Literature DB >> 28551048

Randomized Study Comparing Equal Height Staples With Graduated Height Staples in Bronchial Closure.

Jiro Okami1, Toshiteru Tokunaga2, Takashi Kanou2, Hidenori Kunou2, Daisuke Ishida2, Ayako Fujiwara2, Yuri Ito3, Masahiko Higashiyama2.   

Abstract

BACKGROUND: There are two forms of staple cartridge available for lung cancer operations, the flat face with equal height staples and the stepped face with graduated height staples. The objective of the study was to evaluate their stapling ability in lobectomy.
METHODS: A randomized prospective study was conducted to compare two types of stapling reloads for bronchial closure. Stapling ability was evaluated by the staple formation on the resected bronchial stump. The stumps were sliced along the staple line, and the shapes of the staples were recorded by roentgenogram. Then, the staple formation was scored as 0 to 4 points with 4 approximating a perfect B-shaped staple.
RESULTS: A total of 61 patients were randomly assigned to the equal height staples (n = 30) or the graduated height staples (n = 31). From 61 patients, 183 staple lines, which included 1,144 staples, were evaluated. The case scores were significantly lower in the equal height staples than in the graduated height staples (2.17 versus 2.88, p = 0.0003), respectively. The percentage of staples that formed a complete "B" shape was significantly higher in the graduated height staples than in the equal height staples (25.3% versus 10.0%, p = 0.000). No considerable difference was found between the two groups concerning postoperative complications. No bronchopleural fistula was observed.
CONCLUSIONS: The graduated height staples had significantly higher scores of the staple formation on the bronchia stump than the equal height staples. From the clinical point of view, both the equal height staples and the graduated height staples showed acceptable performance in the stapling ability.
Copyright © 2017 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28551048     DOI: 10.1016/j.athoracsur.2017.02.070

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  6 in total

1.  Stapled fascial suture: ex vivo modeling and clinical implications.

Authors:  Enrico Lauro; Ilaria Corridori; Lorenzo Luciani; Alberto Di Leo; Alberto Sartori; Jacopo Andreuccetti; Diletta Trojan; Giovanni Scudo; Antonella Motta; Nicola M Pugno
Journal:  Surg Endosc       Date:  2022-05-16       Impact factor: 4.584

2.  Prospective Randomized Comparison of Linear Endostaplers During Laparoscopic Sleeve Gastrectomy.

Authors:  Yannis Raftopoulos; Shruthi Rajkumar; Elana Davidson; Pavlos Papasavas
Journal:  Obes Surg       Date:  2022-08-16       Impact factor: 3.479

3.  Customized airway stenting for bronchopleural fistula after pulmonary resection by interventional technique: single-center study of 148 consecutive patients.

Authors:  Xinwei Han; Meipan Yin; Lei Li; Ming Zhu; Kewei Ren; Yu Qi; Xiangnan Li; Gang Wu
Journal:  Surg Endosc       Date:  2018-03-30       Impact factor: 4.584

4.  Short-Term Clinical Outcomes of Robotic Sleeve Gastrectomy in Super- and Super-Superobese Patients With and Without Tissue Reinforcement. A Prospective Cohort Matched Study.

Authors:  Antonio Gangemi; Daniel F Suarez
Journal:  Obes Surg       Date:  2021-07-31       Impact factor: 4.129

5.  Modified silicone stent for the treatment of post-surgical bronchopleural fistula: a clinical observation of 17 cases.

Authors:  Junli Zeng; Xuemei Wu; Zhide Chen; Meihua Zhang; Mingyao Ke
Journal:  BMC Pulm Med       Date:  2021-01-06       Impact factor: 3.317

6.  Micro-CT imaging as a method for comparing perfusion in graduated-height and single-height surgical staple lines.

Authors:  Matthew Eschbach; Gregory M Sindberg; Marisha L Godek; Matthew Nagelschmidt; Nicholas Paquette; Michael Wegener; James Alberino; Jane Mayotte; Amit Vasanji; Andrew M Miesse
Journal:  Med Devices (Auckl)       Date:  2018-08-15
  6 in total