Quality assurance of the jaws only-intensity modulated radiation therapy plans for head-and-neck cancer.

Duong Thanh Tai1, Nguyen Dong Son2, Truong Thi Hong Loan3, H P W Anson4.   

Abstract

Intensity-modulated radiation therapy (IMRT) is a treatment technique which has become routine in developed countries. In most centers this technique is delivered with multi-leaf collimators (MLCs). However, the use of MLCs is not mandatory. Several oncology centres in developing countries are still using linear accelerators (LINAC) without MLCs, and can potentially deliver IMRT plans with the use of collimator jaws. In this report, we present the results of quality assurance of this Jaws-Only-IMRT (JO-IMRT) technique in treating nasopharyngeal carcinoma (NPC) patients. Twenty-five plans of nasopharyngeal patients were randomly chosen. For each patient, a JO-IMRT plan was generated and a series of pre-treatment verification measurements was performed including (1) point dose measurement with an ionization chamber, (2) planar dose measurement with a 2D-array detector and (3) 3-dimensional dose measurement using a rotatable phantom with a 2D-array detector. The average differences between the measured and TPS-calculated point doses were found to be 1.26±0.77%, which is within the institution's dose constraint limits. For the planar dose and 3D dose measurements, the average gamma index based on 3%/3mm criteria were 96.77±2.33% and 94.72±2.67%, respectively. Our measurements showed that the JO-IMRT treatment plans applied to the H&N patients were accurate for the treatment delivery based on our established pass criteria.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Intensity-modulated radiation therapy (IMRT); Jaws only IMRT (JO-IMRT); Quality assurance (QA)

Mesh:

Year:  2017        PMID: 28571708     DOI: 10.1016/j.ejmp.2017.05.059

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  Dosimetric and radiobiological comparison in head-and-neck radiotherapy using JO-IMRT and 3D-CRT.

Authors:  Duong Thanh Tai; Luong Thi Oanh; Pham Hoai Phuong; Abdelmoneim Sulieman; Fouad A Abolaban; Hiba Omer; James C L Chow
Journal:  Saudi J Biol Sci       Date:  2022-06-02       Impact factor: 4.052

2.  Dosimetric and Monte Carlo verification of jaws-only IMRT plans calculated by the Collapsed Cone Convolution algorithm for head and neck cancers.

Authors:  Duong Thanh Tai; Luong Thi Oanh; Nguyen Dong Son; Truong Thi Hong Loan; James C L Chow
Journal:  Rep Pract Oncol Radiother       Date:  2018-11-28
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.