Technical Note: Transconvolution based equalization of positron energy effects for the use of 68 Ge/68 Ga phantoms in determining 18 F PET recovery.

George A Prenosil1, Michael Hentschel1, Markus Fürstner1, Thomas Krause1, Thilo Weitzel1, Bernd Klaeser1.   

Abstract

PURPOSE: Avoiding measurement variability from 18 F phantom preparation by using 68 Ge/68 Ga phantoms for the determination of 18 F recovery curves (RC) in clinical quality assurance measurements and for PET/CT site qualification in multicentre clinical trials.
METHODS: RCs were obtained from PET/CT measurements of seven differently sized phantom spheres filled either with 18 F or with 68 Ga. RCs for the respective other isotope were then determined by two different methods: In the first method, images were convolved with positron range transconvolution functions derived from positron annihilation distributions found in literature. This method generated recasted images matching images using the respective other isotope. In the second method, the PET/CT system's isotope independent (intrinsic) point spread function was determined from said phantom measurements and convolved with numerical representations simulating hot spheres filled with the respective other isotope. These simulations included the isotope specific positron annihilation distributions. Recovered activity concentrations were compared between recasted images, simulated images, and the originally acquired images.
RESULTS: 18 F and 68 Ga recovery was successfully determined from image acquisitions of the respective opposite isotope as well as from the simulations. 68 Ga RCs derived from 18 F data had a normalized root-mean-square deviation (NRMSD) from real 68 Ga measurements of 0.019% when using the first method and of 0.008% when using the second method. 18 F RCs derived from 68 Ga data had a NRMSD from real 18 F measurements of 0.036% when using the first method and of 0.038% when using the second method.
CONCLUSIONS: Applying the principles of transconvolution, 18 F RCs can be recalculated from 68 Ga phantom measurements with excellent accuracy. The maximal additionally introduced error was below 0.4% of the error currently accepted for RCs in the site qualification of multicentre clinical trials by the EARL program of the European Association of Nuclear Medicine (EANM). Therefore, our methods legitimately allow for the use of long-lived solid state 68 Ge/68 Ga phantoms instead of manually prepared 18 F phantoms to characterize comparability of 18 F measurements across different imaging sites or of longitudinal 18 F measurements at a single PET/CT system.
© 2017 The Authors. Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Keywords:  PET/CT; partial volume effect; positron range function; quantitative imaging; recovery curve; transconvolution

Mesh:

Year:  2017        PMID: 28494090     DOI: 10.1002/mp.12330

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  3D printing of radioactive phantoms for nuclear medicine imaging.

Authors:  Tilman Läppchen; Lorenz P Meier; Markus Fürstner; George A Prenosil; Thomas Krause; Axel Rominger; Bernd Klaeser; Michael Hentschel
Journal:  EJNMMI Phys       Date:  2020-04-22

2.  EARL compliance measurements on the biograph vision Quadra PET/CT system with a long axial field of view.

Authors:  George A Prenosil; Michael Hentschel; Thilo Weitzel; Hasan Sari; Kuangyu Shi; Ali Afshar-Oromieh; Axel Rominger
Journal:  EJNMMI Phys       Date:  2022-04-08

3.  Towards guidelines to harmonize textural features in PET: Haralick textural features vary with image noise, but exposure-invariant domains enable comparable PET radiomics.

Authors:  George Amadeus Prenosil; Thilo Weitzel; Markus Fürstner; Michael Hentschel; Thomas Krause; Paul Cumming; Axel Rominger; Bernd Klaeser
Journal:  PLoS One       Date:  2020-03-16       Impact factor: 3.240

4.  Standardisation of conventional and advanced iterative reconstruction methods for Gallium-68 multi-centre PET-CT trials.

Authors:  Georgios Krokos; Lucy C Pike; Gary J R Cook; Paul K Marsden
Journal:  EJNMMI Phys       Date:  2021-07-17
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.