Impacts of 3D bone-to- implant contact and implant diameter on primary stability of dental implant.

Jui-Ting Hsu1, Yen-Wen Shen2, Chih-Wei Kuo3, Ruei-Teng Wang4, Lih-Jyh Fuh5, Heng-Li Huang6.   

Abstract

BACKGROUND/
PURPOSE: This study investigated the effects of three three-dimensional (3D) bone-to-implant contact (BIC) parameters-potential BIC area (pBICA), BIC area (BICA), and 3D BIC percentage (3D BIC%; defined as BICA divided by pBICA)-in relation to the implant diameter on primary implant stability, as well as their correlations were also evaluated.
METHODS: Dental implants with diameters of 3.75, 4, 5, and 6 mm and artificial bone specimens were scanned by microcomputed tomography to construct 3D models for calculating pBICA, BICA, and 3D BIC%. Indexes of the primary implant stability including the insertion torque value (ITV), Periotest value (PTV), and implant stability quotient (ISQ) were measured after implants with various diameters were placed into bone specimens. The Kruskal-Wallis test, Wilcoxon rank-sum test with Bonferroni adjustment, and Spearman correlations were all performed as statistical and correlation analyses.
RESULTS: The implant diameter significantly influenced pBICA and BICA, but not 3D BIC%. ITV and PTV were more sensitive to implant diameter than was ISQ. The coefficients of determination were high (>0.92) for the correlations between pBICA (or BICA) and indexes of the primary implant stability.
CONCLUSION: This study revealed how the implant diameter and the three-dimensional (3D) BIC influence the primary stabilities of dental implant. ITV and PTV were more sensitively influenced by the implant diameter than ISQ. The pBICA and BICA seem to be more important than 3D BIC % for using wider implant in treatment plan, since those two parameters are highly predictive of variations in the primary stability of dental implant.
Copyright © 2017. Published by Elsevier B.V.

Entities:  

Keywords:  3D bone-to-implant contact; Implant diameter; Implant stability quotient; Insertion torque; Microcomputed tomography; Periotest

Mesh:

Substances:

Year:  2017        PMID: 28551316     DOI: 10.1016/j.jfma.2017.05.005

Source DB:  PubMed          Journal:  J Formos Med Assoc        ISSN: 0929-6646            Impact factor:   3.282


  5 in total

1.  Assessment of Implant Stability in Immediate Implant Placement using Different Bone Grafting materials: A Clinical Study.

Authors:  Mohammad Jalaluddin; Saurabh Sathe; Joyce Thomas; Shaista Haleem; Sachin Naik; Madhuniranjanswamy Mahalakshmamma Shivanna
Journal:  J Pharm Bioallied Sci       Date:  2021-06-05

2.  Bone quality affects stability of orthodontic miniscrews.

Authors:  Wan-Ping Yu; Ming-Tzu Tsai; Jian-Hong Yu; Heng-Li Huang; Jui-Ting Hsu
Journal:  Sci Rep       Date:  2022-02-18       Impact factor: 4.379

3.  Non-Destructive Removal of Dental Implant by Using the Cryogenic Method.

Authors:  Burak Ak; Emre Gürkan Eroğlu; Abdullah Seckin Ertugrul; Ayla Batu Öztürk; Şakir Necat Yılmaz
Journal:  Medicina (Kaunas)       Date:  2022-06-25       Impact factor: 2.948

4.  Effects of implant length and 3D bone-to-implant contact on initial stabilities of dental implant: a microcomputed tomography study.

Authors:  Jui-Ting Hsu; Aaron Yu-Jen Wu; Lih-Jyh Fuh; Heng-Li Huang
Journal:  BMC Oral Health       Date:  2017-11-21       Impact factor: 2.757

5.  Digital occlusal analysis of pre and post single posterior implant restoration delivery: A pilot study.

Authors:  Ting Zhou; Jirapa Wongpairojpanich; Maytha Sareethammanuwat; Charukrit Lilakhunakon; Borvornwut Buranawat
Journal:  PLoS One       Date:  2021-07-02       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.