Literature DB >> 28495299

A randomized comparison of a novel nitinol-frame low-dose-copper intrauterine contraceptive and a copper T380S intrauterine contraceptive.

Matthew F Reeves1, Bob H Katz2, Juan M Canela3, Mark J Hathaway4, Michael G Tal2.   

Abstract

OBJECTIVE: We sought to compare VeraCept (VC175), a novel nitinol intrauterine contraceptive (IUC) with 175 square-mm of copper surface area, to a copper T380S IUC. STUDY
DESIGN: We enrolled parous women into a randomized subject-blinded comparison of VC175 and a copper T380S in a 2:1 fashion at a single clinic. The primary outcomes were total adverse events and continuation at 12 months. We also examined pain on insertion, ease of placement, expulsion, tolerability and pregnancy. Subjective ratings were on a 5-point Likert scale (0, no pain to 5, worst pain). We followed subjects through 24-month follow-up.
RESULTS: We enrolled 300 women with 199 randomized to VC175 and 101 to the T380S. Insertion was successful in 198 subjects for VC175 and 100 for the T380S. Mean age was 25 years (range 18, 41), and median parity was 2 (range 1, 8), with 39% having only had Cesarean deliveries. No subjects developed clinical infection or reported serious adverse events. In the VC175 and T380S groups, mean pain at insertion was 1.4 and 2.4, respectively (p<.01). At the 12-month primary endpoint for VC175 and T380S, respectively, continuation was 84% and 68% (p<.002) with expulsions in 5.0% and 12.0% (p<.05) and removal for pain/bleeding in 3.5% and 17.0% (p<.01). At the 24-month visit for VC175 and T380S, respectively, continuation was 77% and 62% (p<.02 by log-rank). One ectopic pregnancy was identified at the 12-month follow-up in a VC175 user. No other pregnancies were diagnosed. With 297.3 and 132.4 woman-years, pregnancy rates were 0.3 and 0.0 per 100 woman-years for VC175 and T380S, respectively.
CONCLUSIONS: VC175 resulted in less pain at insertion, fewer expulsions and higher total continuation than the T380S, with similar contraceptive efficacy. IMPLICATIONS: VC175 is a promising new intrauterine copper contraceptive on a nitinol frame that warrants further clinical trials.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Copper; Insertion; Intrauterine contraception; Nitinol; Randomized controlled trial

Mesh:

Substances:

Year:  2017        PMID: 28495299     DOI: 10.1016/j.contraception.2017.04.002

Source DB:  PubMed          Journal:  Contraception        ISSN: 0010-7824            Impact factor:   3.375


  3 in total

1.  Efficacy, Safety, and Tolerability of a New Low-Dose Copper and Nitinol Intrauterine Device: Phase 2 Data to 36 Months.

Authors:  David K Turok; Anita L Nelson; Clint Dart; Courtney A Schreiber; Kevin Peters; Mary Jo Schreifels; Bob Katz
Journal:  Obstet Gynecol       Date:  2020-04       Impact factor: 7.623

2.  Discontinuation rates of intrauterine contraception due to unfavourable bleeding: a systematic review.

Authors:  Dustin Costescu; Rajinder Chawla; Rowena Hughes; Stephanie Teal; Martin Merz
Journal:  BMC Womens Health       Date:  2022-03-21       Impact factor: 2.809

3.  Contraceptive Technologies: Looking Ahead to New Approaches to Increase Options for Family Planning.

Authors:  Lisa B Haddad; John W Townsend; Regine Sitruk-Ware
Journal:  Clin Obstet Gynecol       Date:  2021-09-01       Impact factor: 1.966