SU-E-T-446: Effect of Dose Calculation Grid Size Variability on the Specification of Spinal Cord Dose Tolerance for Spinal Stereotactic Body Radiation Therapy.

S Gros1, M Descovich1, I Barani1, L Ma1.   

Abstract

PURPOSE: To investigate the effect of the dose grid resolution variability on the spinal cord dose for spinal SBRT treatments.
METHOD: 10 CyberKnife plans were selected for the proximity of the PTV to the spinal cord. All dose distributions were calculated with Monte Carlo using high spatial resolution and minimal relative uncertainty. The plans were renormalized to a 16 Gy prescription dose and to ensure a target coverage > 95% in order to compare the calculated dose distributions. Each dose matrix was resampled 12 times, covering a grid resolution range of 0.95 mm to 13.52 mm. The spinal cord DVHs were generated for each resampled dose grid. The variations of the maximum point dose (DmaxCord) and dose-coverage to partial volumes (D[V]) up to 5 cc were investigated against the grid resolution.
RESULTS: The mean variation of DmaxCord with grid resolution is characterized by an inverse power law, with a sharp initial decrease leading to potentially large underestimates of DmaxCord (24%, 40% and 55% at resolutions of 2 mm, 4 mm and 8 mm). The variability of mean D[V] values decreases from smaller to larger grid resolutions, however large disparities are observed between patient plans. We introduced the variability threshold volume (Vth) as a constraint to express the dose coverage independently from the grid resolution. For resolutions up to 8 mm, the mean Vth value is (0.96±0.10) cc with a corresponding dose coverage of (26± 12) % relative to the initial DmaxCord value.
CONCLUSIONS: Dose distributions calculated with grid resolutions larger than 2 mm could Result in significant underestimates of DmaxCord. Furthermore, the sensitivity of the dose coverage to grid resolution variability is patient dependent. Consequently, a specified cord dose tolerance should be quoted at a particular grid resolution uniformly adopted between institutions; 2 mm or less is an appropriate value.
© 2012 American Association of Physicists in Medicine.

Entities:  

Keywords:  Dosimetry; Radiation therapy; Radiation treatment; Spatial resolution

Year:  2012        PMID: 28517484     DOI: 10.1118/1.4735535

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Study on the ability of 3D gamma analysis and bio-mathematical model in detecting dose changes caused by dose-calculation-grid-size (DCGS).

Authors:  Han Bai; Sijin Zhu; Xingrao Wu; Xuhong Liu; Feihu Chen; Jiawen Yan
Journal:  Radiat Oncol       Date:  2020-07-06       Impact factor: 3.481

2.  Dosimetric comparison of Acuros XB with collapsed cone convolution/superposition and anisotropic analytic algorithm for stereotactic ablative radiotherapy of thoracic spinal metastases.

Authors:  Heming Zhen; Brian Hrycushko; Huichen Lee; Robert Timmerman; Arnold Pompoš; Strahinja Stojadinovic; Ryan Foster; Steve B Jiang; Timothy Solberg; Xuejun Gu
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.