Sensitive albuminuria analysis using dye-binding based test strips.

Joris R Delanghe1, Jonas Himpe2, Naomi De Cock2, Sigurd Delanghe3, Kevin De Herde2, Veronique Stove2, Marijn M Speeckaert3.   

Abstract

BACKGROUND: Populations at increased risk for chronic kidney disease should be screened for albuminuria. Possibilities of advanced urine strip readers based on complementary metal oxide semiconductor (CMOS) sensor technology were investigated for obtaining quantitative albuminuria results.
METHODS: Reflectance data of test strips (Sysmex UFC 3500 reader+CMOS) were compared with albuminuria (BNII) and with proteinuria (Cobas 8000). Urinary creatinine was assayed using a Jaffe-based creatinine assay (Cobas 8000).
RESULTS: Calibration curve was made between 11.5 and 121.5mg/L with detection limit of 5.5mg/L. Within-run CV values of reflectance data were 0.21% (UC-Control L; 10mg/L) and 0.37% (UC-Control H; >150mg/L) for albumin, and 0.71%/3.97% for creatinine. Between-run CV values were 0.24%/0.42% for albumin and 0.93%/5.13% for creatinine. A strong correlation (r=0.92) was obtained between albuminuria (BNII) and protein strip reflectance data. Creatinine reflectance data correlated well with Jaffe-based urinary creatinine data (r=0.90). Albumin:creatinine ratio obtained by test strip and by wet chemistry showed a good correlation (r=0.59). Carbamylated, glycated and partially hydrolyzed isoforms of albumin could be detected by test strip.
CONCLUSIONS: Dye-binding based albumin test strip assay in combination with a CMOS based reader would potentially allow quantitative analysis of albuminuria and determination of albumin:creatinine ratio.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Albuminuria; Chronic kidney disease; Diabetes mellitus; Reflectance data; Test strips

Mesh:

Substances:

Year:  2017        PMID: 28554541     DOI: 10.1016/j.cca.2017.05.032

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  6 in total

1.  Urine test strips vs. pyrogallol red-molybdate assays for proteinuria: a critical approach.

Authors:  Joris R Delanghe; Marc L De Buyzere; Matthijs Oyaert; Sigurd E Delanghe; Marijn M Speeckaert
Journal:  Clin Exp Nephrol       Date:  2020-02-26       Impact factor: 2.801

Review 2.  Progress in Automated Urinalysis.

Authors:  Matthijs Oyaert; Joris Delanghe
Journal:  Ann Lab Med       Date:  2019-01       Impact factor: 3.464

3.  Semiquantitative, fully automated urine test strip analysis.

Authors:  Matthijs Oyaert; Joris R Delanghe
Journal:  J Clin Lab Anal       Date:  2019-02-25       Impact factor: 2.352

4.  Internal Quality Control Data of Urine Reagent Strip Tests and Derivation of Control Rules Based on Sigma Metrics.

Authors:  Haeil Park; Younsuk Ko
Journal:  Ann Lab Med       Date:  2021-09-01       Impact factor: 3.464

5.  Comparison of three types of analyzers for urine protein-to-creatinine ratios in dogs.

Authors:  Sumin Ji; Yeseul Yang; Yeji Jeong; Sung Hyun Hwang; Myung Chul Kim; Yongbaek Kim
Journal:  J Vet Sci       Date:  2021-01       Impact factor: 1.672

6.  A Novel Substrate-Inspired Fluorescence-Based Albumin Detection Improves Assessment of Clinical Outcomes in Hemodialysis Patients Receiving a Nursing Nutrition Intervention.

Authors:  Lei You; Xia Wang; Wenhong Wang
Journal:  Med Sci Monit       Date:  2021-08-10
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.