| Literature DB >> 28491731 |
Jeremy W Docekal1, Mark C Haigney2, Joseph C Lee1, Matthew Needleman1.
Abstract
Entities:
Keywords: Implantable cardioverter-defibrillator; Inappropriate shock; Post-shock redetection interval; T-wave oversensing; TWave
Year: 2016 PMID: 28491731 PMCID: PMC5419968 DOI: 10.1016/j.hrcr.2016.07.004
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Figure 1Rate histogram from the first through fifth shock. The patient was initially in atrial fibrillation, and the first defibrillation restored sinus rhythm. After the first shock, there are at least 2 different populations of “R” waves seen representing T-wave oversensing. The arrows at the conclusion of the slide show the classic “train tracking” seen in T-wave oversensing.
Figure 2A representative sample of the intracardiacs during the second through fifth defibrillation. During the nominal redetection time, the oversensed T waves are sensed late in the pseudo-“R-R” interval. The TWave algorithm does not classify these short coupled pseudo-“R-R” intervals as T waves, resulting in inappropriate T-wave oversensing. This failure of the TWave algorithm, coupled with the nominal (12 of 16) redetection times, resulted in the inappropriate shocks.
Methods for prevention of T-wave oversensing
| Manufacturer | Program Name | TWOS Algorithm | Programmable |
|---|---|---|---|
| Biotronik | Enhanced T-wave suppression (TWS) | With enhanced T-wave suppression algorithm enabled, high pass filtering is increased in order to reduce detection of low-frequency T waves. Additionally, the upper threshold for R-wave detection is increased to 75% of the previously measured R wave. | Yes |
| Boston Scientific | Not applicable | Boston Scientific Cognis and Teligen defibrillators avoid TWOS through a narrow-frequency Band Pass filter, which detects R wave signals in the range of 20–85 Hz. Low-frequency T waves are typically below this detection window, and are excluded. | No |
| Medtronic | TWave | Avoids TWOS through the application of a differential frequency filter, which amplifies the R wave signal, as compared to the T wave signal. Additional rate and pattern criteria are applied to confirm T-wave detection | Yes |
| St Jude Medical | SenseAbility | Reduces TWOS by tailoring post-R-wave adjustable sensitivity using the St Jude Medical SenseAbility proprietary algorithm. In addition, St Jude programming provides a low-frequency attenuation filter to exclude T wave signals. | Yes |
TWOS = T-wave oversensing.
KEY TEACHING POINTS
Despite modern T-wave oversensing (TWOS) prevention algorithms, TWOS remains an important clinical issue for physicians following patients with defibrillators. The TWave discrimination algorithm failed to prevent inappropriate shocks owing to late sensed T waves and, potentially, variation in the R-wave and T-wave amplitude post shock. Clinicians who follow defibrillators with TWOS need to be aware of this algorithm and should consider lengthening the redetect intervals to allow the algorithm more time to make a new template post therapy. |