WE-G-BRB-08: TG-51 Calibration of First Commercial MRI-Guided IMRT System in the Presence of 0.35 Tesla Magnetic Field.

S Goddu1, O Pechenaya Green1, S Mutic1.   

Abstract

PURPOSE: The first real-time-MRI-guided radiotherapy system has been installed in a clinic and it is being evaluated. Presence of magnetic field (MF) during radiation output calibration may have implications on ionization measurements and there is a possibility that standard calibration protocols may not be suitable for dose measurements for such devices. In this study, we evaluated whether a standard calibration protocol (AAPM- TG-51) is appropriate for absolute dose measurement in presence of MF.
METHODS: Treatment delivery of the ViewRay (VR) system is via three 15,000Ci Cobalt-60 heads positioned 120-degrees apart and all calibration measurements were done in the presence of 0.35T MF. Two ADCL- calibrated ionization-chambers (Exradin A12, A16) were used for TG-51 calibration. Chambers were positioned at 5-cm depth, (SSD=105cm: VR's isocenter), and the MLC leaves were shaped to a 10.5cm × 10.5 cm field size. Percent-depth-dose (PDD) measurements were performed for 5 and 10 cm depths. Individual output of each head was measured using the AAPM- TG51 protocol. Calibration accuracy for each head was subsequently verified by Radiological Physics Center (RPC) TLD measurements.
RESULTS: Measured ion-recombination (Pion) and polarity (Ppol) correction factors were less-than 1.002 and 1.006, respectively. Measured PDDs agreed with BJR-25 within ±0.2%. Maximum dose rates for the reference field size at VR's isocenter for heads 1, 2 and 3 were 1.445±0.005, 1.446±0.107, 1.431±0.006 Gy/minute, respectively. Our calibrations agreed with RPC- TLD measurements within ±1.3%, ±2.6% and ±2.0% for treatment-heads 1, 2 and 3, respectively. At the time of calibration, mean activity of the Co-60 sources was 10,800Ci±0.1%.
CONCLUSIONS: This study shows that the TG- 51 calibration is feasible in the presence of 0.35T MF and the measurement agreement is within the range of results obtainable for conventional treatment machines. Drs. Green, Goddu, and Mutic served as scientific consultants for ViewRay, Inc. Dr. Mutic is on the clinical focus group for ViewRay, Inc., and his spouse holds shares in ViewRay, Inc.
© 2012 American Association of Physicists in Medicine.

Entities:  

Keywords:  Calibration; Dosimetry; Intensity modulated radiation therapy; Ionization; Ionizing radiation; Magnetic field measurements; Magnetic fields; Radiation therapy; Radiation therapy equipment; Thermoluminescent dosimeters

Year:  2012        PMID: 28519599     DOI: 10.1118/1.4736194

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

Review 1.  Magnetic resonance image guidance in external beam radiation therapy planning and delivery.

Authors:  Ilamurugu Arivarasan; Chandrasekaran Anuradha; Shanmuga Subramanian; Ayyalusamy Anantharaman; Velayudham Ramasubramanian
Journal:  Jpn J Radiol       Date:  2017-06-13       Impact factor: 2.374

2.  Synthetic Computed Tomography Generation from 0.35T Magnetic Resonance Images for Magnetic Resonance-Only Radiation Therapy Planning Using Perceptual Loss Models.

Authors:  Xue Li; Poonam Yadav; Alan B McMillan
Journal:  Pract Radiat Oncol       Date:  2021-08-24

Review 3.  The role of imaging in the clinical practice of radiation oncology for pancreatic cancer.

Authors:  Eugene J Koay; William Hall; Peter C Park; Beth Erickson; Joseph M Herman
Journal:  Abdom Radiol (NY)       Date:  2018-02

4.  Characterization of a 0.35T MR system for phantom image quality stability and in vivo assessment of motion quantification.

Authors:  Daniel L Saenz; Yue Yan; Neil Christensen; Margaret A Henzler; Lisa J Forrest; John E Bayouth; Bhudatt R Paliwal
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

5.  MR-based treatment planning in radiation therapy using a deep learning approach.

Authors:  Fang Liu; Poonam Yadav; Andrew M Baschnagel; Alan B McMillan
Journal:  J Appl Clin Med Phys       Date:  2019-03       Impact factor: 2.102

6.  Characteristics of the Exradin W1 scintillator in the magnetic field.

Authors:  Jeongmin Yoon; Jung-In Kim; Chang Heon Choi; Jong Min Park
Journal:  J Appl Clin Med Phys       Date:  2019-08-28       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.