Neal S Gerstein1, A Robb McLean2, Eric C Stecker3, Peter M Schulman4.
Abstract
The simultaneous use of 2 external defibrillators to administer either dual or sequential cardioversion or defibrillation for refractory cardiac arrhythmias is increasing in both the out-of-hospital and inhospital settings. Using 2 defibrillators to administer higher energy levels than can be achieved with a single defibrillator is considered off-label and is currently not part of published advanced cardiac life support guidelines. We report the first case in which the use of dual-dose cardioversion was associated with external defibrillator damage. Because defibrillator damage, especially if undetected, jeopardizes patient safety and off-label medical product use may void the manufacturer's warranty, this case should urge users to proceed with caution when contemplating this technique.Entities:
Mesh:
Year: 2017 PMID: 28559035 DOI: 10.1016/j.annemergmed.2017.04.005
Source DB: PubMed Journal: Ann Emerg Med ISSN: 0196-0644 Impact factor: 5.721