Quality of routine diagnostic abdominal images generated from a novel detector-based spectral CT scanner: a technical report on a phantom and clinical study.

Mojgan Hojjati1, Steven Van Hedent1, Negin Rassouli1, Curtis Tatsuoka1, David Jordan1, Amar Dhanantwari2, Prabhakar Rajiah3,4.   

Abstract

PURPOSE: To evaluate the image quality of routine diagnostic images generated from a novel detector-based spectral detector CT (SDCT) and compare it with CT images obtained from a conventional scanner with an energy-integrating detector (Brilliance iCT), Routine diagnostic (conventional/polyenergetic) images are non-material-specific images that resemble single-energy images obtained at the same radiation,
METHODS: ACR guideline-based phantom evaluations were performed on both SDCT and iCT for CT adult body protocol. Retrospective analysis was performed on 50 abdominal CT scans from each scanner. Identical ROIs were placed at multiple locations in the abdomen and attenuation, noise, SNR, and CNR were measured. Subjective image quality analysis on a 5-point Likert scale was performed by 2 readers for enhancement, noise, and image quality.
RESULTS: On phantom studies, SDCT images met the ACR requirements for CT number and deviation, CNR and effective radiation dose. In patients, the qualitative scores were significantly higher for the SDCT than the iCT, including enhancement (4.79 ± 0.38 vs. 4.60 ± 0.51, p = 0.005), noise (4.63 ± 0.42 vs. 4.29 ± 0.50, p = 0.000), and quality (4.85 ± 0.32, vs. 4.57 ± 0.50, p = 0.000). The SNR was higher in SDCT than iCT for liver (7.4 ± 4.2 vs. 7.2 ± 5.3, p = 0.662), spleen (8.6 ± 4.1 vs. 7.4 ± 3.5, p = 0.152), kidney (11.1 ± 6.3 vs. 8.7 ± 5.0, p = 0.033), pancreas (6.90 ± 3.45 vs 6.11 ± 2.64, p = 0.303), aorta (14.2 ± 6.2 vs. 11.0 ± 4.9, p = 0.007), but was slightly lower in lumbar-vertebra (7.7 ± 4.2 vs. 7.8 ± 4.5, p = 0.937). The CNR of the SDCT was also higher than iCT for all abdominal organs.
CONCLUSION: Image quality of routine diagnostic images from the SDCT is comparable to images of a conventional CT scanner with energy-integrating detectors, making it suitable for diagnostic purposes.

Entities:  

Keywords:  Blended; Diagnostic; Dual energy; Spectral

Mesh:

Year:  2017        PMID: 28493070     DOI: 10.1007/s00261-017-1170-z

Source DB:  PubMed          Journal:  Abdom Radiol (NY)


  4 in total

1.  The application of dual-layer spectral detector computed tomography in solitary pulmonary nodule identification.

Authors:  Qingyun Wen; Yong Yue; Jin Shang; Xiaomei Lu; Lu Gao; Yang Hou
Journal:  Quant Imaging Med Surg       Date:  2021-02

Review 2.  Spectral detector CT applications in advanced liver imaging.

Authors:  Noor Fatima Majeed; Marta Braschi Amirfarzan; Christoph Wald; Jeremy R Wortman
Journal:  Br J Radiol       Date:  2021-05-28       Impact factor: 3.629

3.  The optimal monoenergetic spectral image level of coronary computed tomography (CT) angiography on a dual-layer spectral detector CT with half-dose contrast media.

Authors:  Xin Huang; Sizhe Gao; Yue Ma; Xiaomei Lu; Zheng Jia; Yang Hou
Journal:  Quant Imaging Med Surg       Date:  2020-03

Review 4.  Detector-based spectral CT with a novel dual-layer technology: principles and applications.

Authors:  Negin Rassouli; Maryam Etesami; Amar Dhanantwari; Prabhakar Rajiah
Journal:  Insights Imaging       Date:  2017-10-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.