A comparative study using thrombin generation and three different INR methods in patients on Vitamin K antagonist treatment.

N Riva1, K Vella2, S Meli3, K Hickey4, D Zammit2, C Calamatta3, M Makris4, S Kitchen4, W Ageno5, A Gatt1,2,3.   

Abstract

INTRODUCTION: Vitamin K antagonist (VKA) treatment requires routine monitoring using the international normalized ratio (INR). However, different INR assays may vary in their results. The aim of this study was to assess the agreement of three different INR methods, compared with thrombin generation, in patients on VKA treatment.
METHODS: Sixty patients attending the Anticoagulation Clinic at Mater Dei Hospital (Msida, Malta) for VKA monitoring between August and September 2015 were enrolled. The INR was tested using a point-of-care (POC) device (CoaguChek XS Plus, Roche Diagnostics) for both capillary and venous blood samples, a photo-optical (Sysmex CS-2100i/CA-1500, Siemens) and a mechanical clot detection system (Thrombolyzer XRC, Behnk Elektronik). All assays used human recombinant thromboplastin as reagent. Thrombin generation was performed using the calibrated automated thrombogram.
RESULTS: There was a negative curvilinear correlation between the endogenous thrombin potential and different INR assays (r≤-.75) and a strong positive linear correlation between the CoaguChek XS Plus on capillary samples and the other INR methodologies (r≥.96).
CONCLUSION: All different INR assays showed good correlation with the thrombin generation potential. The POC INR showed one of the highest correlation coefficients with thrombin generation, confirming the POC devices as an accurate, valid alternative to laboratory INR in VKA patients.
© 2017 John Wiley & Sons Ltd.

Entities:  

Keywords:  accuracy; international normalized ratio; point-of-care systems; thrombin generation; warfarin

Mesh:

Substances:

Year:  2017        PMID: 28500649     DOI: 10.1111/ijlh.12668

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  3 in total

1.  Patients' satisfaction associated with portable coagulometers for warfarin monitoring: a cross-sectional study.

Authors:  Nicoletta Riva; Christian Borg Xuereb; Walter Ageno; Michael Makris; Alex Gatt
Journal:  Blood Transfus       Date:  2020-06-04       Impact factor: 3.443

2.  Thrombolysis in stroke patients: Comparability of point-of-care versus central laboratory international normalized ratio.

Authors:  Ramona C Dolscheid-Pommerich; Sarah Dolscheid; Lars Eichhorn; Birgit Stoffel-Wagner; Ingo Graeff
Journal:  PLoS One       Date:  2018-01-10       Impact factor: 3.240

Review 3.  A Review of the Extraction and Determination Methods of Thirteen Essential Vitamins to the Human Body: An Update from 2010.

Authors:  Yuan Zhang; Wei-E Zhou; Jia-Qing Yan; Min Liu; Yu Zhou; Xin Shen; Ying-Lin Ma; Xue-Song Feng; Jun Yang; Guo-Hui Li
Journal:  Molecules       Date:  2018-06-19       Impact factor: 4.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.