An integrated blood pressure measurement system for suppression of motion artifacts.

Huthaifa N Abderahman1, Hilmi R Dajani2, Miodrag Bolic2, Voicu Z Groza2.   

Abstract

Accuracy in blood pressure (BP) estimation is essential for proper diagnosis and management of hypertension. Motion artifacts are considered external sources of inaccuracy and can be due to sudden arm motion, muscle tremor, shivering, and transport vehicle vibrations. In the proposed work, a new algorithmic stage is integrated in a non-invasive BP monitor. This stage suppresses the effect of the motion artifact and adjusts the pressure estimation before displaying it to users. The proposed stage is based on a 3-axis accelerometer signal, which helps in the accurate detection of the motion artifact. Both transient motion artifacts and artifact due to vibrations are suppressed using algorithms based on Empirical Mode Decomposition (EMD). Measurements with human subjects show that the proposed algorithms considerably improved the accuracy of the blood pressure estimates in comparison with the commonly-used conventional oscillometric algorithm that does not include an EMD-based stage for artifact suppression, and allowed the estimates to meet the requirements of the international ANSI/AAMI/ISO standard.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Accelerometer; Blood pressure measurement; EMD; Motion artifacts; Vibration

Mesh:

Year:  2017        PMID: 28552114     DOI: 10.1016/j.cmpb.2017.03.007

Source DB:  PubMed          Journal:  Comput Methods Programs Biomed        ISSN: 0169-2607            Impact factor:   5.428


  1 in total

1.  Implementation and Validation of a Novel Noninvasive blood Pressure Measurement Algorithm Based on Oscillometric Method with Obtaining International Hypertension Standards.

Authors:  Sahar Karimian; Mohammadreza Yazdchi; Reza Hajian
Journal:  J Med Signals Sens       Date:  2022-05-12
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.