Mette Thrane Foged1,2, Ulrich Lindberg1,3, Kishore Vakamudi4,5, Henrik B W Larsson2,3, Lars H Pinborg1,2, Troels W Kjær2,6, Martin Fabricius6, Claus Svarer1, Brice Ozenne7, Carsten Thomsen2,8, Sándor Beniczky6,9,10, Olaf B Paulson1,2, Stefan Posse4,5,11.
Abstract
PURPOSE: Concurrent EEG and fMRI is increasingly used to characterize the spatial-temporal dynamics of brain activity. However, most studies to date have been limited to conventional echo-planar imaging (EPI). There is considerable interest in integrating recently developed high-speed fMRI methods with high-density EEG to increase temporal resolution and sensitivity for task-based and resting state fMRI, and for detecting interictal spikes in epilepsy. In the present study using concurrent high-density EEG and recently developed high-speed fMRI methods, we investigate safety of radiofrequency (RF) related heating, the effect of EEG on cortical signal-to-noise ratio (SNR) in fMRI, and assess EEG data quality.Entities:
Mesh:
Year: 2017 PMID: 28552957 PMCID: PMC5446172 DOI: 10.1371/journal.pone.0178409
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
fMRI acquisition parameters.
| Scanner (array coil) | Philips Achieva (32 cha) | Siemens Prisma (64 cha) | ||||||
|---|---|---|---|---|---|---|---|---|
| fMRI Method | EPI | EPI | MEPI | MEPI | MB4-EPI | MS-EVI | EPI | MB8-EPI |
| TR [ms] | 3034 ( | 402 | 3034 | 376 | 450 ( | 487( | 1980 | 280 |
| TE [ms] | 35 | 35 | 9.6 | 9.6 | 35 | 35 | 35 | 35 |
| Number of Tes | 1 | 1 | 4 | 3 | 1 | 1 | 1 | 1 |
| Slices | 37 | 6 | 37 | 6 | 24 | 24 | 30 | 32 |
| Number of volumes | 200 | 3000 | 200 | 3200 | 1400 | 895 | 150 | 1100 |
| Scan time [min:sec] | 10:16 | 20:00 | 10:16 | 20:03 | 10:36 | 04:34 | 10:14 | 19:11 |
| Voxel size [mm] | 3.1 | 3.1 | 3.1 | 3.1 | 2.88 | 4 | 4 | 4 |
| Slice thickness [mm] | 3.1 | 3.1 | 3.1 | 3.1 | 3.3 | 4 | 4 | 4 |
| Slice gap [mm] | 0.31 | 0.31 | 0.31 | 0.31 | 0.7 | 0 | 1 | 1 |
| Matrix size | 76x73 | 76x73 | 76x73 | 76x73 | 80x79 | 64x61 | 64x64 | 64x64 |
| Flip angle [degrees] | 90 | 30 | 90 | 30 | 15 | 30 | 90 | 30 |
* In some cases a TR of 2000 ms was used
** In one case a TR of 381 ms was use
*** In some cases a TR of 300 ms was used.
Cha = Channel, TR = Repetition time, TE = Echo time (minimum TE in case of MEPI), EPI = Echo planar imaging, MEPI = Multi echo planar imaging, MB4-EPI = Multiband-4 echo planar imaging, MS-EVI = Multi-slab echo volumar imaging, MB8EPI = Multiband-8 echo planar imaging.
Fig 1Flowchart of the scan protocol during the systematically temperature measurements.
Temperature measurements and SAR values during MB-EPI sequences in concurrent EEG and MRI in the Phillips Achieva scanner demonstrating the impact of increased flip angle.
| Subject | Sequence | Flip angle (degree) | Scan Duration [min] | SAR [%] (Mean/SD) | SAR [%] (Max) | ΔT [°C] (Mean/SD) | ΔT [°C] (Max) | ||
|---|---|---|---|---|---|---|---|---|---|
| Phantom (n = 1) | MB-EPI | 256 | 33.00 | 51 | 0 | 51 | 0.8 | 0.3 | 1.1 |
| Phantoms (n = 2) | MB-EPI | 120 | 20.0/30.0 | 35 | 6 | 39 | 0.5 | 0 | 0.5 |
| Control (n = 1) | MB-EPI | 120 | 4.27 | 31 | 0 | 31 | 0.8 | 0.5 | 1.1 |
| Control + patients (n = 4) | MB-EPI | 15 | 10.0/20.0 | 18 | 2 | 21 | 0.2 | 0.3 | 0.8 |
Measurements with increased flip angle are in italics. Temperature measurements from all probes were combined.
ΔT: Absolute temperature change.
Summary of temperature measurements.
| Scanner (head RF array coil) | Siemens Prisma (64 ch) (N = 3) | Phillips Achieva (32 ch) (N = 3) | ||||
|---|---|---|---|---|---|---|
| Frontal | Occipital | Temporal | Frontal | Occipital | Temporal | |
| Mean outside | 1.3 | 1.5 | 1.3 | 1.2 | 0.0 | 2.2 |
| Mean inside w/o scanning | 0.1 | 0.6 | 0.9 | 0.1 | 1.5 | 0.4 |
| Mean EPI with cap | 0.2 | 0.4 | 0.6 | 0.2 | 0.1 | 0.8 |
| Mean MB-EPI with cap | 0.1 | 0.6 | 0.5 | 0.5 | 0.4 | 0.9 |
| Mean MS-EVI with cap | 0.1 | 0.2 | 1.2 | |||
| Mean EPI w/o cap | 0.2 | 1.9 | 1.4 | -0.1 | 0.3 | |
| Mean MB-EPI w/o cap | 0.5 | 1.7 | 0.7 | -0.1 | 1.2 | 0.2 |
| Mean MS-EVI w/o cap | 0.2 | 0.2 | ||||
| Mean outside | 0.041 | 0.048 | 0.041 | 0.038 | 0.000 | 0.073 |
| Mean inside w/o scanning | 0.010 | 0.057 | 0.087 | 0.005 | 0.150 | 0.035 |
| Mean EPI with cap | 0.007 | 0.013 | 0.019 | 0.005 | 0.002 | 0.025 |
| Mean MB-EPI with cap | 0.003 | 0.018 | 0.016 | 0.016 | 0.013 | 0.028 |
| Mean MS-EVI with cap | 0.002 | 0.006 | 0.036 | |||
| Mean EPI w/o cap | 0.008 | 0.058 | 0.042 | -0.002 | 0.009 | |
| Mean MB-EPI w/o cap | 0.017 | 0.058 | 0.022 | -0.003 | 0.036 | 0.006 |
| Mean MS-EVI w/o cap | 0.007 | 0.006 | ||||
Fig 2Temperature measurement during a 30-minute EPI sequence with EEG cap on, in the Siemens Prisma scanner.
Fig 6Temperature measurement during a 30-minute EVI sequence with EEG cap on, in the Phillips Achieva scanner.
Summary of SAR values.
| SAR [%] | |||||
|---|---|---|---|---|---|
| Scanner | EEG System | Sequence / EEG cap | Mean | SD | Max |
| Philips Achieva | Brain Products 64 channel | EPI_with_cap | 24 | 1 | 25 |
| EPI_without_cap | 25 | ||||
| MB-EPI_with_cap | 18 | 2 | 21 | ||
| MB-EPI_without_cap | 17 | ||||
| MS-EVI_with_cap | 13 | 3 | 17 | ||
| MS-EVI_without_cap | 14 | ||||
| Siemens Prisma | EGI 256 channel | EPI_with_cap | 19 | 3 | 23 |
| EPI_without_cap | 8 | 0 | 8 | ||
| MB-EPI_with_cap | 16 | 3 | 18 | ||
| MB-EPI_without_cap | 7 | 1 | 8 | ||
Fig 7Posterior dominant (alpha) rhythm in an EEG recorded in the MR scanner, during a high-speed sequence.
Left: Power Spectrum at EEG electrodes demonstrates a clear peak at 9.44 Hz, at the occipital electrodes. Right: power-map of alpha activity showing its 3D distribution.
EEG quality using conventional versus high-speed MR sequences and dependence on head motion.
| Conventional | High-speed | Total | |
|---|---|---|---|
| Number of scans | 30 | 20 | 50 |
| 18 (60%) | 11 (55%) | 29 (58%) | |
| 0.1 (0.1) | 0.2 (0.1) | ||
| 1.4 (1.1) | 0.9 (0.6) |
Relative and absolute motion are reported using mean and standard deviation.
EPI = Echo planar imaging. MEPI = Multi-echo planar imaging. MS-EVI = Multi-slab echo volumar imaging. MB-EPI = Multi-band echo planar imaging.
†Data from 4 scans missing.