Lam4, Hein J Verberne5.
Abstract
BACKGROUND: The significance of and threshold values for the standardized uptake value (SUV) in FDG PET/CT to diagnose prosthetic heart valve (PHV) endocarditis (PVE) are unclear at present.Entities:
Keywords: 18F-fluorodeoxyglucose; EARL; computed tomography; endocarditis; positron emission tomography; prosthetic heart valve
Mesh:
Substances:
Year: 2017 PMID: 28585026 PMCID: PMC6280770 DOI: 10.1007/s12350-017-0932-x
Source DB: PubMed Journal: J Nucl Cardiol ISSN: 1071-3581 Impact factor: 5.952
Study characteristics
| Study | Year | Patients | Valves | Patient preparation | FDG dose | Time to acquisition | Reconstruction method reported | EARL-accredited site | Camera type |
|---|---|---|---|---|---|---|---|---|---|
| Mathieu et al | 2017 | 51 | 54 | 6 hour fast* | 4 MBq/kg | 60 minute | 3-dimensional time-of-flight ordered-subsets expectation maximization | Yes | GE discovery 690 |
| Jiménez-Ballvé et al | 2016 | 41 | 42 | 48 hour HFLC diet, 12 hour fast, 50 IU/kg heparin pre-administration | 5 MBq/kg | 45–60 minute | Iterative reconstruction | No | Siemens biograph 6 TruePoint |
| Salomäki et al | 2015 | 23 | 16 | 24 hour LC diet, 10 hour fast | mean 280 MBq | mean 72 minute | Ordered subset expectation maximization | Yes | GE discovery VCT |
| Fagman et al | 2015 | 30 | 30 | HFLC meal, 18 hour fast (N = 11) | 4 MBq/kg | 60 minute | Default iterative | Yes | Siemens biograph TruePoint 64 |
| 6 hour fast (N = 19)† | |||||||||
| Pizzi et al | 2015 | 92 | 65 | 12 hour fast, 50 IU/kg heparin pre-administration | 3.7 MBq/kg | 60 minute | Iterative TrueX + Time-of-Flight | No | Siemens biograph mCT 64S |
| Rouzet et al | 2014 | 39 | 45 | HFLC meal, 12 hour fast | 4 MBq/kg | 60 minute | Iterative 3-dimensional reconstruction algorithm of the system software | Yes | GE discovery 690 |
| Bartoletti et al | 2014 | 6 | 6 |
|
|
|
| No |
|
| Saby et al | 2013 | 72 | 72 | HFLC meal, 12 hour fast | 5 MBq/kg | 60 minute | Ordered subset expectation maximization | No | GE discovery ST |
* Scans with intense, homogenous uptake in myocardium excluded from analysis
† Controls based on oncologic scan protocols
HFLC, high fat, low carbohydrate; LCD, low carbohydrate diet; IU, International Units; MBq, Megabecquerel
Figure 1Reported SUVs in the eligible studies as median, interquartile ranges and total ranges unless noted otherwise. Green values were reported by EARL-accredited centers. ‡Values reported as median with total ranges. †Values reported as median with quartile ranges
Figure 2Example of proposed measurement standardization in a mitral valve prosthesis. Fused PET/CT images (upper row) and corresponding attenuation-corrected PET images (below) showing whole-valve measurement VOI (in this case a self-expanding VOI set to include voxels ≥40% of maximum) and VOI sphere in the descending aorta (lower right). Uptake in myocardium (hardly present here) and the aortic wall are excluded