| Literature DB >> 28489356 |
Piia Liigand1, Karl Kaupmees1, Kristjan Haav1, Jaanus Liigand1, Ivo Leito1, Marion Girod2, Rodolphe Antoine3, Anneli Kruve1,4.
Abstract
For the first time, the electrospray ionization efficiency (IE) scales in positive and negative mode are united into a single system enabling direct comparison of IE values across ionization modes. This is made possible by the use of a reference compound that ionizes to a similar extent in both positive and negative modes. Thus, choosing the optimal (i.e., most sensitive) ionization conditions for a given set of analytes is enabled. Ionization efficiencies of 33 compounds ionizing in both modes demonstrate that, contrary to general practice, negative mode allows better sensitivity for 46% of such compounds whereas the positive mode is preferred for only 18%, and for 36%, the results for both modes are comparable.Entities:
Year: 2017 PMID: 28489356 DOI: 10.1021/acs.analchem.7b00096
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986