Comparison of longevity and clinical outcomes of implantable cardioverter-defibrillator leads among manufacturers.

Satoshi Kawada1, Nobuhiro Nishii2, Yoshimasa Morimoto3, Akihito Miyoshi3, Motomi Tachibana3, Hiroyasu Sugiyama3, Koji Nakagawa3, Atsuyuki Watanabe3, Hiroshi Morita2, Hiroshi Ito3.   

Abstract

BACKGROUND: An early failure of the Biotronik Linox S/SD implantable cardioverter-defibrillator (ICD) lead has been reported. We have also experienced several cases with early failure of Linox leads.
OBJECTIVE: Our aim was to assess the longevity of Linox S/SD (Biotronik, Berlin, Germany) compared with Sprint Fidelis (Medtronic, Minneapolis, MN), Sprint Quattro (Medtronic), and Endotak Reliance (Boston Scientific, Natick, Massachusetts) leads.
METHODS: We retrospectively reviewed patients who had undergone implantation of Linox S/SD (n = 90), Sprint Fidelis (n = 37), Sprint Quattro (n = 27), or Endotak Reliance (n = 50) leads between June 2000 and December 2013 at our hospital. Variables associated with lead failure were assessed by the Kaplan-Meier method and Cox survival modeling.
RESULTS: Failure rates of Linox, Sprint Fidelis, and Endotak leads were 3.2%/year (7-year survival rate, 81.0%), 3.4%/year (7-year survival rate, 77.2%), and 0.61%/year (7-year survival rate, 95.8%), respectively. No lead failure was found with Sprint Quattro leads. The survival probability of Linox leads was significantly lower than that of Endotak leads (P = .049) and comparable to that of Sprint Fidelis leads (P = .69). In univariate analysis, age was the only predictor of Linox lead failure. Patients <58 years old were at significantly increased risk of lead failure compared with patients ≥58 years old (hazard ratio, 9.0; 95% confidence interval, 1.13-71.3; P = .037).
CONCLUSION: In our single-center experience, the survival rate of Linox leads was unacceptably low. The only predictor of Linox lead failure was age at implantation. This is the first description of a lower survival rate for Linox leads in an Asian population.
Copyright © 2017 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Endotak Reliance; Implantable cardioverter-defibrillator; Linox; Longevity of ICD lead; Sprint Fidelis; Sprint Quattro

Mesh:

Year:  2017        PMID: 28502870     DOI: 10.1016/j.hrthm.2017.05.020

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  1 in total

1.  First time description of early lead failure of the Linox Smart lead compared to other contemporary high-voltage leads.

Authors:  Vanessa Weberndörfer; Tobias Nyffenegger; Ian Russi; Miriam Brinkert; Benjamin Berte; Stefan Toggweiler; Richard Kobza
Journal:  J Interv Card Electrophysiol       Date:  2018-05-01       Impact factor: 1.900

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.