Literature DB >> 28576455

Efficacy of the SENSIMED Triggerfish® in the postoperative follow-up of PHACO-ExPRESS combined surgery.

V Osorio-Alayo1, V T Pérez-Torregrosa2, R Clemente-Tomás2, Á Olate-Pérez2, M Cerdà-Ibáñez2, A Gargallo-Benedicto2, A Barreiro-Rego2, A Duch-Samper2.   

Abstract

OBJECTIVE: To determine the usefulness of the SENSIMED Triggerfish® system in the postoperative control of combined phacoemulsification and ExPRESS implant (PHACO-ExPRESS) surgery in patients with cataract and chronic open angle glaucoma (COAG) during a 2 months follow-up.
METHODS: A prospective study conducted on 15 eyes that were subjected to PHACO-ExPRESS combined surgery. Using the SENSIMED Triggerfish® system, two records of the circadian patterns of intraocular pressure (IOP) were performed, one before and one after surgery. A record was made of the best corrected visual acuity (BCVA), comorbidities, previous IOP, and 7-30-60 days after surgery, as well as any hypotensive drugs and complications.
RESULTS: The final sample was 12 eyes. The mean pre-operative BCVA (log MAR chart) before surgery was 0.5±0.2, and after surgery 0.14±0.1 (P=.02). The previous IOP was 18.7±3.8mmHg with 2.9±0.7 drugs. The mean IOP at 7, 30, and 60 days after surgery decreased to 13±4.1mmHg (P=.002), 13.5±2mmHg (P=.001), and 13.9±2.5mmHg (P=.001), respectively. The amplitudes of the circadian curves changed significantly after surgery (P=.007). The mean values between daytime and night-time periods decreased significantly from 146.8±80.9 mVeq and 61.2±92.mVeq before surgery to 36.4±36 mVeq (P=.000), and -23,2±47.6mVeq (P=.014) after surgery, respectively. There were complications in one patient.
CONCLUSIONS: The SENSIMED Triggerfish® monitoring system showed changes in the curves of the circadian patterns, as well as decreased mean amplitudes after the combined PHACO-ExPRESS technique, suggesting that it may become a useful tool for postoperative follow-up of COAG.
Copyright © 2017 Sociedad Española de Oftalmología. Publicado por Elsevier España, S.L.U. All rights reserved.

Entities:  

Keywords:  24horas; 24hours; Contact lens sensor; ExPRESS; Fluctuación presión intraocular; Glaucoma; Glaucoma open-angle; Glaucoma ángulo abierto; Glaucoma/cirugía; Glaucoma/surgery; Intraocular pressure fluctuation; Sensimed triggerfish; Sensor lentes de contacto

Mesh:

Year:  2017        PMID: 28576455     DOI: 10.1016/j.oftal.2017.04.003

Source DB:  PubMed          Journal:  Arch Soc Esp Oftalmol        ISSN: 0365-6691


  2 in total

Review 1.  Miniaturization in Glaucoma Monitoring and Treatment: A Review of New Technologies That Require a Minimal Surgical Approach.

Authors:  H Burkhard Dick; Tim Schultz; Ronald D Gerste
Journal:  Ophthalmol Ther       Date:  2019-02-06

2.  The effect of daily life activities on intraocular pressure related variations in open-angle glaucoma.

Authors:  Kevin Gillmann; Robert N Weinreb; Kaweh Mansouri
Journal:  Sci Rep       Date:  2021-03-23       Impact factor: 4.379

  2 in total