A quantitative method to the analysis of MLC leaf position and speed based on EPID and EBT3 film for dynamic IMRT treatment with different types of MLC.

Yinghui Li1, Lixin Chen2, Jinhan Zhu2, Bin Wang2, Xiaowei Liu1.   

Abstract

A quantitative method based on the electronic portal imaging system (EPID) and film was developed for MLC position and speed testing; this method was used for three MLC types (Millennium, MLCi, and Agility MLC). To determine the leaf position, a picket fence designed by the dynamic (DMLC) model was used. The full-width half-maximum (FWHM) values of each gap measured by EPID and EBT3 were converted to the gap width using the FWHM versus nominal gap width relationship. The algorithm developed for the picket fence analysis was able to quantify the gap width, the distance between gaps, and each individual leaf position. To determine the leaf speed, a 0.5 × 20 cm2 MLC-defined sliding gap was applied across a 14 × 20 cm2 symmetry field. The linacs ran at a fixed-dose rate. The use of different monitor units (MUs) for this test led to different leaf speeds. The effect of leaf transmission was considered in a speed accuracy analysis. The difference between the EPID and film results for the MLC position is less than 0.1 mm. For the three MLC types, twice the standard deviation (2 SD) is provided; 0.2, 0.4, and 0.4 mm for gap widths of three MLC types, and 0.1, 0.2, and 0.2 mm for distances between gaps. The individual leaf positions deviate from the preset positions within 0.1 mm. The variations in the speed profiles for the EPID and EBT3 results are consistent, but the EPID results are slightly better than the film results. Different speeds were measured for each MLC type. For all three MLC types, speed errors increase with increasing speed. The analysis speeds deviate from the preset speeds within approximately 0.01 cm s-1 . This quantitative analysis of MLC position and speed provides an intuitive evaluation for MLC quality assurance (QA).
© 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  EPID; FILM; MLC position; MLC speed

Mesh:

Year:  2017        PMID: 28517613     DOI: 10.1002/acm2.12102

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  4 in total

1.  Evaluation of QA software system analysis for the static picket fence test.

Authors:  Julien Boudet; Léone Aubignac; Amandine Beneux; Frédéric Mazoyer; Igor Bessieres
Journal:  J Appl Clin Med Phys       Date:  2022-05-15       Impact factor: 2.243

2.  Plan quality comparison for cervical carcinoma treated with Halcyon and Trilogy intensity-modulated radiotherapy.

Authors:  Chengqiang Li; Jinhu Chen; Jian Zhu; Guanzhong Gong; Cheng Tao; Zhenjiang Li; Jie Lu; Yong Yin
Journal:  J Cancer       Date:  2019-10-15       Impact factor: 4.207

3.  Examination of conversion method of dose distribution of lung cancer IMRT using fluence reversible calculation function in O-ring type linac and C-type linac.

Authors:  Toshiya Rachi; Raturi Vijay Parshuram; Yuki Tanaka; Haruki Togo
Journal:  Phys Eng Sci Med       Date:  2022-04-19

4.  Effect of setup error in the single-isocenter technique on stereotactic radiosurgery for multiple brain metastases.

Authors:  Hisashi Nakano; Satoshi Tanabe; Satoru Utsunomiya; Takumi Yamada; Ryuta Sasamoto; Toshimichi Nakano; Hirotake Saito; Takeshi Takizawa; Hironori Sakai; Atsushi Ohta; Eisuke Abe; Motoki Kaidu; Hidefumi Aoyama
Journal:  J Appl Clin Med Phys       Date:  2020-10-29       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.