Drilling resistance: A method to investigate bone quality.

Waqas A Lughmani1, Farukh Farukh2, Kaddour Bouazza-Marouf1, Hassan Ali3.   

Abstract

PURPOSE: Bone drilling is a major part of orthopaedic surgery performed during the internal fixation of fractured bones. At present, information related to drilling force, drilling torque, rate of drill-bit penetration and drill-bit rotational speed is not available to orthopaedic surgeons, clinicians and researchers as bone drilling is performed manually.
METHODS: This study demonstrates that bone drilling force data if recorded in-vivo, during the repair of bone fractures, can provide information about the quality of the bone. To understand the variability and anisotropic behaviour of cortical bone tissue, specimens cut from three anatomic positions of pig and bovine were investigated at the same drilling speed and feed rate.
RESULTS: The experimental results showed that the drilling force does not only vary from one animal bone to another, but also vary within the same bone due to its changing microstructure. Drilling force does not give a direct indication of bone quality; therefore it has been correlated with screw pull-out force to provide a realistic estimation of the bone quality. A significantly high value of correlation (r2 = 0.93 for pig bones and r2 = 0.88 for bovine bones) between maximum drilling force and normalised screw pull-out strength was found.
CONCLUSIONS: The results show that drilling data can be used to indicate bone quality during orthopaedic surgery.

Entities:  

Mesh:

Year:  2017        PMID: 28552924

Source DB:  PubMed          Journal:  Acta Bioeng Biomech        ISSN: 1509-409X            Impact factor:   1.073


  3 in total

1.  A New Site Preparation Protocol That Supports Bone Quality Evaluation and Provides Predictable Implant Insertion Torque.

Authors:  Stefan Velikov; Cristiano Susin; Peter Heuberger; Ainara Irastorza-Landa
Journal:  J Clin Med       Date:  2020-02-11       Impact factor: 4.241

2.  Effects of Different Undersizing Site Preparations on Implant Stability.

Authors:  Bernardo Ferreira Lemos; Paula Lopez-Jarana; Carlos Falcao; Blanca Ríos-Carrasco; Javier Gil; José Vicente Ríos-Santos; Mariano Herrero-Climent
Journal:  Int J Environ Res Public Health       Date:  2020-12-02       Impact factor: 3.390

3.  Effects of the Insertion Type and Depth on the Pedicle Screw Pullout Strength: A Finite Element Study.

Authors:  K Jendoubi; Y Khadri; M Bendjaballah; N Slimane
Journal:  Appl Bionics Biomech       Date:  2018-07-26       Impact factor: 1.781

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.