Literature DB >> 28517544

SU-E-T-645: Treatment of Multiple Brain Metastases Using Stereotactic Radiosurgery with Single-Isocenter Volumetric Modulated Arc Therapy: Comparison with Conventional Dynamic Conformal Arc and Static Beam Stereotactic Radiosurgery.

C Huang1, L Ren1, J Kirkpatrick1, Z Wang1.   

Abstract

PURPOSE: To investigate the treatment of multiple brain metastases using stereotactic radiosurgery with single-isocenter volumetric modulated arc therapy (VMAT) compared with conventional multi-isocenter dynamic conformal arc therapy (DCAT) and three-dimensional conformal radiation therapy (3D-CRT).
METHODS: Seventeen patients with 2 to 5 brain metastatic lesions were studied. The number of patients with 5, 4, 3, and 2 lesions were 4, 5, 4, and 4, respectively. For patients treated with DCAT/3D-CRT plans, VMAT plans were retrospectively generated, and vice versa. Single-isocenter set up was employed in VMAT plans while the number of isocenters was proportional to the number of lesions in DCAT/3D-CRT plans. The DCAT/3D-CRT and VMAT plans were generated using iPlan® RT Dose Version 4.1.1 (BrainLAB, Germany) and Eclipse™ Version 8.6 (Varian, USA) treatment planning system, respectively. All plans were designed to be delivered on Novalis Tx™ system (Varian, USA and BrainLAB, Germany), in which the accelerator equipped with a high definition multileaf collimator (HDMLC).
RESULTS: Conformity index for VMAT plans were equivalent to or better than that for DCAT/3D-CRT plans. While VMAT and DCAT/3D-CRT plans were similar in target coverage, quality of coverage for VMAT plans was better. However, the volume receiving 5Gy was 46% larger for VMAT plans. In addition, the distance from individual lesion to the VMAT isocenter has no impact on VMAT plans. Compared with DCAT/3D-CRT plans, the mean monitor units (MU) decreased by 42% and the estimated treatment time decreased by 49% for VMAT plans.
CONCLUSIONS: This work suggests that single-isocenter VMAT is promising for stereotactic radiosurgery in the treatment of multiple brain metastases. Single-isocenter VMAT is able to achieve comparable conformity, target coverage and quality of coverage with significantly superior delivery efficiency.
© 2012 American Association of Physicists in Medicine.

Entities:  

Keywords:  Brain; Conformal radiation therapy; Dosimetry; Medical treatment planning; Multileaf collimators; Radiation treatment; Radiosurgery; Therapeutics

Year:  2012        PMID: 28517544     DOI: 10.1118/1.4735734

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Single-isocenter multiple-target stereotactic radiosurgery for multiple brain metastases: dosimetric evaluation of two automated treatment planning systems.

Authors:  Giorgio Hamid Raza; Luca Capone; Paolo Tini; Martina Giraffa; Piercarlo Gentile; Giuseppe Minniti
Journal:  Radiat Oncol       Date:  2022-07-01       Impact factor: 4.309

2.  Initial Experience With Single-Isocenter Radiosurgery to Target Multiple Brain Metastases Using an Automated Treatment Planning Software: Clinical Outcomes and Optimal Target Volume Margins Strategy.

Authors:  Giuseppe Minniti; Luca Capone; Filippo Alongi; Vanessa Figlia; Barbara Nardiello; Randa El Gawhary; Claudia Scaringi; Federico Bianciardi; Barbara Tolu; Piercarlo Gentile; Sergio Paolini
Journal:  Adv Radiat Oncol       Date:  2020-07-14

Review 3.  The feasibility and benefits of using volumetric arc therapy in patients with brain metastases: a systematic review.

Authors:  Adriana Andrevska; Kellie A Knight; Charlotte A Sale
Journal:  J Med Radiat Sci       Date:  2014-09-26

4.  Dosimetric Effect of Rotational Setup Errors in Single-Isocenter Volumetric-Modulated Arc Therapy of Multiple Brain Metastases.

Authors:  Kasirajan Tamil Selvan; Ganesan Padma; Murugesan Kesavan Revathy; N Arunai Nambi Raj; K Senthilnathan; P Ramesh Babu
Journal:  J Med Phys       Date:  2019 Apr-Jun

Review 5.  Frameless Image-Guided Radiosurgery for Multiple Brain Metastasis Using VMAT: A Review and an Institutional Experience.

Authors:  Samir Abdallah Hanna; Anselmo Mancini; Alisson Henrique Dal Col; Rie Nadia Asso; Wellington Furtado Pimenta Neves-Junior
Journal:  Front Oncol       Date:  2019-08-07       Impact factor: 6.244

6.  In vitro quality assurance of three-dimensional conformal radiotherapy mono-isocentric plan for simultaneous treatment of two targets.

Authors:  Misba Hamid Baba; Benoy Kumar Singh
Journal:  Int J Health Sci (Qassim)       Date:  2022 Jan-Feb

7.  Radiosurgery for multiple brain metastases using volumetric modulated arc therapy: a single institutional series.

Authors:  Rie Nadia Asso; Anselmo Mancini; Daniel Moore Freitas Palhares; Wellington Furtado Pimenta Palhares Neves Junior; Gustavo Nader Marta; João Luis Fernandes da Silva; Bibiana Ferreira Gouvea Ramos; Rafael Gadia; Samir Abdallah Hanna
Journal:  Rep Pract Oncol Radiother       Date:  2022-09-19

8.  The use of collimator angle optimization and jaw tracking for VMAT-based single-isocenter multiple-target stereotactic radiosurgery for up to six targets in the Varian Eclipse treatment planning system.

Authors:  Lauren M M Pudsey; Dean Cutajar; Alex Wallace; Anastasia Saba; Laurel Schmidt; Andrej Bece; Catherine Clark; Yoshiya Yamada; Giordano Biasi; Anatoly Rosenfeld; Joel Poder
Journal:  J Appl Clin Med Phys       Date:  2021-07-19       Impact factor: 2.102