Literature DB >> 28509509

Fracture of Zirconia Abutment with Metallic Insertion on Anterior Single Titanium Implant with Internal Hexagon: Retrieval Analysis of a Failure.

Marco Antonio Bottino1, Regina Furbino Villefort Rocha1, Lilian Costa Anami1, Mutlu Özcan2, Renata Marques de Melo1.   

Abstract

This case report presents a retrieval analysis of a screw-retained one-piece restoration with the veneering ceramic fired directly onto the zirconia abutment that fractured during insertion. A patient who experienced root fracture of a maxillary left central incisor received a titanium implant on the same day as extraction. After delayed implant loading, a two-piece zirconia abutment with metallic insertion was customized. Upon installation, a horizontal fracture of the crown just above the metallic portion was detected. The retrieval analysis of a fractured zirconia abutment showed crack formation and diffusion of glaze material that expanded the crack line during firing. Copyright
© 2016 Dennis Barber Ltd.

Entities:  

Keywords:  Dental Implant-Abutment Design; Dental Implants; Dental Porcelain; Periprosthetic Fractures; Prosthesis Failure Analyses; Zirconia

Mesh:

Substances:

Year:  2016        PMID: 28509509     DOI: 10.1922/EJPRD_01556Bottino05

Source DB:  PubMed          Journal:  Eur J Prosthodont Restor Dent        ISSN: 0965-7452


  3 in total

1.  Influence of different restorative materials on the stress distribution in dental implants.

Authors:  Carlos-Eduardo Datte; João-Paulo-Mendes Tribst; Amanda-Maria-de Oliveira Dal Piva; Renato-Sussumu Nishioka; Marco-Antonio Bottino; Alexandre-Duarte M Evangelhista; Fabrício M de M Monteiro; Alexandre-Luiz-Souto Borges
Journal:  J Clin Exp Dent       Date:  2018-05-01

2.  Survival Probability, Weibull Characteristics, Stress Distribution, and Fractographic Analysis of Polymer-Infiltrated Ceramic Network Restorations Cemented on a Chairside Titanium Base: An In Vitro and In Silico Study.

Authors:  João P M Tribst; Amanda M O Dal Piva; Alexandre L S Borges; Lilian C Anami; Cornelis J Kleverlaan; Marco A Bottino
Journal:  Materials (Basel)       Date:  2020-04-16       Impact factor: 3.623

3.  Fracture strength analysis of titanium insert-reinforced zirconia abutments according to the axial height of the titanium insert with an internal connection.

Authors:  Seung-Rye Song; Kyeong-Mee Park; Bock-Young Jung
Journal:  PLoS One       Date:  2021-04-01       Impact factor: 3.240

  3 in total