Results from multiple prospective single-center clinical trials of the off-the-shelf p-Branch fenestrated stent graft.

Mark A Farber1, Matthew J Eagleton2, Tara M Mastracci3, James F McKinsey4, Raghuveer Vallabhaneni5, Björn Sonesson6, Nuno Dias6, Timothy Resch6.   

Abstract

OBJECTIVE: The purpose of this study was to report prospective data of an off-the-shelf fenestrated endograft (Zenith p-Branch; Cook Medical, Bloomington, Ind) from four centers for the treatment of patients with pararenal abdominal aortic aneurysms.
METHODS: Data were combined from four single-center investigational studies conducted in the United States and Europe. The p-Branch endograft consists of a proximal off-the-shelf component incorporating a scallop for the celiac artery, a superior mesenteric artery fenestration, and two conical pivot fenestrations to preserve flow to the renal vessels. The device is available in two configurations, a left renal fenestration at the same (configuration A) or lower (configuration B) longitudinal position than the right to accommodate varied anatomy of the patients.
RESULTS: Between August 2011 and September 2015, 76 patients (82% male; mean age, 72 years; 65 elective and 11 emergent) were enrolled, with 55% implanted with option A and 45% with B. The device was deployed successfully in all patients, and stents were placed in all target vessels except in three cases (one elective, two emergent): a left kidney was sacrificed in one patient, and a right renal artery was left unstented in two patients during the index procedure. There was no 30-day mortality. During follow-up (mean, 25 ± 13 months), 10 late deaths occurred (6 elective, 4 emergent; none related to device or procedure), and there were no ruptures or conversions to open repair. Two patients experienced bowel ischemia; one case resolved with nonoperative treatment and one required superior mesenteric artery and celiac artery angioplasty and stent placement. Renal artery occlusion occurred in eight patients (11%) and was deemed procedure related in 63% (5/8) of these patients. Four of these were successfully intervened on with preservation of renal function. The overall renal insufficiency incidence was 7% (5/76). One patient developed renal failure requiring dialysis.
CONCLUSIONS: Early results incorporating learning curves for physicians with a new device and delivery system indicate that the use of the Zenith p-Branch device is feasible and safe. Long-term follow-up is needed to assess the effectiveness and durability of this treatment strategy and to refine the indications for use.
Copyright © 2017 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28559176     DOI: 10.1016/j.jvs.2017.01.068

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  3 in total

1.  The state of complex endovascular abdominal aortic aneurysm repairs in the Vascular Quality Initiative.

Authors:  Thomas F X O'Donnell; Virendra I Patel; Sarah E Deery; Chun Li; Nicholas J Swerdlow; Patric Liang; Adam W Beck; Marc L Schermerhorn
Journal:  J Vasc Surg       Date:  2019-02-02       Impact factor: 4.268

2.  Use of fenestrated-branched endovascular aneurysm repair to treat Carrel patch aneurysmal degeneration after open thoracoabdominal aortic aneurysm repair.

Authors:  Tammy T Nguyen; Jessica P Simons; Andres Schanzer
Journal:  J Vasc Surg Cases Innov Tech       Date:  2019-04-28

3.  Systematic Review and Meta-Analysis of Published Studies on Endovascular Repair of Abdominal Aortic Aneurysm With the p-Branch.

Authors:  Haoliang Wu; Liwei Zhang; Mingxing Li; Shunbo Wei; Cong Zhang; Hualong Bai
Journal:  Front Surg       Date:  2022-04-29
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.