| Literature DB >> 28480331 |
Stefanie J Hectors1,2, Mathilde Wagner1,2, Idoia Corcuera-Solano1,2, Martin Kang3, Alto Stemmer4, Michael A Boss5, Bachir Taouli1,2.
Abstract
Diagonal diffusion-weighted imaging (dDWI) uses simultaneous maximized application of 3 orthogonal gradient systems as opposed to sequential acquisition in 3 directions in conventional 3-scan trace DWI (tDWI). Several theoretical advantages of dDWI vs. tDWI include reduced artifacts and increased sharpness. We compared apparent diffusion coefficient (ADC) quantification and image quality between monopolar dDWI and tDWI in a dedicated diffusion phantom (b = 0/500/900/2000 s/mm2) and in the abdomen (b = 50/400/800 s/mm2) and pelvis (b = 50/1000/1600 s/mm2) of 2 male volunteers at 1.5 T and 3.0 T. Phantom estimated signal-to-noise ratio (eSNR) was also measured. Two independent observers assessed the image quality on a 5-point scale. In the phantom, image quality was similar between tDWI and dDWI, with equivalent ADC quantification (mean coefficient of variation [CV] between sequences: 1.4% ± 1.2% at 1.5 T and 0.7% ± 0.7% at 3.0 T). Phantom eSNR was similar for both tDWI and dDWI, except for a significantly lower eSNR for b900 of dDWI at 3.0 T (P = .006). In the volunteers, the CV values between tDWI and dDWI were higher than those in the phantom (CV range: abdominal organs, 1.3%-13.3%; pelvic organs, 0.6%-5.7%). A trend toward significant better image quality for dDWI compared with tDWI was observed for b800 (abdomen) at 3.0 T and for b1000 and b1600 (pelvis) at 1.5 T (P = .063 to .066). Our data suggest that dDWI may provide better image quality than tDWI without affecting ADC quantification, needing confirmation in a future clinical study.Entities:
Keywords: apparent diffusion coefficient; diffusion-weighted imaging; phantom; volunteers
Year: 2016 PMID: 28480331 PMCID: PMC5416814 DOI: 10.18383/j.tom.2016.00229
Source DB: PubMed Journal: Tomography ISSN: 2379-1381
Acquisition Parameters for tDWI and dDWI Pulse Sequences in Diffusion Phantom and In Vivo
| Phantom | In Vivo (Abdomen) | In Vivo (Pelvis) | ||||
|---|---|---|---|---|---|---|
| tDWI | dDWI | tDWI | dDWI | tDWI | dDWI | |
| Orientation | Coronal | Coronal | Axial | Axial | Axial oblique | Axial oblique |
| Diffusion scheme | Monopolar | Monopolar | Monopolar | Monopolar | Monopolar | Monopolar |
| Gradient directions [x,y,z] | [1.0, 1.0, −0.5], [1.0, −0.5, 1.0],[−0.5, 1.0, 1.0] | [1.0, 1.0, 1.0] | [1.0, 1.0, −0.5], [1.0, −0.5, 1.0],[−0.5, 1.0, 1.0] | [1.0 1.0 1.0] | [1.0, 1.0, −0.5], [1.0, −0.5, 1.0],[−0.5, 1.0, 1.0] | [1.0, 1.0, 1.0] |
| Read-out | SS SE-EPI | SS SE-EPI | SS SE-EPI | SS SE-EPI | SS SE-EPI | SS SE-EPI |
| b Values (s/mm2) | 0, 500, 900, 2000 | 0, 500, 900, 2000 | 50, 400, 800 | 50, 400, 800 | 50, 1000, 1600 | 50, 1000, 1600 |
| Averages | 4, 4, 4, 4 | 4, 12, 12, 12 | 1, 2, 4 | 3, 6, 12 | 1, 5, 10 | 3, 15, 30 |
| TE (ms) | 110 (1.5 T); 99 (3.0 T) | 106 (1.5 T); 95 (3.0 T) | 73 (1.5 T); 70 (3.0 T) | 70 (1.5 T); 68 (3.0 T) | 83 (1.5 T); 71 (3.0 T) | 79 (1.5 T); 69 (3.0 T) |
| TR (ms) | 10 000 | 10 000 | 5000 | 5000 | 5000 | 5000 |
| Slice thickness (mm) | 5 | 5 | 5 | 5 | 5 | 5 |
| Number of slices | 5 | 5 | 30 | 30 | 20 | 20 |
| Matrix | 128 × 128 | 128 × 128 | 128 × 128 | 128 × 128 | 114 × 114 | 114 × 114 |
| FOV (mm2) | 210 × 210 | 210 × 210 | 370 × 250 | 370 × 250 | 250 × 250 | 250 × 250 |
| GRAPPA acceleration | 2 | 2 | 2 | 2 | 2 | 2 |
| Fat suppression | None | None | SPAIR | SPAIR | SPAIR | SPAIR |
| Acquisition time (min) | 7:10 | 7:10 | 2:12 | 2:12 | 4:27 | 4:27 |
The in vivo acquisitions were performed during free breathing.
Abbreviations: dDWI, diagonal diffusion-weighted imaging; FOV, field of view; GRAPPA, generalized autocalibrating partial parallel acquisition; SPAIR, spectral attenuated inversion recovery; tDWI, 3-scan trace diffusion-weighted imaging; SS SE-EPI, single-shot spin-echo echo planar imaging; TE, echo time; TR, repetition time.
Average ADC and CV Values for Phantom Test-Retest Acquisitions
| Vial No. | PVP % | 1.5 T | 3.0 T | 1.5 T vs. 3.0 T | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| tDWI | dDWI | CV tDWI–dDWI | tDWI | dDWI | CV tDWI–dDWI | tDWI | dDWI | ||||||||||
| ADC Test | ADC Retest | CV Test–Retest | ADC Test | ADC Retest | CV Test–Retest | ADC Test | ADC Retest | CV Test–Retest | ADC Test | ADC Retest | CV Test–Retest | CV 1.5 T–3.0 T | CV 1.5 T–3.0 T | ||||
| 1 | 0 | 1.13 ± 0.01 | 1.13 ± 0.02 | 0.0 | 1.14 ± 0.02 | 1.15 ± 0.02 | 1.0 | 0.1 | 1.13 ± 0.01 | 1.13 ± 0.01 | 0.1 | 1.13 ± 0.01 | 1.14 ± 0.01 | 0.7 | 0.2 | 0.0 | 0.5 |
| 2 | 0 | 1.12 ± 0.02 | 1.13 ± 0.02 | 0.6 | 1.12 ± 0.02 | 1.14 ± 0.02 | 1.3 | 0.2 | 1.13 ± 0.01 | 1.12 ± 0.01 | 0.4 | 1.12 ± 0.01 | 1.15 ± 0.01 | 1.6 | 0.7 | 0.3 | 0.0 |
| 3 | 0 | 1.15 ± 0.01 | 1.17 ± 0.02 | 1.5 | 1.13 ± 0.01 | 1.16 ± 0.02 | 1.4 | 1.0 | 1.14 ± 0.01 | 1.16 ± 0.03 | 1.4 | 1.13 ± 0.01 | 1.17 ± 0.01 | 2.4 | 0.5 | 0.6 | 0.5 |
| 4 | 10 | 0.86 ± 0.01 | 0.88 ± 0.01 | 1.8 | 0.86 ± 0.01 | 0.88 ± 0.01 | 1.6 | 0.1 | 0.86 ± 0.01 | 0.87 ± 0.01 | 0.6 | 0.85 ± 0.01 | 0.87 ± 0.01 | 2.0 | 1.0 | 1.2 | 1.6 |
| 5 | 10 | 0.85 ± 0.01 | 0.88 ± 0.01 | 2.4 | 0.85 ± 0.01 | 0.88 ± 0.01 | 2.5 | 0.3 | 0.84 ± 0.01 | 0.87 ± 0.01 | 2.2 | 0.84 ± 0.01 | 0.88 ± 0.01 | 3.0 | 0.1 | 0.7 | 0.5 |
| 6 | 20 | 0.63 ± 0.01 | 0.64 ± 0.01 | 1.0 | 0.64 ± 0.01 | 0.65 ± 0.01 | 1.5 | 0.9 | 0.63 ± 0.01 | 0.64 ± 0.00 | 0.9 | 0.63 ± 0.01 | 0.64 ± 0.01 | 1.8 | 0.4 | 1.3 | 1.8 |
| 7 | 20 | 0.63 ± 0.00 | 0.65 ± 0.01 | 1.7 | 0.66 ± 0.01 | 0.67 ± 0.01 | 1.4 | 2.3 | 0.62 ± 0.01 | 0.65 ± 0.01 | 2.7 | 0.64 ± 0.01 | 0.68 ± 0.01 | 3.6 | 2.5 | 1.2 | 0.5 |
| 8 | 30 | 0.42 ± 0.01 | 0.44 ± 0.01 | 2.5 | 0.44 ± 0.01 | 0.45 ± 0.01 | 2.7 | 2.1 | 0.43 ± 0.01 | 0.44 ± 0.01 | 1.7 | 0.43 ± 0.01 | 0.45 ± 0.01 | 2.1 | 0.5 | 0.2 | 1.2 |
| 9 | 30 | 0.41 ± 0.01 | 0.43 ± 0.01 | 2.5 | 0.43 ± 0.01 | 0.44 ± 0.01 | 2.1 | 2.3 | 0.41 ± 0.01 | 0.42 ± 0.01 | 2.4 | 0.41 ± 0.01 | 0.43 ± 0.01 | 2.8 | 0.4 | 0.7 | 2.2 |
| 10 | 40 | 0.25 ± 0.01 | 0.25 ± 0.01 | 1.0 | 0.26 ± 0.01 | 0.26 ± 0.01 | 0.0 | 1.7 | 0.24 ± 0.01 | 0.24 ± 0.01 | 1.4 | 0.24 ± 0.01 | 0.25 ± 0.01 | 3.9 | 0.1 | 2.8 | 2.8 |
| 11 | 40 | 0.23 ± 0.01 | 0.24 ± 0.01 | 2.0 | 0.24 ± 0.01 | 0.25 ± 0.01 | 1.8 | 1.5 | 0.23 ± 0.01 | 0.24 ± 0.01 | 1.5 | 0.23 ± 0.01 | 0.24 ± 0.01 | 1.8 | 0.1 | 0.9 | 2.2 |
| 12 | 50 | 0.13 ± 0.02 | 0.12 ± 0.02 | 1.1 | 0.12 ± 0.02 | 0.13 ± 0.02 | 2.1 | 2.1 | 0.13 ± 0.02 | 0.12 ± 0.02 | 4.0 | 0.12 ± 0.02 | 0.12 ± 0.02 | 1.6 | 1.0 | 1.4 | 0.7 |
| 13 | 50 | 0.13 ± 0.01 | 0.13 ± 0.02 | 1.6 | 0.12 ± 0.01 | 0.13 ± 0.01 | 4.8 | 4.2 | 0.12 ± 0.01 | 0.11 ± 0.01 | 1.0 | 0.12 ± 0.01 | 0.12 ± 0.01 | 1.2 | 1.1 | 8.7 | 3.9 |
ADC is expressed in 10−3 mm2/s. CV values (%) mentioned in the table are between test–retest acquisitions, between tDWI and dDWI, and between 1.5 T and 3.0 T.
Abbreviations: ADC, apparent diffusion coefficient; dDWI, diagonal diffusion-weighted imaging; PVP, polyvinylpyrrolidone; tDWI, 3-scan trace diffusion-weighted imaging.
Figure 1.Bland–Altman plots comparing 3-scan trace diffusion-weighted imaging (tDWI) and diagonal diffusion-weighted imaging (dDWI) phantom apparent diffusion coefficient (ADC) measurements at 1.5 T (A) and at 3.0 T (B), illustrating similar ADC quantification between both techniques at both field strengths.
Average eSNR Values in the Diffusion Phantom
| Vial No. | 1.5 T | 3.0 T | ||||||
|---|---|---|---|---|---|---|---|---|
| eSNRtDWI b900 | eSNRdDWI b900 | eSNRtDWI b2000 | eSNRdDWI b2000 | eSNRtDWI b900 | eSNRdDWI b900 | eSNRtDWI b2000 | eSNRdDWI b2000 | |
| 1 | 55.9 ± 2.5 | 46.6 ± 0.9 | 31.7 ± 5.8 | 30.0 ± 1.7 | 49.9 ± 10.8 | 45.4 ± 9.8 | 43.3 ± 9.7 | 35.7 ± 6.3 |
| 2 | 32.9 ± 6.7 | 22.7 ± 1.4 | 24.2 ± 4.6 | 19.6 ± 2.1 | 28.3 ± 0.8 | 30.0 ± 7.3 | 27.1 ± 0.8 | 35.3 ± 7.3 |
| 3 | 23.7 ± 2.2 | 21.7 ± 0.6 | 20.3 ± 4.4 | 19.0 ± 1.2 | 23.3 ± 8.2 | 21.4 ± 8.5 | 18.9 ± 10.2 | 20.1 ± 8.4 |
| 4 | 47.4 ± 1.1 | 40.0 ± 8.9 | 37.0 ± 2.7 | 38.7 ± 7.3 | 64.8 ± 6.3 | 57.0 ± 5.3 | 68.4 ± 10.3 | 61.5 ± 12.5 |
| 5 | 48.5 ± 6.4 | 41.3 ± 4.8 | 45.1 ± 1.6 | 38.6 ± 5.3 | 20.0 ± 0.8 | 20.6 ± 3.5 | 20.6 ± 1.2 | 23.0 ± 6.4 |
| 6 | 67.2 ± 8.6 | 76.6 ± 22.2 | 58.4 ± 8.9 | 54.1 ± 14.8 | 82.5 ± 21.3 | 81.4 ± 9.0 | 98.6 ± 10.7 | 91.9 ± 16.5 |
| 7 | 18.1 ± 0.3 | 18.6 ± 0.7 | 18.6 ± 0.1 | 19.1 ± 1.4 | 18.5 ± 1.8 | 17.7 ± 1.1 | 18.3 ± 1.8 | 17.1 ± 0.2 |
| 8 | 41.0 ± 4.1 | 42.0 ± 1.6 | 30.2 ± 2.8 | 42.2 ± 2.8 | 40.5 ± 4.1 | 40.0 ± 4.9 | 41.2 ± 2.8 | 41.5 ± 3.3 |
| 9 | 27.3 ± 3.0 | 28.3 ± 4.2 | 29.8 ± 1.3 | 29.1 ± 2.8 | 35.9 ± 3.0 | 30.0 ± 2.7 | 38.4 ± 6.1 | 32.7 ± 3.3 |
| 10 | 30.2 ± 2.1 | 32.7 ± 3.0 | 30.5 ± 3.4 | 32.7 ± 0.4 | 40.5 ± 7.1 | 32.3 ± 0.3 | 38.0 ± 2.6 | 31.0 ± 0.7 |
| 11 | 28.9 ± 1.1 | 32.7 ± 4.0 | 30.9 ± 1.6 | 34.4 ± 0.8 | 48.0 ± 0.1 | 42.8 ± 6.4 | 51.4 ± 0.8 | 44.5 ± 8.8 |
| 12 | 46.5 ± 0.8 | 37.1 ± 6.2 | 48.2 ± 7.4 | 40.5 ± 2.6 | 40.2 ± 1.9 | 39.6 ± 3.3 | 69.3 ± 2.7 | 49.2 ± 4.6 |
| 13 | 42.5 ± 2.3 | 40.9 ± 4.9 | 48.0 ± 1.5 | 46.5 ± 3.1 | 65.7 ± 3.6 | 60.5 ± 1.5 | 54.0 ± 4.1 | 64.9 ± 21.4 |
| Mean | 39.2 ± 13.9 | 37.0 ± 14.8 | 35.6 ± 11.9 | 34.2 ± 10.8 | 42.7 ± 19.2 | 39.9 ± 18.1 | 45.2 ± 23.4 | 42.2 ± 20.8 |
Average SNR values are listed for tDWI and dDWI for the b900 and b2000 images in vitro at 1.5 T and 3.0 T. Wilcoxon signed rank P values between eSNR of tDWI and dDWI are 0.273, 0.340, 0.006, and 0.080 for b900 at 1.5 T, b2000 at 1.5 T, b900 at 3.0 T, and b2000 at 3.0 T, respectively.
Figure 2.Representative diffusion-weighted imaging (DWI) images and ADC maps using tDWI and dDWI of a dedicated NIST diffusion phantom (containing 13 vials) at 1.5 T (see Table 1 for sequence parameters). Similar ADC quantification and image quality was observed between tDWI and dDWI.
Average Quality Scores in the Diffusion Phantom
| 1.5 T | 3.0 T | |||
|---|---|---|---|---|
| tDWI | dDWI | tDWI | dDWI | |
| b500 | ||||
| Anatomic distortion | 4.0 ± 0.0 | 4.0 ± 0.0 | 3.0 ± 0.0 | 3.0 ± 0.0 |
| Image sharpness | 4.0 ± 0.0 | 4.0 ± 0.0 | 3.5 ± 0.7 | 3.0 ± 0.0 |
| Artifacts | 4.0 ± 0.0 | 4.0 ± 0.0 | 4.0 ± 0.0 | 4.0 ± 0.0 |
| Sum | 12.0 ± 0.0 | 12.0 ± 0.0 | 10.5 ± 0.7 | 10.0 ± 0.0 |
| b900 | ||||
| Anatomic distortion | 4.0 ± 0.0 | 4.0 ± 0.0 | 3.0 ± 0.0 | 4.0 ± 0.0 |
| Image sharpness | 4.5 ± 0.7 | 4.0 ± 0.0 | 3.5 ± 0.7 | 4.0 ± 0.0 |
| Artifacts | 4.0 ± 0.0 | 4.0 ± 0.0 | 4.0 ± 0.0 | 4.0 ± 0.0 |
| Sum | 12.5 ± 0.7 | 12.0 ± 0.0 | 10.5 ± 0.7 | 11.0 ± 0.0 |
| b2000 | ||||
| Anatomic distortion | 4.5 ± 0.7 | 4.5 ± 0.7 | 4.0 ± 0.0 | 3.5 ± 0.7 |
| Image sharpness | 5.0 ± 0.0 | 5.0 ± 0.0 | 4.0 ± 0.0 | 4.5 ± 0.7 |
| Artifacts | 4.0 ± 0.0 | 4.0 ± 0.0 | 4.0 ± 0.0 | 4.0 ± 0.0 |
| Sum | 13.5 ± 0.7 | 13.5 ± 0.7 | 12.0 ± 0.0 | 12.0 ± 1.4 |
| ADC | ||||
| Overall quality | 4.5 ± 0.7 | 4.5 ± 0.7 | 3.0 ± 0.0 | 3.5 ± 0.7 |
The average quality scores are those from 2 observers for in vitro tDWI and dDWI measurements at 1.5 T and 3.0 T.
Abbreviations: ADC, apparent diffusion coefficient; dDWI, diagonal diffusion-weighted imaging; tDWI, 3-scan trace diffusion-weighted imaging.
Figure 3.Representative images using tDWI and dDWI in the abdomen and pelvis of a 23-year-old male volunteer at 1.5 T. In the abdomen, the image quality was scored similar for both acquisitions (tDWI: image quality total score = 12.5 at b400, 12.5 at b800, and overall quality score = 5 for ADC; dDWI image quality total score = 12 at b400, 12 at b800, and overall quality score = 5 for ADC). In the pelvis, the image quality scored better in terms of image sharpness and artifacts for dDWI compared with tDWI (tDWI: image quality total score = 8 at b1000, 7 at b1600, and overall quality score = 3 for ADC; dDWI image quality total score = 10 at b1000, 10 at b1600, and overall quality score = 3.5 for ADC).
ROI Size Range, Average ADC and CV Values in the Abdomen and Pelvis
| Tissue | ROI Size (Range of Number of Pixels) | 1.5 T | 3.0 T | CV (1.5 T vs. 3.0 T) | |||||
|---|---|---|---|---|---|---|---|---|---|
| tDWI | dDWI | CV | tDWI | dDWI | CV | tDWI | dDWI | ||
| Liver | 43–55 | 1.24 ± 0.08 | 1.28 ± 0.03 | 2.6 | 1.32 ± 0.06 | 1.20 ± 0.19 | 6.8 | 5.6 | 9.1 |
| Gallbladder | 20–28 | 3.02 ± 0.09 | 3.06 ± 0.33 | 4.0 | 3.22 ± 0.49 | 3.23 ± 0.09 | 6.3 | 9.1 | 6.9 |
| Pancreas | 16–39 | 1.39 ± 0.01 | 1.43 ± 0.09 | 2.6 | 1.63 ± 0.26 | 1.36 ± 0.25 | 13.3 | 10.9 | 12.4 |
| Spleen | 36–51 | 0.86 ± 0.04 | 0.77 ± 0.03 | 7.6 | 0.76 ± 0.15 | 0.73 ± 0.01 | 10.0 | 9.1 | 4.1 |
| Renal cortex | 2–5 | 1.94 ± 0.05 | 1.96 ± 0.10 | 1.3 | 1.91 ± 0.10 | 1.88 ± 0.04 | 1.7 | 1.4 | 2.9 |
| Renal medulla | 4–7 | 1.85 ± 0.04 | 1.93 ± 0.17 | 5.5 | 1.90 ± 0.07 | 1.98 ± 0.11 | 2.9 | 1.7 | 7.1 |
| Prostate PZ | 9–12 | 1.07 ± 0.05 | 1.11 ± 0.00 | 2.5 | 1.15 ± 0.05 | 1.11 ± 0.03 | 2.6 | 4.7 | 1.4 |
| Prostate TZ | 10–13 | 0.90 ± 0.04 | 0.86 ± 0.02 | 2.7 | 0.94 ± 0.01 | 0.99 ± 0.11 | 5.7 | 3.4 | 9.5 |
| Urinary bladder | 34–86 | 1.93 ± 0.04 | 1.94 ± 0.01 | 0.6 | 2.34 ± 0.09 | 2.21 ± 0.29 | 4.6 | 13.5 | 8.6 |
This table presents the ROI size range, average ADC values (10−3 mm2/s), and CV values (%) between tDWI and dDWI and between 1.5 T and 3.0 T in abdominal and pelvic organs of 2 healthy volunteers. For organs/tissues in which multiple ROIs were drawn (ie, liver, pancreas, spleen, renal cortex, renal medulla, and prostate peripheral zone), the average values of all ROIs are listed.
Abbreviations: dDWI, diagonal diffusion-weighted imaging; PZ, peripheral zone; ROI, region of interest; tDWI, 3-scan trace diffusion-weighted imaging; TZ, transitional zone.
Average Quality Scores in the Abdomen and Pelvis
| 1.5 T | 3.0 T | |||||
|---|---|---|---|---|---|---|
| tDWI | dDWI | tDWI | dDWI | |||
| b400 abdomen | ||||||
| Anatomic distortion | 4.5 ± 0.0 | 4.3 ± 0.4 | 0.317 | 3.0 ± 0.0 | 3.3 ± 1.1 | 0.563 |
| Image sharpness | 4.0 ± 0.0 | 4.0 ± 0.0 | 1.000 | 3.0 ± 0.0 | 3.5 ± 0.0 | 0.157 |
| Artifacts | 4.0 ± 0.0 | 4.0 ± 0.0 | 1.000 | 3.0 ± 0.0 | 2.8 ± 0.4 | 0.317 |
| Sum | 12.5 ± 0.0 | 12.3 ± 0.4 | 0.317 | 9.0 ± 0.0 | 9.5 ± 0.7 | 0.414 |
| b800 abdomen | ||||||
| Anatomic distortion | 4.5 ± 0.0 | 4.3 ± 0.4 | 0.317 | 2.8 ± 0.4 | 4.0 ± 0.0 | 0.059 |
| Image sharpness | 4.0 ± 0.0 | 4.0 ± 0.0 | 1.000 | 2.8 ± 1.1 | 3.8 ± 0.4 | 0.102 |
| Artifacts | 4.0 ± 0.0 | 4.0 ± 0.0 | 1.000 | 2.8 ± 0.4 | 3.3 ± 0.4 | 0.157 |
| Sum | 12.5 ± 0.0 | 12.3 ± 0.4 | 0.317 | 8.3 ± 1.8 | 11.0 ± 0.7 | 0.065 |
| ADC abdomen | ||||||
| Overall quality | 5.0 ± 0.0 | 5.0 ± 0.0 | 1.000 | 3.0 ± 0.0 | 3.3 ± 0.4 | 0.317 |
| b1000 pelvis | ||||||
| Anatomic distortion | 2.8 ± 0.4 | 3.0 ± 0.0 | 0.317 | 3.5 ± 0.7 | 3.0 ± 1.4 | 0.157 |
| Image sharpness | 2.0 ± 0.0 | 2.8 ± 0.3 | 0.083 | 3.5 ± 0.0 | 3.0 ± 1.4 | 0.414 |
| Artifacts | 2.5 ± 0.7 | 3.5 ± 0.7 | 0.102 | 3.5 ± 0.7 | 3.8 ± 0.4 | 0.317 |
| Sum | 7.3 ± 1.1 | 9.3 ± 1.1 | 0.063 | 10.5 ± 1.4 | 9.8 ± 3.2 | 0.276 |
| b1600 pelvis | ||||||
| Anatomic distortion | 2.3 ± 0.4 | 3.0 ± 0.0 | 0.083 | 2.5 ± 0.7 | 2.5 ± 0.7 | 1.000 |
| Image sharpness | 2.0 ± 0.0 | 2.8 ± 0.4 | 0.083 | 2.5 ± 0.7 | 2.5 ± 0.7 | 1.000 |
| Artifacts | 2.3 ± 0.4 | 3.8 ± 0.4 | 0.063 | 2.8 ± 0.4 | 2.5 ± 0.7 | 0.563 |
| Sum | 6.5 ± 0.7 | 9.5 ± 0.7 | 0.066 | 7.8 ± 0.4 | 7.5 ± 2.1 | 0.705 |
| ADC pelvis | ||||||
| Overall quality | 3.0 ± 0.0 | 3.5 ± 0.7 | 0.157 | 3.5 ± 0.7 | 3.8 ± 1.1 | 0.317 |
The average quality scores were calculated in the abdomen and pelvis using tDWI and dDWI measurements of 2 volunteers at 1.5 T and 3.0 T.
aWilcoxon signed rank test.