KEY TEACHING POINTSApical hypertrophy is a unique disease with important considerations for device optimization. Progressive V4 R-wave reduction and ST-segment normalization are likely markers of scarring of the hypertrophic apical tissues impacting implantable cardioverter-defibrillator sensing during ventricular fibrillation.An implantable cardioverter-defibrillator sensing lead should be placed in normal tissues once electrocardiogram changes are noted and low-amplitude fractionated activity is recorded during defibrillation testing. Routine defibrillation testing every 2–3 years in patients with atypical anatomy should be considered.Good R-wave sensing in sinus may not necessarily indicate reliable sensing during ventricular fibrillation in patients with abnormal anatomy.Although there is a paucity of randomized trials evaluating outcomes after primary-prevention implantable cardioverter-defibrillator (ICD) placement for patients with hypertrophic cardiomyopathy, guidelines recommend ICD placement for certain patients with the disease. Apical hypertrophic cardiomyopathy (AHCM) is one pattern of disease manifestation with unique characteristics. Hypertrophy involves predominantly the apex of the left ventricle and at times the right ventricle. Whereas some have suggested a benign course of disease in cases of AHCM, there are a few reports of patients developing sustained ventricular arrhythmias and sudden cardiac death.4, 5, 6, 7 We present a case of ventricular fibrillation (VF) and ICD failure because of sensing failure and subsequent ICD manipulation in a patient with AHCM. This case brings to light longitudinal changes in the hypertrophic tissues, electrical changes during VF, and the need for interval testing to ensure proper device function in the setting of tissue remodeling over time.
Case report
A 59-year-old male skier with an athletic lifestyle with known AHCM suffered an out-of-hospital VF arrest. Nine years prior to the VF arrest, he had undergone ICD implantation for primary prevention of sudden cardiac death. His past medical history included isthmus-dependent flutter with successful cavotricuspid isthmus ablation; paroxysmal atrial fibrillation; and Addison disease. He had no known history of sustained ventricular arrhythmias and no family history of sudden cardiac death. At the initial presentation 12 years prior to the event, the electrocardiogram (ECG) showed V4 R wave of 55 mm (5.5 mV), QRS width of 96 msec, and 5 mm precordial ST depression with deeply inverted T waves measuring 20 mm (2 mV), the pattern often seen in AHCM.7, 8 Transthoracic echocardiography showed severe left ventricular apical hypertrophy (13 mm thickness of the septum and posterior walls, but 23 mm at the apex) without any apical aneurysm (Figure 1A).
Figure 1
A: Echocardiographic appearance of this patient’s left ventricular apical hypertrophy. B: Progressive electrical changes over time per surface electrocardiograms (ECGs) (N = 47). R-wave amplitude and T-wave amplitude both declined over time. Units are in mV (millimeters have been converted accordingly from each ECG’s calibration). Examples of V4 complexes are displayed on the right (R-wave reduction −37%, ST −96%).
He underwent implantation of a Guidant ICD at age 50. Coil and sensing lead were placed in the right ventricular (RV) apex, with an R wave of 11 mV, impedance of 800 ohms, and pacing threshold of 1 volt. Active fixation lead was placed in the right atrial appendage with a P wave of 2.6 mV, impedance of 956 ohms, and pacing threshold of 1.2 volts. Defibrillator testing was performed, with 21 J from his device converting him to normal sinus rhythm with impedance of 43 ohms. There appeared to be proper detection with no undersensing.After 4 years of regular follow-up and no shocks, the patient returned for generator change secondary to battery depletion. He had a new generator (Boston Scientific TELIGEN IS-1/DF-1-DR) implanted and connected to his previously placed leads. Testing at that time revealed that the RV apical lead had an R wave of 16.6 mV, impedance of 458 ohms, pacing threshold 0.8 volts. Following the generator change, the patient again underwent defibrillation testing (DFT). Initially there was failure to sense owing to fallout and the patient received external defibrillation at 250 J. The device was tested again with increased sensitivity to 0.5 mV. After VF induction, 31 J from the device converted the patient cleanly to normal sinus rhythm with impedance of 39 ohms. As measured by the RV lead, the maximal peak-to-peak amplitude of the VF was 15 mV. It is worth noting that on careful review of the electrograms there are a few instances of undersensed low-amplitude fractionated activity, but the fractionated activity appeared to be of a low burden of the total activity and did not interfere with the deviceʼs recognition of VF.The patient did well for years with no ICD therapy and routine follow-up. Unfortunately, he suffered an out-of-hospital arrest 5 years after his generator change, at age 59. Civilians found him unresponsive and cardiopulmonary resuscitation was initiated. Emergency medical services arrived and administered 2 external shocks for VF, with return of spontaneous circulation. After induced hypothermia per standard protocol, he regained total neurologic function by 1 month. At this time, the ECG showed V4 R wave of 20 mm, QRS width 106 msec, no ST-segment depression, and T wave now upright (Figure 1B).The ICD data during the arrest was analyzed and showed baseline R wave of 15 mV while in sinus, which was similar to prearrest measurements. Nonetheless, the device shows development of ventricular tachycardia that quickly degenerated into very fine VF. Shortly after VF onset, the maximal peak-to-peak amplitude was approximately 13 mV. Within 3.5 seconds the activity became highly fractionated, with segments of much lower amplitude consistently less than 3 mV (minimal <1 mV) and detection gaps of 443 msec, leading to counter resetting and failure to shock (Figure 2). Eventually the device administered a 41 J shock at approximately 6 minutes into the episode, with brief return of sinus rhythm followed by rapid recurrence of VF. Following another ineffective shock by his device, he received external defibrillation by Emergency Medical Services. In total, he experienced approximately 11 minutes of VF as recorded by his device.
Figure 2
Electrogram as recorded by the device at the onset of ventricular fibrillation (VF) demonstrates failure to sense VF.
After clinical improvement, the patient underwent electrophysiology study with defibrillator testing. Our hypothesis was that progressive apical fibrosis had contributed to undersensing of VF by his apical lead, and our hope was to identify the optimal location of an alternative sensing lead. A quadripolar and a duodeca catheter were placed to simultaneously record the apex and mid and high septum. During induced VF the apex exhibited marked fractionated activity with maximal amplitude of 1.5 mV with a transition to organized, higher-amplitude signal at the upper septum of 4.2 mV as measured by the catheters (Figure 3A and B). Based on these data we placed an active-fixation sense/pace lead with an 11-mm separation (BSC RELIANCE) at the mid RV septum as noted in the posteroanterior chest radiograph image (Figure 3C).
Figure 3
A: Electrograms prior to placing the right ventricular (RV) septal lead. B: Catheter placement is shown with a 20-pole mapping catheter in the right ventricle that has been advanced as distally as possible to approximate sensing at the apical lead and a 4-pole mapping catheter to approximate the high septal activity. The apical activity (DD1–2) is fractionated, low-amplitude signal during ventricular fibrillation (VF). However, the same activity is discrete and high-amplitude at the basilar RV septum. C: New RV septal lead with associated electrogram during induced VF with cleaner signal (red) as compared to the old lead (blue), also with activity during VF.
Following connection of the new sensing lead to the ICD (BSC Incepta IS-1/DF-1 DR), testing showed an R wave of 15.4 mV, impedance of 808 ohms, and pacing threshold of 1.6 volts at 0.5 msec. In DFT, he underwent successful conversion from VF to sinus rhythm with 21 J ICD shock with no undersensing. VF maximal peak-to-peak amplitude as measured by the device was 13.5 mV with markedly less fractionated activity as compared to that measured previously by the apical lead.The patientʼs 47 ECGs taken over 12 years leading to the arrest showed progressive reduction of the V4 R-wave amplitude and normalization of ST strain pattern (R wave −37%, ST −96% reduction; correlation of 0.59 and 0.91, respectively) (Figure 1B).
Discussion
This case has a number of interesting implications. It is possible that progression of fibrosis in AHCM can interfere with proper function of sensing leads over time. Compared with other forms of hypertrophic cardiomyopathy, the apical variant tends to have faster progression of fibrosis in the hypertrophied areas of myocardium, as evaluated by extent of late-gadolinium enhancement on cardiac magnetic resonance imaging over time. Our patient is unable to undergo optimal cardiac magnetic resonance imaging of his apex owing to the presence of the ICD lead interfering with the image acquisition, though we suspect that progression of apical scarring and fibrosis occurred, as evidenced by his progressive ECG changes. High QRS amplitude and giant negative T waves in precordial leads, particularly V4, are commonly seen in AHCM. The dynamic changes in the patientʼs ECG morphology likely represents composite tissue remodeling, which caused rapid disorganization during VF, low amplitude, and device undersensing. Interestingly, though this affected detection of electrical activity during VF, R-wave amplitude and sensing was normal during sinus rhythm, as measured by his device.
Conclusion
In an era where DFT is not consistently done, this case supports consideration of routine DFT in patients with hypertrophic cardiomyopathy. In our patient, although the R wave was consistently measured to be adequate during prior testing, undersensing caused device failure. An early sign of impending failure may have been present during his follow-up DFT at the time of generator change in 2010, when initial DFT failed and adjustment in device sensitivity was required. Additionally, the dynamic changes in his ECG as detailed above may have been another sign of the anatomic changes. It is possible that these ECG changes in patients with AHCM will provide a noninvasive marker that should trigger defibrillation testing. Low-voltage, fractionated activity noted during DFT is likely an indication for placing a sensing lead in normal myocardium. Altogether avoiding apical lead placement in patients with AHCM would be a reasonable approach; however, it is likely to increase the DFT in patients with this particular cardiac anatomy. In addition to AHCM, there is often progressive fibrosis over years in other forms of hypertrophic cardiomyopathy as well. Follow-up DFT testing should be considered in all such patients.Despite excellent R-wave sensing during sinus rhythm, the amplitude in VF rapidly deteriorated and became fractionated, resulting in undersensing. The assumption that high-voltage ventricular sensing in sinus will imply reliable sensing in VF is not appropriate with this anatomy. Clearly, more data would be useful in defining optimal device treatment for patients with AHCM.
KEY TEACHING POINTS
Apical hypertrophy is a unique disease with important considerations for device optimization. Progressive V4 R-wave reduction and ST-segment normalization are likely markers of scarring of the hypertrophic apical tissues impacting implantable cardioverter-defibrillator sensing during ventricular fibrillation.
An implantable cardioverter-defibrillator sensing lead should be placed in normal tissues once electrocardiogram changes are noted and low-amplitude fractionated activity is recorded during defibrillation testing. Routine defibrillation testing every 2–3 years in patients with atypical anatomy should be considered.
Good R-wave sensing in sinus may not necessarily indicate reliable sensing during ventricular fibrillation in patients with abnormal anatomy.
Authors: Maria J Eriksson; Brian Sonnenberg; Anna Woo; Paul Rakowski; Thomas G Parker; E Douglas Wigle; Harry Rakowski Journal: J Am Coll Cardiol Date: 2002-02-20 Impact factor: 24.094
Authors: Douglas P Zipes; A John Camm; Martin Borggrefe; Alfred E Buxton; Bernard Chaitman; Martin Fromer; Gabriel Gregoratos; George Klein; Arthur J Moss; Robert J Myerburg; Silvia G Priori; Miguel A Quinones; Dan M Roden; Michael J Silka; Cynthia Tracy; Sidney C Smith; Alice K Jacobs; Cynthia D Adams; Elliott M Antman; Jeffrey L Anderson; Sharon A Hunt; Jonathan L Halperin; Rick Nishimura; Joseph P Ornato; Richard L Page; Barbara Riegel; Silvia G Priori; Jean-Jacques Blanc; Andrzej Budaj; A John Camm; Veronica Dean; Jaap W Deckers; Catherine Despres; Kenneth Dickstein; John Lekakis; Keith McGregor; Marco Metra; Joao Morais; Ady Osterspey; Juan Luis Tamargo; José Luis Zamorano Journal: J Am Coll Cardiol Date: 2006-09-05 Impact factor: 24.094
Authors: Kyle W Klarich; Christine H Attenhofer Jost; Josepha Binder; Heidi M Connolly; Christopher G Scott; William K Freeman; Michael J Ackerman; Rick A Nishimura; A Jamil Tajik; Steve R Ommen Journal: Am J Cardiol Date: 2013-03-27 Impact factor: 2.778
Authors: Giancarlo Todiere; Giovanni Donato Aquaro; Paolo Piaggi; Francesco Formisano; Andrea Barison; Pier Giorgio Masci; Elisabetta Strata; Lorenzo Bacigalupo; Mario Marzilli; Alessandro Pingitore; Massimo Lombardi Journal: J Am Coll Cardiol Date: 2012-09-04 Impact factor: 24.094