| Literature DB >> 28510374 |
Thomas Haenel1,2,3, Berenika Hausnerová3,4, Johannes Steinhaus3,2, Ing Bernhard Moeginger2.
Abstract
This study investigates two technically simple methods to determine the irradiance distribution of light curing units that governs the performance of a visible-light curing resin-based composites. Insufficient light irradiation leads to under-cured composites with poor mechanical properties and elution of residual monomers. The unknown irradiance distribution and its effect on the final restoration are the main critical issues requiring highly sophisticated experimental equipment. The study shows that irradiance distributions of LCUs can easily be determined qualitatively with generally available equipment. This significantly helps dentists in practices to be informed about the homogeneity of the curing lights. CopyrightKeywords: Depth Of Cure; Irradiance Distribution; Laser-Beam Profiler; Light Curing Units; Single Lens Reflex Camera; iPad
Mesh:
Substances:
Year: 2016 PMID: 28510374 DOI: 10.1922/EJPRD_1513Haenel06
Source DB: PubMed Journal: Eur J Prosthodont Restor Dent ISSN: 0965-7452