SU-E-J-142: Gafchromic Film Dosimetry in Fluoroscopy for Dynamic Tumor Tracking Irradiation of the Lung Using XR-SP2 Model - A Phantom Study.

T Shiinoki1,2,3,4, A Sawada1,2,3,4, Y Ishihara1,2,3,4, Y Miyabe1,2,3,4, T Fujimoto1,2,3,4, T Nakai1,2,3,4, Y Matsuo1,2,3,4, T Mizowaki1,2,3,4, M Kokubo1,2,3,4, M Hiraoka1,2,3,4.   

Abstract

PURPOSE: We have recently developed a dynamic tumor tracking irradiation system using Vero4DRT (MHI-Tm2 000). It is needed to create a 4D correlation model between a fiducial marker implanted near a tumor and an external surrogate as a function of time by continuously acquiring both fluoroscopy images and external surrogate signals. The purpose of this study was to propose a new dosimetry method using Gafchromic XR-SP2 films to measure surface dose by fluoroscopy imaging.
METHODS: First, half-value layers (HVLs) were measured using aluminum (Al) thicknesses (15 mm) at 40125 kVp. Subsequently, several films were irradiated using various milliampere second values on a solid water phantom. The surface air kerma were also measured using the chamber to calculate the surface doses under the same condition. Then, the calibration curve of dose vs. pixel values was calculated. Finally, surface dose by fluoroscopy imaging was measured using several pieces of film taped on the chest phantom. Orthogonal X-ray fluoroscopy imaging was simultaneously performed until completion of data acquisition for creating a 4D correlation model. Those films were scanned after irradiation using a flat-bed scanner and converted to dose by calibration curve.
RESULTS: The HVLs for tube voltage within 40125 kVp ranged from 2.35 to 5.98 mm Al. The calibration curve between surface dose and pixel values was reasonably smooth. The differences between the measured and the calibrated doses were less than 3%. The hot spots with the maximum dose of 37.12 mGy were observed around the area overlapped by both fluoroscopic fields.
CONCLUSIONS: We have proposed a new dosimetry method using Gafchromic XR-SP2 films to measure surface dose by fluoroscopy imaging. This phantom study has demonstrated that it may be feasible to assess surface dose to patients during dynamic tumor tracking irradiation in clinic with ease after further investigation. This research was supported by the Japan Society for the Promotion of Science (JSPS) through its Funding Program for World-Leading Innovation R&D on Science and Technology (FIRST Program). Research sponsored in part by Mitsubishi Heavy Industries, Ltd.
© 2012 American Association of Physicists in Medicine.

Entities:  

Keywords:  Aluminium; Calibration; Cancer; Data acquisition; Dosimetry; Fluoroscopy; Lungs; Medical imaging; Surface dynamics; Surface measurements

Year:  2012        PMID: 28518900     DOI: 10.1118/1.4734978

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Quantification of the kV X-ray imaging dose during real-time tumor tracking and from three- and four-dimensional cone-beam computed tomography in lung cancer patients using a Monte Carlo simulation.

Authors:  Mitsuhiro Nakamura; Yoshitomo Ishihara; Yukinori Matsuo; Yusuke Iizuka; Nami Ueki; Hiraku Iramina; Hideaki Hirashima; Takashi Mizowaki
Journal:  J Radiat Res       Date:  2018-03-01       Impact factor: 2.724

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.