Vitaly Panchuk1, Dmitry Kirsanov2, Ekaterina Oleneva3, Valentin Semenov4, Andrey Legin1.
Abstract
The procedure for transfer of calibration models between different analytical methods is suggested. It is based on the direct standardization (DS) algorithm earlier suggested for data conversion between physically different instruments of the same type. As a result, multivariate regression models obtained e.g. for NIR spectroscopic measurements in one wavelength range can be successfully applied for predictions from the data obtained with another NIR spectrometer in another wavelength range. The performance of the suggested method was tested with two simulated and two real datasets. In the latter case calibration models constructed for energy-dispersive X-ray fluorescence, UV-Vis spectrometry and NIR spectrometry were addressed. The observed performance of the method implies that it can have a broad range of possible applications in analytical chemistry; some of them are suggested in the paper.Keywords: Calibration model; Calibration transfer; Multivariate regression
Year: 2017 PMID: 28501196 DOI: 10.1016/j.talanta.2017.04.039